Creative OXY-110 Manual de instrucciones

User Manual for Handheld Pulse Oximeter
I
Instructions to User
Dear Customer,
Thank you for purchasing this quality product. Please read the
manual very carefully before using this device. Failure to follow
these instructions can cause measuring abnormality or damage to
the Oximeter.
No part of this manual may be photocopied, reproduced or
translated into another language without the prior written consent.
We reserve the right to revise and amend it at any time without
prior notice.
Version of the Manual: Ver 1.0
Issued Date: October 28th, 2016
All rights reserved.
Notes:
The contents contained in this manual are subject to change
without prior notice.
Information furnished by Creative is believed to be accurate
and reliable. However, no responsibility is assumed by Creative
for its use, or any infringements of patents or other rights of
third parties that may result from its use.
3502-2290079

User Manual for Handheld Pulse Oximeter
II
Instructions for Safe Operation
Check the device to make sure that there is no visible
damage that may affect user’s safety and measurement
performance. It is recommended that the device should be
inspected minimally before each use. If there is obvious
damage, stop using the device.
Necessary service must be performed only by qualified
technicians. Users are not permitted to service this device.
The oximeter must not be used with the devices and
accessories not specified in User Manual.
Cautions
Explosive hazard—DO NOT use the oximeter in
environment with inflammable gas such as some ignitable
anesthetic agents.
DO NOT use the oximeter while the Patient is under MRI or
CT scanning. This device is NOT MRI Compatible.
Warnings
Discomfort or pain may occur if using the sensor of this
device continuously on the same location for a long time,
especially for the patients with poor microcirculation. It is
recommended that the Oximeter should not be applied to the
same location for longer than 2 hours or less if any
abnormal condition is found. Frequently check and
re-position the Oximeter sensor.
Misapplication of a SpO2probe with excessive pressure for
prolonged periods can induce pressure injury.
Place the SpO2probe on the finger tightly will cause venous
pulse and effect blood circulation, and lead to interstitial
edema, hypoxia and inaccurate measurement.
Biocompatibility tests have been performed on all the
applied parts, some exceptional allergic patients may still

User Manual for Handheld Pulse Oximeter
III
have anaphylaxis. Do not apply to those who have
anaphylaxis.
For the individual patients, there should be a more prudent
inspecting in the placing process. The sensor can not be
placed on the edema and tender tissue.
The local law should be followed when disposing of the
expired device or its accessories.
DO NOT operate in the environment where strong
electro-magnetic interference exists, such as radiogram,
television, radiophone, etc.
Please pay attention to the SpO2probe cable while using to
avoid strangulating patient.
Attentions
Keep the oximeter away from dust, vibration, corrosive
substances, explosive materials, high temperature and
moisture.
If the Oximeter gets wet, please stop operating it and do not
resume operation until it is dry and checked for correct
operation. When it is carried from a cold environment to a
warm and humid environment, please do not use it
immediately. Allow at least 15 minutes for the Oximeter to
reach ambient temperature.
DO NOT operate the button on the front panel with sharp
materials or sharp point.
DO NOT use high temperature or high pressure steam
disinfection on the oximeter and probes. Refer to related
chapter for instructions regarding cleaning and disinfection.
The intended use of this device is not for therapy purpose.
The equipment is IP22 with protection against harmful solid
foreign objects and ingress of liquid. So that means the

User Manual for Handheld Pulse Oximeter
IV
equipment is protected against solid foreign objects of
12.5mm and greater, and protected against vertically falling
water drops when enclosure tilted up to 15°.
Please pay attention to the effects of lint, dust, light
(including sunlight), etc.
Declaration of Conformity
The manufacturer hereby declares that this device complies with
the following standards:
IEC 60601-1:2005+A1: 2012, IEC60601-1-2:2014,
IEC60601-1-11:2010, ISO 80601-2-61:2011 and follows the
provisions of the council directive MDD93/42/EEC.

User Manual for Handheld Pulse Oximeter
V
Table of Contents
1 Overview...................................................................................... 1
1.1 Appearance............................................................................ 1
1.2 Product Name and Model...................................................... 3
1.3 Structure.................................................................................3
1.4 Features..................................................................................4
1.5 Intended Use.......................................................................... 4
1.6 Working Environment........................................................... 4
2 Preparation.................................................................................. 5
2.1 Charger Base Connection...................................................... 5
2.2 Power Supply.........................................................................5
3 Make Measurement.....................................................................7
3.1 SpO2Measurement................................................................ 7
3.2 Temperature Measurement (optional)....................................9
4 Operation................................................................................... 11
4.1 Power on/off the Oximeter.................................................. 11
4.2 Default Display Screen........................................................11
4.3 Menu....................................................................................14
4.4 Record..................................................................................27
5 Technical Specifications............................................................32
6 Over-limit Indication.............................................................. 34
6.1 Limit settings....................................................................... 34
6.2 Over-limit indication sound mute setting............................ 34
7 Packing List............................................................................... 35
8 Repair and Maintenance.......................................................... 36
8.1 Maintenance.........................................................................36
8.2 Cleaning and Disinfecting Instruction.................................36
9 Troubleshooting.........................................................................37
Appendix....................................................................................... 38
Quality Inspection Certificate..................................................... 45

User Manual for Handheld Pulse Oximeter
1
1 Overview
1.1 Appearance
Power/Left
Mute /Right
Display screen
Figure 1.1 Front View
Rotate-lock /Up
Setting/Down
Alert
indicator
Menu/Confirm
Record/Back
Power indicator

User Manual for Handheld Pulse Oximeter
2
1. Display screen: Display measurement result, trends and menus.
2. (Power/Left): Power on/off the device by longtime
pressing; On menu or sub-menu screen, short time press it to move
the cursor left or adjust the parameter values.
3. (Mute/Right): On data recall screen, longtime press this
key, then the delete dialog pops up; On measuring screen, short
time press it to resume or mute the alert sound. On non-alert status,
short time press it to turn on/off beep sound. On menu or sub-menu
screen, short time press it to move the cursor right or adjust the
parameter values.
Note: when the device alert indication is on, short time pressing
will mute the alert sound, the mute state will persist for about 90s.
After this mute period (90s), then the alert sound will resume.
4. (Rotate-lock/Up): On measuring screen, longtime pressing
to enable or disable the automatic screen orientation (on horizontal
or vertical direction); On menu or sub-menu screen, short time
press it to move the cursor upwards or adjust the parameter value.
Figure 1.2 Upper-side view
Figure 1.3 Bottom side view
SpO2:
SpO2 probe
connector
TEMP:
Temperature
probe
connector
DC power
input socket

User Manual for Handheld Pulse Oximeter
3
5. (Setting/Down): On measuring screen, longtime pressing
to enter into setting screen; On menu or sub-menu screen, short
time press it to move the cursor downwards or adjust the parameter
value.
6. (Menu/Confirm): Short time press it to enter into menu
screen, or to confirm the selection.
7. (Record/Back): Short time press it to enter into SpO2
record list screen, or to back to the previous level of menu.
8. (Alert indicator): If the probe is not well placed
or disconnected, or the measured value exceeds the preset alert
limit value, then the alert indicator will keep orange flashing.
9. (Power saving mode indicator): If the device is set as power
saving mode, then the indicator keeps on. And on measuring screen,
the indicator flashes with the pulse beep.
10. Icon: "SpO2"( ): SpO2Probe Connector.
11. Icon: "TEMP"( ): Temperature Probe Connector.
12. (DC power input): used for connecting
external DC power input for charging the built-in rechargeable
battery.
1.2 Product Name and Model
Name: Handheld Pulse Oximeter
Model: SP-20
1.3 Structure
It consists of the main unit and SpO2probe.
(Note: with optional temperature prob, this Oximeter can make
temperature measurement.)

User Manual for Handheld Pulse Oximeter
4
1.4 Features
It is lightweight, small in size and easy to carry
Color LCD to display plethysmogram and parameters
Measure SpO2, Pulse Rate and Temperature simultaneously
PI (Perfusion Index) display is available
Up to 500 hours data storage for SpO2and PR and can be
recalled.
16 user IDs for making data and can be added.
A built-on holder for convenient standing on desktop and
display viewing.
Real-time battery status display and low battery voltage
indication.
Auto power off is available
Audible and visual alert function is available
Data uploading to PC for management (Optional)
Power saving mode is available
1.5 Intended Use
This Handheld Pulse Oximeter is intended for measuring and
recording the pulse rate, functional oxygen saturation (SpO2) and
temperature (optional). It is applicable for detecting SpO2, pulse
rate and temperature of adult and pediatric patients in clinical
institutions and homes.
1.6 Working Environment
Operating temperature: 5~40℃
Operating humidity: 15%~93% (non-condensing)
Atmospheric pressure: 70kPa~106kPa

User Manual for Handheld Pulse Oximeter
5
2 Preparation
2.1 Charger Base Connection
Put the device into the charger
base properly, and note the
polarity markings, as shown in
figure 2.1.
Figure 2.1 Charger base connection
充电底座连接
Oximeter
Charger base
Tabla de contenidos
Otros manuales de Equipo médico de Creative

Creative
Creative PC 66B Manual de usuario

Creative
Creative POD-1 Manual de usuario

Creative
Creative PC-60N Manual de usuario

Creative
Creative PC-60NW-1 Manual de usuario

Creative
Creative PC-66A Manual de usuario

Creative
Creative PC 66B Manual de instrucciones

Creative
Creative PC-60NW Manual de usuario

Creative
Creative OXY-11 Manual de usuario

Creative
Creative PC-68A Manual de instrucciones

Creative
Creative PC-68B Manual de usuario