PHOTODYNAMIC THERAPY
HELBO®User Manual for HELBO® TheraLite Laser
5
Version 14 06/2017
1.3. CLINICAL PROCEDURE (in red)
Control of relevant process parameters with the HELBO® T-Controller
Brief instructions Activity
(at device)
Display Acoustic
signal
Therapy stage Next stage
Initial situation:
Rinse mouth 2x
None None None Rinse patient’s mouth twice for
approx. 20 sec. with sufficient
water / remove saliva and blood
from the areas to be treated
Switch main switch to „on“
Main switch in “on“
position
Switch main
switch to „on“
Tooth fields 1-8 flash (red) None Select corresponding tooth field
according to the number of teeth
to be treated
Press corresponding tooth
field according to the num-
ber of teeth to be treated
Pressing the correct
tooth field, application
of phenothiazine
chloride dye
Press corre-
sponding tooth
field according
to the number
of teeth to be
treated
Tooth fields according to the
number of teeth to be treated
are permanently lit
None Dye application and distribution
(if necessary) using the air
blower on the multi-functional
syringe to ensure thin, uniform
coverage of the entire surface
Monitoring dye exposure
time by pressing the Go/
Break key
Pressing the Go/Break
key. Phenothiazine
chloride dye exposure
time: 60 -180 sec.
Press Go/Break
button
Tooth fields according to the
number of teeth to be treated
are permanently lit / “photosen-
sitizer” display (blue) is perma-
nently lit for 60 sec. and flashes
slowly for another 120 sec.
Beep Dye exposure (60-180 sec.) Rinsing using the multi-
functional syringe and
suctioning off excess dye
Rinsing and suction of
excess phenothiazine
chloride dye
None “Photosensitizer” display (blue)
flashes rapidly
None Rinsing using the multi-function-
al syringe (including in the
pocket!) and suctioning off
excess dye
Press Go/Break button once
more; exposure of first tooth
with the HELBO® TheraLite
Laser for 60 sec.
Press Go/Break button
once more, exposure
with HELBO® TheraLite
Laser: 60 sec. per tooth/
per cm2
Press Go/Break
button
Tooth field “1” flashes, tooth
fields of the teeth not yet treat-
ed are permanently lit, 60 sec.
display indicates exposure time
in 10 sec. intervals
Beep Exposure of first tooth with
the HELBO® TheraLite Laser for
60 sec.
Exposure of the remaining
teeth with the HELBO®
TheraLite Laser for 60 sec.
each
Exposure of the remain-
ing teeth: 60 sec. per
tooth/per cm2
None Tooth field “2” (3-8) flashes,
tooth fields still requiring treat-
ment are permanently lit, 60
sec. display indicates exposure
time in 10 sec. intervals, tooth
fields of teeth already treated
are not lit
Beep
(every
10 sec.
after each
additional
tooth)
Exposure of the remaining teeth
with the HELBO® TheraLite Laser
for 60 sec. per tooth
End of therapy after last
sound is heard (all tooth
fields 1-8 flash)
Conclusion of therapy:
Rinse
Turn off device
by pressing
„on/off“ button
None None Thoroughly rinse patient’s mouth None
Other func-
tions:
Press Go/Break
button during
treatment
Tooth fields still to be treated
are permanently lit / 60 sec.
display flashes
None
Press RESET
button during
treatment
Starting point: All tooth fields
flash
„Check Bat-
tery“ warning
light
Lights up early when battery
power is low => change battery
None