Error functions for MARK III PUMP in APo1 mode only
Error Message Description Action
Err Unfeasible operation No action required
Er 2 Safety system error Press button
Er 3 Irregularity of motor circuit Press button
Er 4 Obstruction of plunger during retraction Remove obstruction. Restart PUMP by removing battery and reinserting after 15 seconds
Er 5 Irregularity of plunger mechanism Press button
Er 6 Motor irregularity Restart PUMP by removing battery and reinserting after 15 seconds
Er 7 Internal communication error Press button
Er 8 Reading memory error Restart PUMP by removing battery and reinserting after 15 seconds
Er 9 Motor circuit error Restart PUMP by removing battery and reinserting after 15 seconds
OCCL Occlusion of infusion line Clear occlusion and press button
Er,d Bolus dose set to 0.00ml Function not available
Error functions
Error functions for MARK I PUMP only
Error Message Description Action
Err Unfeasible operation No action required
Err 1 Occurs when battery is inserted, some
initial setting may be wrong Remove and reinsert battery after 15 seconds
Err 2 The motor has stopped in order to avoid
overrunning Press button to restart
Err 3 Irregularity of motor circuit Press button to restart
Err 4 Obstruction of plunger during END
phase of retraction Remove the obstruction. Restart PUMP by removing battery and reinserting after 15 seconds
Err 5 Irregularity of plunger during retraction.
Plunger may be damaged Press button. Alternatively, remove battery and re-insert after 15 seconds
Err 7 Mechanical/motor block during backwards
movement Restart PUMP by removing battery and reinserting after 15 seconds
Err 8 Reading memory error Restart PUMP by removing battery and reinserting after 15 seconds
Err 9 Motor circuit error Restart PUMP by removing battery and reinserting after 15 seconds
Err 11 Flow rate is above maximum setting Restart PUMP by removing battery and reinserting after 15 seconds
OCCL Occlusion of infusion line Clear occlusion and press button
7 8
Administering the bolus dose
While the pump is on, press and hold until you hear a long beep - then release
The screen will show the bolus dose being infused with dflashing. Beeps will sound at every 0.11ml with a long beep when
the full dose has been administered. If required, the bolus dose can be stopped pressing the button
Note: If a lock out time has been set (APo3 mode only) this will limit the time between bolus doses
Refer to Setting the bolus dose lockout time on page 4 for more details
PAUSING THE INFUSION While the infusion is running press and hold ,
the display will now show S oP
The PUMP can be restarted by pressing and holding
Pausing the infusion
Retracting the plunger to the start position
NOTE: The infusion line must always be DISCONNECTED from the patient before the plunger is retracted
The syringe should not be removed until the plunger is fully retracted
RETRACTING IF THE INFUSION HAS NOT ENDED:
Stop by pressing and holding - display shows S oP, press to stop beeping
Disconnect the infusion line
End infusion by pressing and together - a long beep will sound and display will show End
then I---- as the plunger withdraws.
A beep will sound when completed and the display returns to OFF
RETRACTING AT THE END OF THE INFUSION:
NB - An intermittent beep will sound at 10 minutes and at 5 minutes before the end of infusion.
At the end of infusion an on-going beep will sound and the display will show
End
PUMP display shows End
Disconnect infusion line
Press and hold to stop beeping and retract plunger - display will show I----
Plunger retracts and beeps when it has reached the start position - display shows OFF
APo1 MODE APo3 MODE