Daiichi Electronics SDLC-105A Manual de usuario

SDLC-207-104
INSTRUCTION MANUAL
LCD ALARM SETTER
SDLC-105A

SDLC-207-104
1
Introduction
Thank you for your purchase of our product.
Read this instruction manual carefully before installation, wiring, and using this product.
Keep this instruction manual handy for reference at any time.
Have a contact with us or sales agent in case that this instruction manual is lost or damaged.
<Caution>
Have a contact with us when you have any questions or are aware of missing article.
Safety precaution
Important contents are mentioned in this instruction manual to prevent any damage/use this product appropriately.
Keep the following safety precaution in mind after understanding each sign.
Improper use may lead to death or severe injury.
Improper use may possibly lead to death or severe injury.
Improper use may lead to medium injury.
We are not responsible for the damage caused by following condition(earthquake/fire which is not caused by us,
action by third party, other accident, damage caused by our customer, misuse, product usage under abnormal
condition).
We are not responsible for secondary damage caused by product use/product malfunction (loss of profit, halt of
business operation). We are also not responsible for damage caused by false operation in combination with connecting
equipment which is beyond our control.
Do not disassemble, remodel and repair this product.
Have a contact with us or sales agent when product failure happens to prevent fire/electric shock/injury.
Do not get this product wet to prevent heat generation/ignition/product failure. When this product gets wet,
stop using it.
Do not connect metal excepting wiring to terminal in order to prevent heat generation/ignition.
Do not get this product near the inflammables/combustible chemicals/gas to prevent fire.
Connect specified power supply.
Connecting power supply beyond specification causes fire/product failure.
When dust is on the terminal, wipe it off after power is OFF to prevent fire.
Follow the below-mentioned procedure when abnormality (fuming/bad odor) happens.
(1) Stop using after power and input are OFF.
(2) Have a contact with us or sales agent.
Do not use this product in a environment of high temperature/high humidity to prevent any damage.
Do not touch the terminal during operation to prevent electric shock.
Do not pull/bend connecting cable with force. Cable damage causes heat generation/burn and contact failure
leads to equipment damage.
Do not connect/inspect with wet hands to prevent electric shock.
Other precaution
Don't mount or store this unit in the following environment.
Places where corrosive gas (SO
2
/ H
2
S / etc.)(1) is generated.
Places where dust is generated.
Places with much vibration and shock.
Places with influence of external magnetic field (2).
Note (1) Corrosive gas = Sulfur dioxide SO
2
/ Hydrogen sulfide H
2
S / etc.
Note (2) Large current bus / saturable reactor / etc.
Wipe off dirt on the surface with dry cloth softly. Keep in mind that strong rubbing of nameplate leads to
character disappearance. Organic solvent is not appropriate for cleaning.
Mercury component, Nickel-cadmium battery are not used in this product.
This product is disposed as an industrial waste (Non-inflammable).
This product is not dust-proof structure and not waterproof structure. In using by the outdoor panel, please
avoid the place, which dust causes.
And, please install in the place that requires neither rain nor waterdrop.

SDLC-207-104
2
Content
1. Explanation of product
1.1 Product outline ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
1.2 Features of product ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
1.3 Composition of type and specification code ・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
2. Specification and performance
2.1 Specification of alarm output ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4
2.2 Operation of alarm output ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4
2.3 Relay operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 6
2.4 Prealarm function ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7
2.5 Setting value ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 8
2.6 Performance ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 9
3. Installation
3.1 Outline dimension drawing ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 10
3.2 Mounting dimension drawing ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 10
3.3 Connection ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 11
3.4 Cautions on mounting ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 11
3.5 LCD viewing angle ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 12
4. Operation method
4.1 The name and function of each part ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 12
4.2 Each display mode and operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 13
4.3 Measurement mode ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 13
4.4 Set value display mode ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 14
4.5 Set value change mode ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 15
5. Block-diagram and principle of operation
5.1 Block-diagram ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 17
5.2 Principle of operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 17
6. Inspection
6.1 Receiving inspection ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 18
6.2 Periodic inspection ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 18
7. Check, calibration and maintenance
7.1 Check ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 18
7.2 Calibration ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 18
7.3 Trouble shooting ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 19
7.4 Countermeasures against troubles ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 19
8. About unit symbol seal ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 20

SDLC-207-104
3
1. Explanation of product
1.1 Product outline
This product is the alarm setter of the plug-in structure which considers a DC voltage (or DC current) signal
as an input, and outputs the relay contact signal of four points (maximum) as compared with the set-up operation
set value.
This product is the type corresponding to the software that has a LCD function and a scale can set up arbitrarily
according to the amount of processes. The real scale display function of an input, the dead band setting function,
the contact delay setting function, and the starting delay setting function are built-in. It can correspond
to various process instrumentation flexibly.
1.2 Features of product
Setting accuracy is ±0.1% (% for input span). Display accuracy is ±0.1%±1digit (% for input span).
Between input, output, auxiliary supply and cases, withstand voltage is AC2000V design.
The display of real scale by LCD. And, setting can carry out freely by the front button switch.
It can correspond to scaling change etc.
Nonvolatile RAM offers the power failure guarantee of the set value.
Sufficient derating of uses parts, and reduction of inside generation of heat improves reliability.
The back light is turned on at the case of key operation. 30 seconds after key operation is completed, the
back light puts out the light.
By the pre-alarm function, the efflux in contact delay time is displayed by blinking of a monitor lamp.
(ON/OFF of prealarm can be selected by setting.)
1.3 Composition of type and specification code
Type Specification code
SDLC-105A-
①
②
③
① Input (Input resistance) ② Output contact composition ③ Auxiliary supply
A3:DC0‐60mV (About 1MΩ) 1:Two alarm outputs (c contact) F:AC・DC 80‐264V 3.6VA
A5:DC0‐1V (About 1MΩ) 2:Three alarm outputs (a contact) Rated voltage
A6:DC0‐5V (About 1MΩ) 3:Three alarm outputs (b contact) AC100/110V,50/60Hz
AC200/220V,50/60Hz
DC100/110V
A7:DC0‐10V (About 1MΩ) 4:Four alarm outputs (a contact)
A8:DC1‐5V (About 1MΩ) 5:Four alarm outputs (b contact)
B3:DC±60mV (About 1MΩ) 0:Other
B5:DC±1V (About 1MΩ) 5:DC19‐29V 2.5W
B6:DC±5V (About 1MΩ) (Rated voltage DC24V)
B7:DC±10V (About 1MΩ)
C3:DC0‐1mA (About 100Ω)
C4:DC0‐5mA (About 100Ω)
C5:DC0‐10mA (About 100Ω)
C6:DC0‐16mA (About 100Ω)
C7:DC4‐20mA (About 100Ω)
00:Other
Production range
Voltage input:60mV to 60V
Current input:1mA to 50mA

SDLC-207-104
4
2. Specification and performance
2.1 Specification of alarm output
Item Specification
Output contact points 2 contact outputs,3 contact outputs,4 contact outputs
Contact composition
2 contact outputs:Each 1c contact
3 contact outputs:Each 1a contact or each 1b contact
4 contact outputs:Each 1a contact (Two-points of one side is common.) or
Each 1b contact (Two-points of one side is common.)
Contact capacity
Maximum switching load AC120V 1A (COSφ=1) ,DC30V 1A (Resistance load) or
DC125V 150mA (Resistance load)
Minimum switching load DC5V 10mA
Electric life 500,000 times or more (Switching frequency, 30 times/minute)
Output mode
Setting in five kinds of modes as follows is possible for operation of detection.
① High limit setting (H),Operation of detection:Relay excitation
② High limit setting (H),Operation of detection:Relay non-excitation
③ Low limit setting (L) ,Operation of detection:Relay excitation
④ Low limit setting (L) ,Operation of detection:Relay non-excitation
⑤ OFF ,It is always relay non-excitation.
Relay excitation
status display
In 4 contacts (maximum) of each, It is monitor lamp turn on the light at the case of relay
excitation. It is blinking among delay time at the case of the prealarm setting ON.
Relay and monitor lamp
operation
At the auxiliary supply OFF,
or at the relay non-excitation. Relay b contact "ON", Monitor lamp "OFF".
At the relay excitation Relay a contact "ON", Monitor lamp "ON".
Output mode Reference to "2.2 Operation of alarm output".
Relay contact status Reference to "2.2 Operation of alarm output".
2.2 Operation of alarm output
① At the auxiliary supply OFF, or at the relay non-excitation:Relay b contact "ON", Monitor lamp "OFF".
② At the relay excitation:Relay a contact "ON", Monitor lamp "ON".
Output mode Relay and monitor lamp operating state
▽ Operation value setting. Input →
Excitation
H
Monitor lamp ●:Turn on the light
Relay ○:Turn off the light
L
Monitor lamp
Relay
Non-excitat
ion
H
Monitor lamp
Relay
L
Monitor lamp
Relay
OFF
(Excitation・
Non-excitation)
Monitor lamp
Relay
Non-excitation Excitation
Turn off the light Turn on the light
Non-excitation
Excitation
Turn off the light
Turn on the light
Non-excitation
Excitation
Turn off the light
Turn on the light
Non-excitation Non-excitation
Turn off the li
g
ht Turn off the light
Non-excitation Excitation
Turn off the light Turn on the light

SDLC-207-104
5
● Relay contact status
Status 4 contact outputs:Each 1a contact 4 contact outputs:Each 1b contact
Auxiliary supply OFF
or relay
non-excitation
Terminal №
AL4AL3C2
10911
AL2AL1C1
213
Terminal №
AL4AL3C2
10911
AL2AL1C1
213
Relay excitation
Terminal №
AL4AL3C2
10911
AL2AL1C1
213
Terminal №
AL4AL3C2
10911
AL2AL1C1
213
Status 3 contact outputs:Each 1a contact 3 contact outputs:Each 1b contact
Auxiliary supply OFF
or relay
non-excitation
Terminal №
AL3
10911
AL1AL2
213
Terminal №
AL3
10911
AL1AL2
213
Relay excitation
Terminal №
AL3
10911
AL1AL2
213
Terminal №
AL3
10911
AL1AL2
213
Status 2 contact outputs:Each 1c contact
Auxiliary supply OFF or
relay non-excitation
Terminal №
AL2
10911213
c2 a2 b2
AL1
c1 a1 b1
Relay excitation
Terminal №
AL2
10911213
c2 a2 b2
AL1
c1 a1 b1

SDLC-207-104
6
2.3 Relay operation
■ Contact delay function
Relay operation in over-input detection. (Output mode:Excitation, H (3))
Operation value
setting
Relay action ON
OFF
Setting value
Input
Contact delay
setting time
Contact delay
setting time
Contact delay
setting time
Dead band
Relay operation in shortage-input detection. (Output mode:Excitation, L (3))
Operation value
setting
Relay action ON
OFF
Input
Contact delay
setting time
Contact delay
setting time
Contact delay
setting time
Dead band
Setting value
Note(3) In case of non-excitation for relay excitation operation setting, action of relay becomes reverse.
■ Starting delay function
After an auxiliary supply apply, detection action is
performed after a starting delay time.
Auxiliary supply
Detect action
Closing
Starting delay setting time

SDLC-207-104
7
2.4 Prealarm function
A monitor lamp is blinked among contact delay time until excitation of a relay after operation value detection.
Blinking of a monitor lamp will be gradually fast, and when the point of contact delay time passes, a monitor
lamp is lights up.
If contact delay time setting is made into 0 second, a prealarm function constitutes OFF.
・Relay action and pre-alarm in over-input detection.
(Output mode setting:Excitation H, Pre-alarm setting:ON)
Operation value
setting
Relay
action
ON
OFF
Dead band
Setting value
Input
Monitor
lamp
ON
OFF
Blinking
Lights up
Blinking
Contact delay
setting time
Contact delay
setting time
Contact delay
setting time
Lights up
・Action of the prealarm in contact delay time (blinking)
180s
3s
Blinking cycle
6s
Blinking cycle
4s
Blinking cycle
2s Blinking becomes
gradually fast.
3s
120s
3s
3s
2s
2s
1s
1s
2s
2s
60s 10s 0s

SDLC-207-104
8
2.5 Setting value
№ Contents Mark
Initial value
Setting range
2 contact
outputs
3 contact
outputs
4 contact
outputs
1 Unit display UNIT % %,m,℃,m3/h,ppm,pH,TPm,×10kg,Ω,kW
2 Minimum value scaling MIN 0.0 -9999 to 9999 -9.99 to 99.99
-99.9 to 999.9 0.000 to 9.999
3 Maximum value scaling MAX 100.0
4
AL1
Output mode ALMD1 High High High High:H operation,Low:L operation,
OFF :With no operation
5 Operation value COMP1 70.0 80.0 80.0 It sets in the measurement display range
of -25 to +125%, and is a real scale.
6 Dead band D.B.1 3.0 0.5 to 50.0%
7 Excitation mode EXMD1 0N ON:Excitation,OFF:Non-excitation
8 Contact delay time (CD) C.D.1 0 0 to 180s
9
AL2
Output mode ALMD2 Low High High High:H operation,Low:L operation,
OFF :With no operation
10 Operation value COMP2 30.0 70.0 70.0 It sets in the measurement display range
of -25 to +125%, and is a real scale
11 Dead band D.B.2 3.0 0.5 to 50.0%
12 Excitation mode EXMD2 0N ON:Excitation,OFF:Non-excitation
13 Contact delay time (CD) C.D.2 0 0 to 180s
14
AL3
Output mode ALMD3 - Low Low High:H operation,Low:L operation,
OFF :With no operation
15 Operation value COMP3 - 30.0 30.0 It sets in the measurement display range
of -25 to +125%, and is a real scale
16 Dead band D.B.3 - 3.0 0.5 to 50.0%
17 Excitation mode EXMD3 - ON ON:Excitation,OFF:Non-excitation
18 Contact delay time (CD) C.D.3 - 0 0 to 180s
19
AL4
Output mode ALMD4 - Low High:H operation,Low:L operation,
OFF :With no operation
20 Operation value COMP4 - 20.0 It sets in the measurement display range
of -25 to +125%, and is a real scale
21 Dead band D.B.4 - 3.0 0.5 to 50.0%
22 Excitation mode EXMD4 - ON ON:Excitation,OFF:Non-excitation
23 Contact delay time (CD) C.D.4 - 0 0 to 180s
24 Starting delay time (SD) S.D. 5 1 to 180s
25 Bias point calibration CBIAS 0.00 -9.99 to 9.99% (% for input span)
26 Span point calibration CSPAN 0.00 -9.99 to 9.99% (% for input span)
27 Average number AVE 1 1,4,8,16,32
28 Prealarm PREAL OFF ON:Prealarm ON, OFF:Prealarm OFF
< Notes of setting >
・The decimal point position change of scaling is made by the minimum value scaling.
If the decimal point position is changed, the decimal point position of maximum value scaling and operation values
1 to 4 will also change. If the decimal point position is changed, please be sure to check the set value of
a maximum value scaling and an operation value.
・An operation value can be set up in -25 to +125% of the measurement display range (minimum value scaling to
maximum value scaling).
(Example) Input 4 to 20mA, Operation value -25.0, Output mode is "Low" setting.
It becomes L detection in 0mA of inputs, it is also possible to use it as
disconnection detection.
・After setting change, in case a set value combination error occurs, it becomes a setting error display.
Please push the key and set up the right set value (in set value change mode).
125%
100%
50%
0%
-25%
24mA
20mA
12mA
4mA
0mA

SDLC-207-104
9
2.6 Performance
Item Specification
Setting accuracy (4) ±0.1% (% for input span)
Display accuracy (5)(6) ±0.1% (% for input span) ±1digit
Reproducibility of the
operating point (4) ±0.1% (% for input span)
Operating time accuracy ±0.25 seconds of contact delay time set value
(However, in case of set value = 0 second. 0.5±0.25 seconds)
Reset time Less than 0.5 seconds
Starting delay time
accuracy ±0.25 seconds of starting delay time set value
Influence of temperature 0.2% (For input span) /23±10℃
Influence of auxiliary
supply voltage variation 0.1% (For input span) /Within variation range
Operation period About 0.1 second
Response time About 0.5 second
(In case 90 to 110% of step variation of operation value setting is given, in CD=0 second)
Overload
capacity
Voltage input 2 times 10 seconds and 1.5 times continuation of rated voltage.
Current input 10 times 5 seconds and 1.5 times continuation of rated current.
Aux. supply 1.5 times 10 seconds of rated voltage. And upper limit of the variation range is continued.
Insulation resistance
Between electric circuit and case.
Above 50MΩ at DC500V.
Between input, output terminals and auxiliary supply terminals.
Between input terminals and output terminals.
Between AL1 output terminals and AL2 output terminals.
(Only two alarm contact output specification)
Between AL1 output terminals and AL2 output terminals
and AL3 output terminals.
(Only three alarm contact output specification)
Between AL1・AL2 output terminals and AL3・AL4 output terminals.
(Only four alarm contact output specification)
Power frequency
withstand voltage
Between electric circuit and case.
AC2000V (50/60Hz) 1 minute.
There must not be malfunction.
Between input, output terminals and auxiliary supply terminals.
Between input terminals and output terminals.
Between AL1 output terminals and AL2 output terminals.
(Only two alarm contact output specification)
Between AL1 output terminals and AL2 output terminals
and AL3 output terminals.
(Only three alarm contact output specification)
Between AL1・AL2 output terminals and AL3・AL4 output terminals.
(Only four alarm contact output specification)
Impulse withstand
voltage Between electric circuit and case.
5kV 1.2/50
μ
s.
Positive and negative
polarity for each 3 times.
There must not be malfunction.
Vibration Vibration of vibration frequency 16.7Hz,Double amplitude 1mm. In the direction of X Y Z for
each 10 minutes. There must not be malfunction.
Shock Error:98m/s
2
, X, Y, Z direction for each 3 times.
Duration:294m/s
2
, X, Y, Z direction for each 3 times.
Power failure guarantee Each set value is data-saved by non-volatilized memory.
Material of case Case:Flame resisting ABS resin , Name plate:Polyester film , Socket:Glass filled PBT resin
Appearance color Case, Socket:Munsell N1.5 (Black) , Name-plate:Gray
Operating temperature
and humidity limits -10 to +55℃ , 5 to 90%RH (Non condensing)
Storage temperature
limits -25 to +70℃
Mass Approx. 350g
Accessories Socket,Unit symbol sheet,Instruction manual
Note(4) Input is DC60mV to less than 1V, DC±60mV to less than ±1V. ±0.2% (% for input span)
Note(5) Input is DC60mV to less than 1V, DC±60mV to less than ±1V. ±0.2% (% for input span) ±1digit.
note(6) In more than 10000 (5 digits except decimal point), a measurement display span constitutes ±0.1% (%
for input span)±2digit. And, if input is less than 1V and measurement display span is 10000 or more.
±0.2% (% for input span) ±2digit.
Tabla de contenidos
Otros manuales de Instrumento de medición de Daiichi Electronics
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario