User’s Guide
Version 1.1
Date: 2015-07-28

2 | P a g e
Table of Contents
1. Probe Description ............................................................................ 3
2. TECA™ Technology ........................................................................... 4
3. Getting Started ............................................................................... 6
4. Setting Up Your Inspection................................................................ 11
4a. Opening an Existing Setup.......................................................................11
4b. Creating a New Setup with the Setup Wizard ...............................................13
5. Nulling the Sharck Probe .................................................................. 18
6. Calibrating the Sharck Probe ............................................................. 19
7. Collecting Data .............................................................................. 21
8. Saving Data................................................................................... 24
9. Analyzing Data............................................................................... 25
9a. Available C-Scans..................................................................................25
9b. Analysis Methodology.............................................................................25
9c. Analyzing Data with Reddy ......................................................................27
10. Saving and Reporting ..................................................................... 30

3 | P a g e
1. Probe Description
The medium Sharck™probe used in this guide contains two rows of 11 independent fingers. This
array probe (SHARCK-W053-R-N05S) offers a coverage or active area of 53 mm (2.1 in). Other
sizes of Sharck array probes are also available.
Figure 1 Medium Shark probe
Note
The small Sharck probe is compatible with the 32-channel and 64-channel Reddy™, while the
medium Sharck is best suited for the 64-channel Reddy. The large Sharck probe is not
compatible with Reddy, as it requires 128 channels (Ectane® 128 or 256).
All the fingers of Sharck probes (patent pending) contain a three-coil assembly operating in a
specific transmit/receive mode called tangential that enables measuring variations of the eddy
currents distribution. This is referred to as tangential eddy current array or TECA™. TECA was
developed to detect and size the length and depth of axial surface-breaking linear indications
in carbon steel. It is also capable of also detecting transverse cracks (perpendicular to the scan
axis). TECA’s coils use two fixed frequencies: 20 kHz and 80 kHz.

4 | P a g e
All Sharck probes are equipped with a high-resolution magnetic wheel encoder
(20.53 points/mm). SHARCK-W053-R-N05S comes with a 12-pin encoder connector for Reddy,
and a standard 5 m cable. The encoder (and guiding wheels) can easily be removed for cleaning
and maintenance.
Figure 2 Sharck encoder and guiding wheels are removable
For more information about the Sharck (description, specifications and typical performance),
refer to the Probe Information document provided with the probes and the 4-page brochure.
2. TECA™ Technology
TECA’s coil arrangement and tangential operation mode enable obtaining a very particular eddy
current signal for surface-breaking cracks in carbon steel. The TECA pancake coil’s axis is
parallel to the surface under test instead of being perpendicular, as is typical in eddy current
testing when using pancake coils. As illustrated below, this coil orientation helps eddy currents
to dive underneath cracks, offering valuable information about their depth.
Each finger of the Sharck probes contain three coils: two tangential operating in
transmit/receive mode and a third whose axis is perpendicular to the surface under test. The
tangential coils are used to measure the depth of cracks, while the third coil is used to position
cracks, measure their length, and to detect transverse cracks.
Figure 3 Simplified TECA working principle
Coils scanning the surface
Induced eddy currents

5 | P a g e
The main characteristics of TECA signals are:
•Almost flat liftoff signal
•Crack-like indications approximately 90° relative to the liftoff signal
•All crack-like indications feature the same phase shift
•Signal amplitude linked to the defect depth
Figure 4 Liftoff and phase shift
Obviously, the vertical amplitude of a given indication is impacted by lift off. As shown above,
the vertical component of the signal can be reduced by a factor 2 or 3 when there is significant
lift off. However, since the coil design allows for monitoring the lift off, it can be measured in
order to adjust depth sizing accordingly. Ultimately, the technique relies on multiple depth
curves that are calibrated with various lift off values. The result is equivalent to a 3D “depth
plane” allowing to correlate crack depth, vertical amplitude and lift off (horizontal component),
as illustrated in the next figure.
Figure 5 A 3D “depth plane”is integrated into the software to correlate
crack depth, vertical amplitude and liftoff

6 | P a g e
3. Getting Started
Figures below summarise the names given to the various part of the Reddy’s interface.
Figure 6 Backstage window
Figure 7 Front stage window
Ribbons
Status
icons
Display
area
File
navigator
Collapsable tables
Readings

7 | P a g e
1. Connect the Sharck probe to Reddy: both the 160-pin ECA connector and the 12-pin encoder
connector (I/O).
2. Turn on Reddy. (Refer to the Reddy user’s guide for details.)
Notes
•You can connect and disconnect the Sharck probe while the instrument is running.
•Reddy performs a compatibility check on the probe to verify that the current setup is valid.
3. In the backstage view, tap Project.
4. Create a new inspection folder or open an existing one by tapping Create Inspection or
Open Inspection, respectively.
Figure 8 Backstage view: Create/Open inspection

8 | P a g e
Figure 9 Creating an inspection project
5. If you are creating a new inspection folder, type its name in the Inspection text box.
6. Tap OK.
Note
When you open a new inspection folder for the first time and press on Start Working, the
Acquisition Summary dialog box appears. The information you specify here is associated to this
inspection folder. You can tap Acquisition Summary in the backstage view to open it again at
a later time. When you are done, tap OK.
Figure 10 Acquisition Summary dialog box
7. Create a new setup or open an existing setup by tapping Create New Setup or Open Setup,
respectively.

9 | P a g e
Figure 11 Backstage view: Create/Open setup
8. Once you are done, tap Start Working.
Figure 12 Backstage view: Start Working

10 | P a g e
Figure 13 Front stage view
Tabla de contenidos
Otros manuales de Instrumento de medición de eddyfi
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario