Denver Instrument MAXX Manual de usuario

Denver MAXX
Electronic Precision Scales/Balances
Elektronische Präzisionswaagen
Balances électroniques d’analyse et de précision
Balanzas electrónicas de precisión
Operating Instructions/Betriebsanleitung/
Mode d’emploi/Manual de instrucciones

English page 3
In cases involving questions of
interpretation, the German-language
version shall prevail.
Deutsch Seite 20
Im Auslegungsfall ist die
deutsche Sprache maßgeblich.
Français page 37
En cas de questions concernant
l’interprétation, la version
en langue allemande fera autorité.
Español página 54
En caso de interpretación,
la versión en lengua alemana
será determinante.
2

3
Contents
Page
Warnings and Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Getting Started. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Basic Weighing Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Description of the Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Application Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Toggling between Weigh Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Counting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Weighing in Percent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Display Lock Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Total/Formulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Specific Gravity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Calibration/Span Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Advanced Features (Setup Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Error Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
CMarking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

4
Installation
– It is recommended to connect
only Denver accessories and
options, as these are optimally
designed for use with your
balance/scale.
– Do not open the balance/scale
housing as this will void the
manufacturer’s warranty.
Symbols
The following symbols are used
in these instructions:
●indicates steps you must
perform
$indicates steps you must
perform only under certain
conditions
>describes what happens after
you have performed a certain
step
–indicates an item in a list
!indicates a hazard
Warnings and Safety Precautions
Safety Information
●To prevent damage to the equip-
ment, please read these operat-
ing instructions carefully before
using your balance/scale.
!Do not use this equipment in
hazardous areas.
!Make sure the voltage rating
printed on the power supply is
identical to your local line volt-
age.
!Use only commercially available
9V batteries (not included).
If desired, you can use a
rechargeable battery.
– The balance/scale is energized at
all times unless you disconnect
the AC adapter and, if connect-
ed, the battery.
– Protect the AC adapter from
contact with liquid
!Exposure to excessive electro-
magnetic interference can cause
the readout value to change.
Once the disturbance has
ceased, the instrument can be
used again in accordance with
its intended purpose.

5
Getting Started
Equipment Supplied
– Balance/scale
– Weighing pan
– AC adapter
Additional equipment with models
MXX-612, MXX-412, MXX-212:
– Calibration weight
Additional equipment with model
MXX-123:
– Calibration weight
– Round glass draft ring
– Level indicator and adjustable feet
Installation
Choose a location that is not subject
to the following negative influences:
– Heat (heater or direct sunlight)
– Drafts from open windows and
doors
– Extreme vibrations during
weighing
– Excessive moisture
Setting Up the Balance/Scale
●Place the components on the
balance/scale in the following
order:
– Reversible weighing pan
– Round glass draft ring on model
MXX-123

6
Connecting the Balance/Scale
to AC Power
!Use only the AC adapter for
optimal performance and safety.
§Insert plug into the jack (located
on back of balance/scale)
§Plug the AC adapter into an
electrical outlet
!LISTED power supply 11 V–21 V
compliant with NEC Class 2 out-
put.
Below-Balance/Under-Scale Weighing
A port for a below-balance weighing
hanger is located on the bottom of
the balance/scale.
§Open the cover plate on the
bottom of the balance/scale.
§Attach the sample (e.g., using
a suspension wire) to the hanger.
$Install a shield for protection
against drafts if necessary.
Installing the Battery
(not for model MXX-123)
$Batteries are not included with
the equipment supplied
!Use only commercially available
9V batteries.
!If you use a rechargeable bat-
tery, recharge it with an external
battery charger.
§Lay the balance/scale on its side
§Open the battery compartment:
remove the compartment cover
§Install the battery in the com-
partment
$Make sure the polarity is correct.
§Close the battery compartment:
slide the cover into position until
it snaps into place
!Do not throw away used batter-
ies with normal household
waste. Rechargeable batteries
contain toxic materials and must
be disposed of in accordance
with local waste disposal regula-
tions.

7
Leveling the Balance/Scale
(only for model MXX-123)
§Always level the balance/scale
again any time after it has been
moved to a different location.
§Turn the front feet as shown in the
diagram until the air bubble is cen-
tered within the circle of the level
indicator.
§Lower all feet to touch the counter-
top so the balance/scale does not
rock.
> In most cases this will require
several adjustment steps.
Removing Weights for
Calibration/Span Adjustment
(only for models MXX-612, MXX-412,
MXX-212, MXX-123)
§Grasp the tab to turn and pull out
the weight compartment.
§Always wear gloves or use tweez-
ers when handling weights.
$Follow instructions on page 15 for
calibration/span adjustment.
Warmup Time for Model MXX-123:
To ensure accurate results, the scale
must warm up for 1 hour before
operation. Only after this time will
the scale have reached the required
operating temperature.

Operation
Basic Weighing Function
Features
– Zeroing the balance/scale
You can zero the balance/scale with-
in the entire weighing range, up to
the maximum capacity.
Preparation
§Switch on the balance/scale:
press the (ON/OFF) key
§If necessary, zero the balance/
scale: press the (ZERO) key
$If necessary, change the configura-
tion settings: see the chapter
entitled “Configuration”
$If desired, reset the factory settings:
see the chapter entitled
“Advanced Features,” menu item
9.– 1
Additional functions:
$Switching off the balance/scale:
press the (ON/OFF) key
$Battery operation: automatic shut-off
after 2, 5 or 10 minutes; see chapter
on configuration. Example: 2 min-
utes. If the weight readout remains
unchanged and no keys are pressed
for at least two minutes, the battery
symbol “ “ starts flashing.
After another 2 seconds, the
balance/scale shuts off automati-
cally, unless a key is pressed.
Description of the Keys
(ON/OFF) On/off key: switches the
balance/scale on and off or
switches it to the standby
mode.
Battery operation: on;
turns backlight on; off
(ZERO) Zeros the balance/scale;
press and hold 2 seconds:
enters application menu
(CAL) Starts calibration/adjust-
ment
(F) Starts an application pro-
gram; scrolling in applica-
tion menu, configuration
menu and calibration menu
(ENTER) Confirms the selected
setting; exits application,
configuration & calibration
menu if key is pressed
and held for at least
2 seconds.
(PRINT) Generates a printout or
data output
Example: Determine weight of sample
Step Key (or instruction) Display
1. Switch on the balance/scale (ON/OFF)
Self-test runs
Display: Software version r 31.01 g
2. Place container on the balance/scale
(in this example, 52 g) 52.0 g
3. Zero the balance/scale (ZERO) 0.0 g
4. Place sample in container on
balance/scale (in this example, 150.2 g). 150.2 g
8

9
Application Programs
Toggling between Weigh Units
With this application program you can toggle the display of a weight value
back and forth between two weigh units (see table below).
Example: Toggle weigh unit from pounds [lb] (application) to grams [g]
(basic unit)
Step Key (or instruction) Display
1. Select application program (ZERO) ≥2 sec
2. Select Toggling between Weigh Units (F) repeatedly
3. Confirm unit of application* (ENTER)
4. Select weigh unit; in this example: (F) repeatedly
“5. Pounds” (see table below)
5. Confirm weigh unit (pounds) (ENTER)
6. Place sample on balance/scale
7. Toggle weigh unit (F)
Menu code Unit Conversion factor Display
1. user** Grams 1.00000000000 o
2.grams (factory setting) Grams 1.00000000000 g
4.carat Carats 5.00000000000 o
5.Pound Pounds 0.00220462260 lb
6.ounce Ounces 0.03527396200 oz
7.tryo Troy ounces 0.03215074700 ozt
8.tl.hon Hong Kong taels 0.02671725000 tlk
9.tl.sin Singapore taels 0.02645544638 tl
10.tl.ta Taiwanese taels 0.02666666000 tl
11.grai Grains 15.4323583500 GN
12.peny Pennyweights 0.64301493100 dwt
15.tl.ch Chinese taels 0.02645547175 tl
22.pdoz lb/oz 0.03527396200 lb:oz
23.ne t Newtons 0.00980665000 N
* Grams is chosen as 1st unit; to changes 1st unit, see Advanced Features section.
** User-defined conversion can be loaded in balance/scale with RS-232 or USB
program option. Contact Denver Instrument for more information.

10
Counting
Purpose
With the Counting program you can determine the number of parts or pieces
by weight.
Example: Determine the number of uncounted parts; calculate the sample
reference quantity (in this example: 20)
Step Key (or instruction) Display
1. Select application program (ZERO) ≥2 sec
2. Select Counting (F) repeatedly
3. Confirm application (ENTER)
Symbol “ “ on the display:
application is active
4. Place empty container on the
balance/scale
5. Zero the balance/scale (ZERO)
6. Place sample reference quantity (20)
on the balance/scale
7. Select sample reference quantity:
in increments of 1 (1, 2, 3, ..., 99) or (F) repeatedly (briefly) or
in increments of 10 (10, 20, 30, ...,100) (F) ≥2 sec.
8. Confirm selected sample reference
quantity (ENTER)
9. Place uncounted parts on balance/scale
10. Toggle display between mean
piece weight, total weight, and quantity (F) repeatedly
11. Unload the balance/scale
12. Start new Count application
or exit application (ENTER) ≥2 sec
13. Reactivate Counting (if no other (F)
application program has been selected)
14. Repeat procedure starting from Step 5.
Tabla de contenidos
Otros manuales de Escala de Denver Instrument

Denver Instrument
Denver Instrument Tl-series Manual de usuario

Denver Instrument
Denver Instrument XE-100A Manual de usuario

Denver Instrument
Denver Instrument TB-124 Manual de usuario

Denver Instrument
Denver Instrument TP Series Manual de usuario

Denver Instrument
Denver Instrument Summit Manual de usuario

Denver Instrument
Denver Instrument ACA-100 Manual de usuario

Denver Instrument
Denver Instrument PK-Series Manual de usuario

Denver Instrument
Denver Instrument TB Series Manual de usuario

Denver Instrument
Denver Instrument XP Series Manual de usuario

Denver Instrument
Denver Instrument P-114 Manual de usuario