
Pre-Humidifier 3110
5 Pre-Humidifier Operation
5.1 Operation Overview
Two manual needle valves are provided on the front panel to control the dry and wet air flow rates that
mix together to create the desired sample flow RH. The wet flow is created using a 36” long nafion tube
submerged in a water bath maintained at 40◦C . Dry compressed air enters the nafion tube in the water
bath and becomes saturated with water vapor.
When attempting high RH operation, be sure to tightly close the Dry Flow valve to limit the amount of dry
flow supplied to the sample humidifier.
The saturated air flow from the wet bath is mixed with a dry air flow and the mixed flow is delivered to a
second box housing another 24” coil of nafion. The particle sample flow is passed through the 24” long
nafion. By manually adjusting the ratio of dry and wet air flow rates, different RH values can be attained.
Because of the nature of the nafion tubing, the system is relatively slow response, be sure to wait several
minutes after making changes to the air flow rates to determine the change in RH. An RH sensor monitors
the RH of the sample flow just before the exit port on the front panel. The RH sensor reading is displayed
on the front panel.
Warning: Operating the Pre-Humidifier at RH above 80% over long periods of time (days) can possibly
result in water condensation inside the sample flow nafion box. To operate under dry conditions, dry air flow
may be required for a substantial amount of time to evaporate any liquid condensate.
Figure 5: Front view of the Pre-Humidifier chassis.
Connect your aerosol instrument to the sample flow out port of the Pre-Humidifier . In the case of the
Brechtel Model 3100 HTDMA, the Pre-Humidifier is installed between the SEMS and HSEMS so that the
dry monodisperse flow from the SEMS, after the MCPC flow splitter, is attached to the Pre-Humidifier
sample flow in port. The sample flow out port of the Pre-Humidifier is connected to the HSEMS sample
flow in port. The RH setting of the Pre-Humidifier can be manually entered into the HSEMS AutoScheduler
software. Adding the Pre-Humidifier to the HTDMA allows particle efflorescence studies to be conducted by
exposing the dry SEMS-selected particles to a high RH in the Pre-Humidifier and then exposing the sample
flow to a lower RH in the HSEMS.
83-00029-01 Page 7 of 9