Emec ECL6 Manual de usuario

OPEN AMPEROMETRIC CELLS
OPERATING MANUAL
EN
ECL6 - ECL6E - ECL12 - ECL12E

INDEX
GENERAL SAFETY GUIDELINES ...................3
PURPOSE OF USE AND SAFETY ...................4
ENVIRONMENTAL SAFETY...........................5
SPARE PARTS...............................................5
INTRODUCTION...........................................7
Open amperometric cells ................................ 7
MODELS ........................................................ 8
Packaging ...................................................... 8
ECL6.............................................................. 9
ECL6/E........................................................... 9
ECL12............................................................ 10
ECL12/E......................................................... 10
Operating principle......................................... 11
Precautions .................................................... 11
Installation..................................................... 12
pH/ORP/temperature probes installation ......... 13
Commissioning............................................... 13
Electrode replacing......................................... 14
Electrode cleaning.......................................... 14
Calibration ..................................................... 15
Decommissioning .......................................... 15
MAINTENANCE ...........................................16
Maintenance schedule.................................... 16
Maintenance inspection ................................ 16
Calibration procedure..................................... 16
TROUBLESHOOTING....................................17
Repair service................................................. 17
ECL6 ............................................................. 18
ECL6/E........................................................... 19
ECL12............................................................ 20
ECL12/E......................................................... 21

3
⎘
This operating instructions contains safety information that if ignored can endanger life or result
in serious injury.
The original instruction is in English.
Read these instructions carefully before use and keep them for future reference.
Information and specifications on this manual could be uncorrect or could have printing errors.
Specifications are subject to change without notice.
Version: R1-02-17
D.M. 7 Febbraio 2012 n.25 – D.M.6 Aprile 2004 n.174 – Regolamento UE 10/2011
Apparecchiature finalizzate al trattamento dell’acqua destinata al consumo umano/
Equipment intended to come into contact with food/
Materiales y objetos plasticos destinados a entrar en contacto con alimentos.
Danger!
Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
Warning!
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
Important - A practice not related to personal injury or additional information.
Cross reference - An instance which refers to related information elsewhere in the same
document
GENERAL SAFETY GUIDELINES
Operating, installing, or maintaining the unit in any way that is not covered in this manual could
cause death, serious personal injury, or damage to the equipment.
This manual use the following safety message icon:
ICON

4
EQUIPMENT INTENDED FOR THE MEASUREMENT OF CHLORINE IN WATER.
Do not use in explosive area (EX).
Do not use with flammable chemicals.
Do not use with radioactive chemicals.
Use the equipment in accordance with the data and specifications printed on the label.
Do not modify or use in a manner inconsistent with the provisions of the operating manual.
When using this product with aggressive chemicals observe the regulations concerning
the transport and storage of aggressive fluids.
When installing always observe national regulations.
Manufacturer is not liable for any unauthorized use or misuse of this product that may
cause injury, damage to persons or materials.
Probes must be serviced and repaired by qualified and authorized personnel only.
Before any operation:
• always read chemical Material Safety Data Sheet (MSDS);
• always wear protective clothing;
• empty and rinse the liquid end before work on a the product which has been
used with hazardous or unknown chemicals.
Avoid grinding / shock / falls / friction.
PURPOSE OF USE
AND SAFETY

5
Work area
Always keep the area clean to avoid and/or discover emissions.
Recycling guidelines
EWC code: 16 02 16
Always recycle according to these guidelines:
1. If the unit or parts are accepted by an authorized recycling company, then follow local recycling
laws and regulations.
2. If the unit or parts are not accepted by an authorized recycling company, then return them to
the nearest representative.
Waste and emissions regulations
Observe these safety regulations regarding waste and emissions:
• Dispose appropriately of all waste.
• Handle and dispose of the dosed chemical in compliance with applicable environmental
regulations.
• Clean up all spills in accordance with safety and environmental procedures.
• Report all environmental emissions to the appropriate authorities.
Environmental
safety
For spare parts orders or any other communication, refer to the pump’s label.
Code (CODE) and serial number (P / N) uniquely identify the probe.
Spare parts

6
Transportation and
storage
Storage
An unsuitable transportation or storage can cause damages.
Use original box to pack the amperometric cell.
Observe storage conditions also for transportation.
Although packed, always protect the unit against humidity and the action of
chemicals.
Before returning the probe to the manufacturer repair service, refer to ⎘
Decommissioning.
DO NOT DISCARD PACKAGING. USE IT TO RETURN THE PROBE.
To store the probe refer to ⎘Decommissioning.
1. Clean the electrode, refer to ⎘ Cleaning procedure.
2. Store the probe in the original box, protected against sun, chemicals and humidity.
Transportation and storage temperature ..... +10°C - +50°C (+32°F - +122°F)
Humidity .................................................... 95% relative humidity (not condensed)

7
ECL Series is designed for measuring free chlorine (both organic and inorganic).
Open amperometric cells comprise an Off-line probe holders, a sensing electrode and a flow electrode.
Probe holders can contain up to three probes (temperature, pH and ORP).
The flow of water within this cell must remain constant and within 40 l/h. A pressure stabilizer is
available for areas subject to sudden pressure changes.
It is recommended to instal a filter before the probe holder.
Suitable for non-clean water thanks to the presence of cleaning balls.
Fig. 1. Description
SEPR
Water inlet valve
Water sampling
valve
Water sampling point
WATER INLET
WATER OUTLET
ELECTRODE
Temperature
pH or ORP probe
brown (+ supply)
blue (- supply)
black (PNP transistor
collector output)
RED (+)
BLACK (-)
Open
amperometric
cells
INTRODUCTION

8
MODELS • ECL6 for free chlorine (organic and inorganic) - pH and ORP probe-holder
• ECL6/E for free chlorine (organic and inorganic)
• ECL12 for free chlorine (organic and inorganic) for salt water - pH and ORP probe-holder
• ECL12/E for free chlorine (organic and inorganic) for salt water
Packaging Packaging includes:
• AMPEROMETRIC CELL
• COD. 11070081: KIT SCREW + WASHER +SCREW COVERS
• COD. 11067291: STANDARD KIT (1/2” PP+FP NIPPLE; 1/2” PVC GREY BALL VALVE M/M; FLAT
GASKET 1/2” F 1/4” M EPDM)
• COD. 02502121: 1/2” 6X8 HOSE HOLDING KIT FOR BLACK PP VALVE
• 4 MT 6X8 PE HOSE

9
ECL6
ECL6/E ECL6 ECL6/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) / BROMINE
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/copper; on request gold/copper)
Ph working range 6-8 pH
Run-in-time First polarization: 2 h about
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
adjustment See Operating manual: “Calibration”
Alcalinity 100 ppm
Working
temperature 5-40° C (41-104°F)
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
holder for pH, ORP and temperature temperature
Fittings for
connection to the
sample pipeline
6x8
Material Electrode: platinum/copper
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Maintenance
Regular control of the signal
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

10
ECL12
ECL12/E ECL12 ECL12/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/silver)
Ph working range 6-8 pH
Run-in-time First polarization: 2 h about
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
adjustment See Operating manual: “Calibration”
Slope calibration See Operating manual: “Calibration”- DPD1 method
Alcalinity 100 ppm
Working
temperature 5-40° C (41-104°F)
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
holder for pH, ORP and temperature temperature
Fittings for
connection to the
sample pipeline
6x8
Material Electrode: platinum/silver
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Maintenance
Regular control of the signal
SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Este manual sirve para los siguientes modelos
3
Tabla de contenidos
Otros manuales de Instrumento de medición de Emec
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario