EMS 110 Manual de usuario

User Manual
PRECISION ULTRASOUND
BALANCE
Model 110
EMS Physio Ltd
.
Grove Technology Park
Downsview Road
Wantage
Oxfordshire OX12 9FE
England

Precision Ultrasound Balance
2
Declaration of Conformity
NOTICE: This equipment has been tested and found to comply with the
limits class A digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial
environment.
This equipment generates, uses, and can radiate radio frequency energy
and if not installed and used in accordance with the instruction manual,
may cause interference to radio communications.
Operation of this equipment in a residential area may cause
unacceptable interference to radio and TV reception requiring the
operator to take whatever steps are necessary to correct the interference
at his own expense.

Precision Ultrasound Balance 3
General Information
This manual provides the necessary information for the installation and
operation of the Precision Ultrasound Balance.
These instructions must be studied before putting the unit into operation.
The information contained in this manual is subject to change without
notice.
No part of this manual may be photocopied, reproduced or translated into
another language without the prior written consent of EMS Physio Ltd.
Record of Amendments
ISSUE COMMENTS DATE
3 Revised 10/04/2007
4 Revised 19/10/2011

Precision Ultrasound Balance
4
Contents
Page
Title 1
Declaration of Conformity 2
General Information 3
Record of Amendments 3
Contents 4
Warranty 5
Introduction 6
Technical Specifications 7
Unpacking 8
Assembled Balance Schematic (Fig 1) 9
Assembly 10
Controls and Markings 11
Operating Instructions 12
Maintenance 13
Troubleshooting 14
Parts List 15

Precision Ultrasound Balance 5
Warranty
This EMS Physio Ltd., (hereinafter called the company) product is
warranted against defects in materials and workmanship for a period of two
years from the date of shipment. The Company will at its option, repair or
replace components which prove to be defective during the warranty
period, provided that the repairs or replacements are carried out by the
Company or its approved agents.
The Company will consider itself responsible for the effects on safety,
reliability and performance of the product:-
only if assembly operations, re-adjustments, modifications or repairs are
carried out by persons authorised by it,
only if the product is used in accordance with the instructions for use,
only if the electrical installation of the relevant room complies with the
appropriate national requirements.
Should the product be returned to the Company for repair it must be sent
carriage paid.

Precision Ultrasound Balance
6
Introduction
The EMS Precision Ultrasound Balance has been developed in response to
the increasing demand for equipment capable of testing and calibrating
ultrasound therapy units.
The balance is based on the principle of measuring the radiation force
produced on a suitable target by an ultrasound transducer.
The force is measured using a precision digital electronic balance
programmed for direct read-out in watts. The unit is portable with carrying
case.
Due to the fact that the unit is factory set, if any problems arise with the
unit it is highly recommended that the unit is returned for service and re-
calibration.
Environmental Considerations
For best results the Precision Ultrasound Balance should only be used in an
environment free from excessive air currents or vibration. Do NOT install
the balance near open windows or doors causing draughts or rapid
temperature fluctuations, near heating or air-conditioning vents, near
vibrating or rotating machinery, or on an uneven surface.

Precision Ultrasound Balance 7
Technical Specification
Accuracy (Scale factor) 10%
Resolution 0.02 watts
Linearity <0.1 watts
Stability (short term) <0.1 watts/10 minutes
Zero Drift ( after warm up) <0.1 watts/hour
Warm up Time <10 minutes
Full Scale 30 watts
Size 222 x 300 x 330 mm
Weight 4.0 kg (complete)
Water Capacity 1 litre
Recommended Frequency Range 0.8 to 3.5 MHz
Mains adaptor
Power input (AC) 100 mA @ 240V
Power output (AC) 500 mA @12V
Other mains adaptors are available on request.
An optional battery pack is available.
All the above are typical at 23 C.
The balance scalefactor has been programmed assuming the
temperature of the de-gassed water is 23 C. Readings will be
approximately 2% low if the actual water temperature is 13 C, and 2%
high for 33 C.
The EMS Precision Ultrasound balance is supplied in a carrying case.

Precision Ultrasound Balance
8
Unpacking
Upon receipt of the unit check for any damage which may have
occurred in transit. If any signs of damage are found then retain all
packing material and inform the carrier and the Company or its agent
from whom the unit was purchased.
Each unit is supplied with a low voltage adaptor and cable. The adaptor,
cable and unit all conform to international safety standards.
Check that the mains voltage and frequency indicated on the mains
adaptor are as required. If the mains supply is NOT within the specified
range DO NOT CONNECT THE MAINS ADAPTOR TO YOUR
SUPPLY.

Precision Ultrasound Balance 9
Target
Assembly Clamp
Support stem
S
upport Arm
Assembly
Container
Electronic Balance
Tare (Zero) Button (x2)
Centralizing
Plate (Various)
Display
ON/OFF
Button
Figure 1 – Assembled Balance
Level Indicator
Enter Button
Select Button

Precision Ultrasound Balance
10
Assembly
Refer to figure 1 - Assembled Balance
1. Withdraw the balance from the carrying case and place on flat and
level surface.
2. Assemble the stand by screwing the support stem into the
mounting block on the corner of the stand.
3. Level the balance using the adjustable feet, and check with the
built in level indicator.
4. Carefully withdraw the target and container assembly from the
carrying case.
5. Place the container on top of the stand, taking care to ensure that
the weighbridge spindle is engaged on top of the loadcell and that
the container is central between the location pegs of the stand.
6. Select a suitable workpiece clamp and fit to the stand support stem
(the unit is supplied with quick release clamps for use with EMS
heads and an adjustable clamp for use with other styles and sizes
of heads).
7. The target container can now be filled with suitably degassed
water. (To degas the water, it should be boiled and left to cool in a
suitable covered container - see IEC 150.) Sodium Sulphide (at a
ratio of 4g/litre) may be added to help degassing the water if
required.
8. Once the container has been filled with degassed water, a suitable
centralising plate may be fitted. (There are five sizes of plates for
use with various EMS heads.)
9. If the source being tested cannot be used with a suitable
centralising plate then ensure that the source is central to the target
otherwise inaccurate readings may be obtained.
10. Before use, the target frame should be checked to see that it is not
in contact with the head centralising plate.
11. Connect the output lead from the mains adaptor to the socket on
the rear of the balance. Connect the mains adaptor to a suitable
mains outlet.
The PRECISION ULTRASOUND BALANCE is now ready for use.