Buffalo filter BOWA Manual de usuario

Operator’s Manual
BOWA- electronic
Smoke Evacuation System
with RFID Technology

Limited Warranty
For a period of one (1) year following the date of delivery BOWA electronic warrants the
BOWA Smoke Evacuation System against any defects in material or workmanship. BOWA
electronic will repair or replace (at BOWA electronic’s option) the same without charge,
provided that routine maintenance as specified in this manual has been performed using
replacement parts approved by BOWA. This warranty is void if the product is used in a
manner or for purposes other than intended.
© 20 BOWA electronic GmbH & Co. KG
Heinrich-Hertz-Strasse 4-10
72810 Gomaringen/Germany
www.bowa-medical.com
TELEFON: +49 (0) 7072-6002-0
Fax: +49 (0) 7072-6002-33
info@bowa-medical.com
The revision level of this manual is specified by the highest revision letter found on either the
inside front cover or enclosed errata pages (if any).
Manual Number: 5(9$0$18$/%2:$23SGI
Manufactured for BOWA electronic GmbH & Co. by:
Buffalo Filter
59*HQHVHH6 WUHW
/DQFDVWHU1< 86 $
Classified by Underwriters Laboratories, Inc. with respect to electric shock, fire and
mechanical hazards only in accordance with UL2601-1
9D93
9D93
2

3
Table of Contents
Page
1.0 Introduction................................................................................................... 1
2.0 Inspection...................................................................................................... 1
3.0 Operational Information............................................................................... 1
4.0 Warnings and Cautions................................................................................ 1- 2
5.0 Operating Instructions..................................................................................2-4
• Key Pad Diagram
• Power ON/ OFF
• Motor ON/ OFF
• Suction Control
• TURBO Mode
• Filter Life Remaining
• Footswitch
• Remote Switch Activator (RSA)
• Circuit Fuses
• Service
6.0 Maintenance.................................................................................................. 5
• Cleaning
• Periodic Inspection
• Troubleshooting
7.0 Specifications................................................................................................6
8.0 Ordering Information.................................................................................... 6
9.0 Equipment Return......................................................................................... 7
10.0 Terms.............................................................................................................8
• Confidential Information
• Consequential Damages/ Limits of Liability
• Entire Agreement
• Jurisdiction
• Shipment of Order
11.0 Warranty........................................................................................................ 8

The BOWA Smoke Evacuation System is designed to remove smoke and aerosols and
noxious odors produced during, but not limited to, laser and electrosurgical procedures.
The BOWA Smoke Evacuation System has been designed with a high suction, high flow rate
centrifugal action pump. The ultra quiet motor is used to draw the smoke plume from the
surgical site through the vacuum tubing/hose into the BOWA
filter where the surgical smoke
is processed by 4 stages of filtration. Built into a single housing, this disposable filter is used
to simplify the installation and removal during filter changes. The filter is completely enclosed
to protect the healthcare personnel from potential contamination during filter changes and is
totally filter life automated via RFID technology.
The BOWA Smoke Evacuation System has been thoroughly tested and inspected before
shipment from the factory. Please check the unit before setup to insure that no damage has
occurred in transit. If damage is evident, please contact BOWA customer service at
+49 (0) 7072-6002-0
The BOWA Smoke Evacuation System complies wi th UL/IEC 60601.1 elect rical standard in
the following systems:
• 220v- 240v 50/60 Hz
Please note that all Cautions and Warnings should be read and understood before any
use of this equipment. This smoke evacuation system should only be operated by
personnel who have been properly trained in its operation.
Warnings:
• The BOWA Smoke Evacuation Filters are completely disposable. Please
handle these disposable filters according to your local codes, regulations,
and institution policy.
• This device is not intended for evacuation of fluid. If fluid is expected to be
aspirated to the BOWA Filter, fluid collection devices must be installed with the
vacuum hose assembly. Failure to install a fluid collection device could cause
filter blockage and electrical damage.
4

• DO NOT OPERATE THIS DEVICE IN THE PRESENCE OF EXPLOSIVE GASES.
EQUIPMENT NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE
ANESTHETIC MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE.
• REFER ROUTINE SERVICING TO QUALIFIED
BIOMEDICAL TECHNICAL PERSONNEL.
• There are no user serviceable components in the BOWA
Smoke Evacuation System.
• Use only with the power cord provided and always plug
into a grounded outlet.
SYMBOL DESCRIPTION/MEANING
TYPE CF Equipment Class I
Medical Equipment with respect to
electrical shock,fire and mechanical
hazards only. In accordance with UL
60601
Cautions:
• Do not block either the tubing or the filter. If either becomes occluded or
significantly restricted, the motor/blower may overheat and cause the unit
to become inoperative.
• Attempting to use any other filter or accessory not supplied by BOWA may
cause damage and/or cause the system to be inoperable and will void the
warranty.
• Care must be exercised in the installation of hoses, adapters, suction
canisters.
Failure to follow the procedures outlined in this m anual may result in over heating of
the motor and may void the unit warranty.
5

The BOWA Smoke Evacuation LCD control display is user-friendly and completely
interactive, allowing for easy to understand use (as displayed on the LCD screen).
Key Pad Diagram
FIGURE 1
The following items are important for the operation of your system.
POWER ON/OFF
The main power switch is located on the rear panel of your smoke evacuator. This switch
completely enables and disables the main power.
When the unit is powered on the
following LCD screen will appear.
After inserting a filter a screen will
flash for a few seconds to indicate
filter type and original life.
Next, the standard screen will
appear which displays
information regarding the unit’s
current activity.
BOWA EL ECT RONI C
INSERT BOWA
FI LTER
BOWA ELECTRON I C
BOWA FI LTER
8 HOUR FI LTER
BOWA FI LTER HR: M
N
T i me R e ma i n i n g 7 : 3 1
Suct i on Set t i ng 100
%
Mo de - TURBO Mo t o r - ON
6

MOTOR ON/OFF
The Motor ON/OFF button is located on the front panel (Figure 1). This button turns the
motor (for suction) on and off.
SUCTION CONTROL
The amount of suction may be adjusted using the UP or DOWN arrow buttons located on
the key pad (Figure 1). Suction may be adjusted up or down by 10% increments. It is
recommended that you select the lowest suction setting that effectively clears the
operative field of surgical smoke.
TURBO MODE
The TURBO button is located on the key pad (Figure 1). The amount of suction is
significantly increased when this button is pressed. To return to NORMAL mode, press the
TURBO button again.
FILTER LIFE REMAINING
The exact number of hours and minutes remaining on the BOWA filter (in use) are displayed
on the LCD screen. When the filter’s life is consumed, time remaining will indicate that there
are 0 hours and 0 minutes remaining. This line will continually flash. When the filter life
reaches this point, a new filter with time remaining should be inserted; however, the suction
will still work. When the unit’s main power is turned off or the system is left idle for more than
6 hours, a new filter or a filter with remaining filter life MUST be inserted into the filter basket
to allow the unit’s suction to turn back on.
When the filter’s life is
exhausted the LCD
screen will flash
“REPLACE FILTER.”
BOWA F I L TER HR : MN
REPLACE F I LTER 0 0 : 0 0
Suct ion Sett ing 100%
Mod e - NORML Mo t o r - ON
7

FOOTSWITCH
A pneumatic footswitc h is provided. To activate the foots witch, simply plug the footswitch
tubing into the jack ( ) on the rear panel. The suction can now be switched ON/OFF by
depressing the footswitch.
REMOTE SWITCH ACTIVATOR
For directions on using the Re mote Switch Activator, see instructions ac companying the
device.
CIRCUIT FUSES
Two 8 amp fuses are located on the rear panel wi thin the power entry module. These circuit
fuses protect both the system and the operator from damage or injury.
SERVICE
When a service message appears, please c ontact BOWA Customer Service for instructions.
Below are some examples of service screens that may appear.
If this screen appears the system may be in need of
service for the motor. Call BOWA.
If this screen appears there may be a problem with your
system or filter. Call BOWA.
Brush Hour s= 500
IF OVER 500HRS CALL:
+49( 0) 7072- 6002- 0
F I LTER I . D . ERROR !
CALL FOR SERV I CE
+49( 0) 7072- 6002- 0
.
B r u s
SETTING THE DEFAULT LANGUAGE
1 Turn the main power on without filter inserted.
2 Scroll through languages using suction control arrow buttons and select a language by
pressing motor button.
3 Smoke Evacuator will use this language as the default language.
4 To change the default language, remove filter and press motor on and turbo button
simultaneously.
5 Repeat step 2.
8

CLEANING
The exterior of the system may be cleaned by unplugging the unit, wiping it with a cloth
that has been dampened (not dripping) with a mild detergent or surface disinfectant.
PERIODIC INSPECTION
The BOWA Smoke Evacuation System should be visually inspected at least once a year.
This inspection should include:
• Inspecting for damage to power cord
• Inspecting for damage to power plug
• Checking the tightness of power plug
• Ensuring there is proper mating, cleanliness, and absence of damage to the filter inlet
• Ensuring there isn’t any obvious external or internal damage to the system
TROUBLESHOOTING THE SYSTEM
The following chart can be used to help troubleshoot typical problems.
PROBLEM POTENTIAL CAUSE CORRECTIVE ACTION
1. System does
not appear to
have any power
(no lights).
1. Not plugged into electrical
outlet or power entry
module.
2. Fuses are blown.
3. Main power switch is in
OFF position.
4. Electronic system failure.
1. Check power outlet and ensure
connection to rear panel of machine.
2. Replace fuses located within the power
entry module.
3. Check main power switch on rear panel
of system to ensure that it is in the ON
( I ) position.
4. Call BOWA Customer Service at:
+49 (0) 7072-6002-0
2. System does
not function even
though suction is
ON.
1. Filter life is expired.
2. An invalid filter has been
inserted into the basket.
3. A filter has not been
inserted into the filter
basket.
1. Replace filter. Insert a BOWA RFID Filter
3. System is ON
but suction is not
working or is
extremely
minimal.
1. Filter is not completely
seated in the basket.
2. Filter is clogged.
3. Vacuum hose is clogged or
“kinked.”
4. Motor/ blower is obstructed.
1. Re-install BOWA Filter. Press firmly in
place within the filter basket.
2. Replace filter. Ensure that a BOWA filter
and a pre-filter are used for Erbium:YAG
laser procedures.
3. Check or replace smoke evacuation
tubing.
4. Call BOWA Customer Service at:
+49 (0) 7072-6002-0
9

The following table outlines the specifications for the BOWA Smoke Evacuation System.
Maximum Flow Setting
Tubing 7/8” (22 mm) Actual Flow Rates:
Normal: 25 CFM (708 lpm)
Turbo: 35 CFM (991 lpm)
Variable Flow Control YES
Continuous
OperationMotor
1 HP (0.75 kw)
Static Suction 85 inches H O (211 mbar)2
Noise Level (dBA) 65dB maximum
Footswitch YES- pneumatic
Certifications UL, cUL, CE
Display LCD
Voltages 220v- 240v, 50/60Hz
Dimensions 43.18cm W x 19.1cm H x 41.80 cm D
Weight 10.7 kg
To re-order disposables (filters, tubes, hoses), obtain replacement parts, or to return a unit
for service, call BOWA Customer Service at: +49 (0) 7072-6002-0, or contact your authorized
BOWA Distributor/Representative.
10
Tabla de contenidos
Manuales populares de Sistema de seguridad de otras marcas

EDM
EDM Solution 6+6 Wireless-AE Manual de usuario

Highway Safety Group
Highway Safety Group EA401 Manual de usuario

Siren
Siren LED GSM Manual de usuario

Detection Systems
Detection Systems 7090i Instrucciones de montaje

Se-Kure Controls
Se-Kure Controls MicroMini SK-4841 Manual de usuario

Siemens
Siemens FDM273 Manual de usuario