CTL- Contactless Temperature Logger detailed user guide version 1.2 Page 5
Push STOP button and/or by mobile app /or Memory full
-40 °C to +80 °C External Sensor
-20 °C to +80 °C Internal Sensor
±0.2°C (max) From –30°C to +60°C
No special interface required
1, Internal Digital or External
4 thresholds: HIGH-HIGH; HIGH; LOW; LOW-LOW
Alarm thresholds with LED visual
Red LEDs - HIGH / HIGH-HIGH
Blue LEDs - LOW / LOW-LOW
Orange (1): for Status
Orange (1): for RF
Full Memory Download Time
2 buttons with different activities:
On : Check status & manual log (less than a second)/RF on & off
(3-4 seconds)/Start (7-9 seconds)/Stop (10-12 seconds)
Status/Refresh : Check status (2 seconds)
CR2450 Panasonic 3.0V coin cell
95mm x 50mm x 9mm (with plastic sleeve)
IP 67, Completely Food Safe
Calibration on customer request
Pre-programmed and
User Programmable (special conditions apply)
Temperature Units °F and °C
Android 4.x or higher, iOS Windows XP, 7, 8, 10
Communicate with device, sync date/time, setup the device and log
parameters, download the records, calculate and analyze the data
and calculate statistic parameters (max, min, avg, MKT, log duration,
individual thermal zone durations)
Show data in table and chart, generate report