3. In the “SELECT BUFFER SET” screen select option 2 and press READ to select a pH 4,7 &10
at 20 degrees.
4. 0r Select Option 3 to input custom Buffer values and Press READ
5. In the USER BUFFER SET screen select the option 1 and press right arrow
6. Press the down arrow to select CANCEL. Press READ to remove the current value.
7. To select a custom buffer highlight the value and press READ.
8. When you have the correct value in the ENTRY box select ENTER and press READ.
9. Repeat for Custom buffers 2 and 3. Note: You can still perform 1,2 or 3 point Calibrations. i.e.
you don’t have to use all 3 buffers.
10. Press READ to return to the PARAMETERS ADJUST screen and press READ again to
revert to the pH Mode.
11. To calibrate follow the Calibration procedure above.
Setting the Time and Date.
1. Switch the unit on using the on/off button.
2. Press the right arrow Key to Enter the PARAMETERS ADJUST SCREEN
3. Scroll down to the DATE FORMAT option and press READ until you arrive at the desired
format.
4. Scroll down to the CLOCK option using the down arrow key.
5. Press READ to enter the CLOCK screen.
6. Select CANCEL using the down arrow key and press READ until the current DATE in the
entry box is removed. Select the current date digit by digit by selecting the correct number and
pressing READ.
7. When the entry box contains the correct date select ENTER and press READ.
8. You are now in the TIME screen. Repeat from 5 above to select the correct time.
9.
From the PARAMETERS ADJUST screen press READ to return to the pH Mode
.
Logging and printing
The QP451 Portable pH Meter has both a log and a print function. You can print and log directly to
an external device when connected via the Mini USB output or straight to file. There is capacity to
store 64 files. The oldest will get overwritten if used files are not deleted.
The data logger will accept up to 10,000 data points. If you have selected a log period /log interval
combination that exceeds this you will be advised to amend your parameters.