Blackrock Microsystems CerePlex E Series Manual de usuario

Manufacturer
630 Komas Drive | Suite 200
Salt Lake City | UT 84108 | USA
P +1 (801) 582-5533 | F +1 (801) 582-1509
www.blackrockmicro.com
Revision 4.00 / LB-0545 – CerePlex E IFU – 2020/03
© 2020 Blackrock Microsystems, LLC
CerePlex E
PN: 7513, 10511 – E96
PN: 8010, 10510 – E128
PN: 10747 – E256
Instructions for Use

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
2
Table of Contents
What this Manual Covers ...................................................... 4
Intended Use and Indications for Use ................................... 4
Warnings and Cautions ......................................................... 4
Warnings ........................................................................................ 4
Precautions ..................................................................................... 5
Specifications ........................................................................ 6
Hardware ............................................................................... 7
Blackrock CerePlex E Setup ........................................................... 7
Instructions for Assembly ................................................................ 8
Description of Switches on the CerePlex E ..................................... 9
Impedance and Recording selector switch ................................... 9
Frequency Range Selector Wide & Norm ..................................... 9
Reference and Ground Selector ................................................. 10
R1 and R2 Reference selector ................................................... 10
Cleaning & Maintenance Instructions .................................. 10
Sterilization ................................................................................... 10
Cleaning and Maintenance ........................................................... 10
Filament Film Disassembly and Replacement .............................. 11
Disposal ........................................................................................ 12
Magnetic Resonance .......................................................... 12
Troubleshooting .................................................................. 12
Non-Green Indicator Light on CerePlex E ..................................... 12
Poor Signal to Noise Ratio / Noisy Signal ..................................... 13
Non-Green Light on Digital Hub Bank ........................................... 13
Return Merchandise Authorization ...................................... 13
Warranty .............................................................................. 14

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
3
Support ................................................................................ 14
Complaints .................................................................................... 14

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
4
What this Manual Covers
The CerePlex E is a Digital Headstage Accessory that provides a compact interface from
a Blackrock Utah Array to the Cerebus or CerePlex Direct data acquisition systems. The
CerePlex E converts analog signals to digital format at the recording site, which can
reduce noise introduced into the signal during transmission to the recording system.
This Instructions for Use will cover important information about the device including
warnings, hardware settings, and maintenance activities.
There are several CerePlex E models that are available in different channel counts.
These are outlined in the table below.
Intended Use and Indications for Use
The CerePlex E is intended to be used with the CerePort Electrode or CerePort
Electrode pedestal connector with third party electrodes.
The CerePlex E supports processing and display of biopotential signals from user
supplied electrodes. Biopotential signals include but are not limited to:
Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG),
electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP).
Warnings and Cautions
Warnings
Only connect Cerebus System components to properly tested, grounded
and dedicated AC outlets to reduce the risk of electrical shock. Do not
use an adapter for ungrounded wall outlets.
Do not connect the Cerebus System to an outlet controlled by a wall
switch.
Do not use the CerePlex E in the presence of flammable anesthetic
agents.
Avoid strong static discharges from sources like televisions or computer
monitors because it can damage the electrical parts of the system.
Model Name
CerePlex E96
v1
CerePlex E128
v1
CerePlex E96
v2
CerePlex E128
v2
CerePlex E256
Model Number
7513 8010 10511 10510 10747
Channel Count
96 128 96 128 256

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
5
Keep the CerePlex E away from liquids. Contact with water, shower
spray. or wet surfaces can lead to the subject receiving an electrical
shock.
Connection of external instruments may compromise electrical safety
compliance with IEC 60601-1.
The CerePlex E digital data cable is long and may snag on equipment in
the room. Secure equipment in place and avoid placing in walkways or
other traffic areas.
The CerePlex E is attached to the CerePort Electrode percutaneous
connector, force applied to the CerePlex E will be applied to this
connector. Avoid large impacts to the device when it is attached to the
subject.
Using the ‘Wide-Band’ filter mode on the device may cause a reduced
signal to noise ratio.
Precautions
Read this entire manual prior to using the device.
Do not allow the digital data cable to be pulled and apply force to the
CerePlex E.
Do not allow any object to impact the CerePlex E while in use.
Do not drop the CerePlex E when handling.
Do not use the CerePlex E with damaged data cables or data cables that
are not supplied by Blackrock, doing so may damage the device.
When replacing the filament film of the device, avoid touching circuit
board components as these can be easily broken and cause damage to
the device.
Do not attempt to use the CerePlex E with connectors other than the
CerePort pedestal connector.
Do not plug the digital data cable into devices other than the Digital Hub
and Blackrock approved headstage accessories.
Avoid spilling fluid onto or into the CerePlex E as doing so may cause
electrical damage to the device and loss of function or electrical shock.

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
6
Specifications
Model Name
CerePlex E
96
CerePlex E128
CerePlex
E256
Model Number
7513/10511 8010/10510 10757
Number of Channels
96 128 256
Power
Requirements
0-5 VDC, 150 mA maximum load 0-5 VDC, 300 mA maximum load
Resolution
16-bit, 250 nV per bit
Sampling Frequency
30,000 Samples per Second
Mode of Operation
Continuous
Input Hardware
Filter
Normal Mode:
0.3 Hz (1st Order) – 7.5 kHz (3rd Order)
Wide-Band Mode:
0.02 Hz (1st Order) – 10 kHz (3rd Order)
Input Filter Type
Butterworth
Input Impedance
1300 MΩ @ 10 Hz, 13 MΩ @ 1 kHz
Imp
edance
Measurement Range
5 kΩ – 1 MΩ
Maximum Input
Range
+/- 8.192 mV with Respect to Reference
Dimensions
L 24.9 mm x W 17.7 mm x H 17.9 mm
Operating
Environment
10˚ C to 40˚ C, 5 to 95% R.H. (non-condensing)
Storage
Environment
-20˚ C to 50˚ C, 5 to 100% R.H. (non-condensing)

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
7
Hardware
Blackrock CerePlex E Setup
The image below demonstrates how the CerePlex E connects with the Digital
Cerebus System or the CerePlex Direct in a lower channel configuration.
Figure 1 - System Diagram of CerePlex E Connected to the Cerebus System or the CerePlex Direct. The
E256 requires two digital hubs.

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
8
Instructions for Assembly
The CerePlex E is designed to mate with the CerePort connector attached to
Blackrock Utah Arrays. The CerePlex E is equipped with three guide posts.
These guide posts must be aligned with the three available notches around the
edge of the CerePort connector to facilitate a correct connection between
CerePlex E and the CerePort array. For a detailed procedure on how to connect
the CerePlex E headstage to CerePort connector please refer to the following
instructions:
Connect the CerePlex E to the host device through an HDMI A-D cable. HDMI
Connector A is associated with banks A, B, C, and D of the LGA connector and
HDMI Connector B is associated with banks E, F, G, and H of the LGA
connector. The other end of the HDMI cable should be connected to the Digital
Hub or CerePlex Direct, with cable A occupying a lower slot and cable B
occupying a higher slot.
Power on the connected host device.
The LED on the CerePlex E should light up as blue.
Roughly align the guide pins with the pedestal before attaching to the pedestal.
Figure 3 - Connect CerePlex E to Pedestal
Hold the CerePlex E housing with one hand and gently rotate the wheel
clockwise with another hand.
While screwing down, slightly rotate the housing left or right until you feel the
guide posts slide into the notches.
Figure 2 – Insert HDMI A cable into CerePlex E

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
9
Once the CerePlex E is aligned with the pedestal, rotate the wheel until the
device is secured to the pedestal.
Figure 4 - Rotate wheel clockwise to connect CerePlex E to Pedestal
The blue LED will turn green when the device is connected correctly.
The instructions above are also detailed in one of the Blackrock training videos
which can be accessed online from the Blackrock Microsystems website under
How-to Videos.
Description of Switches on the CerePlex E
Figure 5 - Available switches on the CerePlex E96, E128, and E256
Impedance and Recording selector switch
When the selector is positioned on “Imp” the CerePlex E enters
impedance measurement mode. This mode enables users to measure
impedance of electrodes via Blackrock’s Central Suite. For information
regarding impedance measurement please refer to the document titled
“LB 0574 Central IFU,” which is available for download from the Blackrock
Microsystems website. To revert the CerePlex E headstage into the
recording mode, simply toggle the switch to “Rec” position (recording).
Frequency Range Selector Wide & Norm
This switch toggles between two different hardware-based filter input
settings. The “Wide” filter setting corresponds to an input frequency range

Revision 4.00 / LB-0545 – CerePlex E IFU
© 2020 Blackrock Microsystems, LLC
10
from 0.02 Hz to 10 kHz. “Norm” (Normal) filter setting is between 0.3 Hz
to 7.5 kHz.
Reference and Ground Selector
This switch allows the user to select between different sources to use as
the input reference. If the switch is toggled to “Gnd” mode, subject ground
will be used as the input reference to the amplifier. Setting the switch
position on “Ref” (Reference), will use one of the implanted reference
wires (R1 or R2) as input reference. R1 and R2 can be selected by
another switch adjacent to reference and ground selector (see below).
R1 and R2 Reference selector
Use this switch to select which of the two available reference wires to use
as the input reference. A switch position to R1 selects reference wire
number 1 as the input reference. The R2 switch selects the reference
wire number 2 as shown in Figure 5. The Ref/Gnd switch must be in Ref
mode for this switch to be effective.
Cleaning & Maintenance Instructions
Sterilization
While not generally required, the device may be sterilized by both Vaporized
Hydrogen Peroxide and Ethylene Oxide Gas. Note that Vaporized Hydrogen
Peroxide may significantly discolor the case of the device. Newer models have a
new case that is immune to this discoloration.
Cleaning and Maintenance
The CerePlex E case may be wiped down with standard sanitizing wipes,
isopropyl alcohol (70%) or ethanol (96%). If necessary, the surfaces and
connection ports may be wiped with enzymatic detergent. Dust and debris on the
exterior connectors/ports may be removed with compressed air.
The collar assembly underneath the CerePlex E should be inspected before each
use and cleaned every time after use. To clean the collar and filament film, follow
the instructions below:
Visually inspect the connector; remove any particles if present.
Visually inspect the film. If the film is swollen or degraded, follow the
section titled “Filament Film Replacement” (see below) to replace the film.
If necessary, clean the connector with a cotton swab and 70% isopropyl
alcohol. Do not let the cleaning solution pool. Allow the solution to dry
before use.
The connector and the film must also be cleaned after use with a cotton
swab and 70% isopropyl alcohol. Do not allow the solution to pool. Allow
the solution to dry before storage.
Otros manuales para CerePlex E Series
1
Este manual sirve para los siguientes modelos
10
Tabla de contenidos
Otros manuales de Unidad de control de Blackrock Microsystems
Manuales populares de Unidad de control de otras marcas

Festo
Festo Compact Performance CP-FB6-E Manual de lista de piezas

Elo TouchSystems
Elo TouchSystems DMS-SA19P-EXTME Manual de usuario

JS Automation
JS Automation MPC3034A Manual de usuario

JAUDT
JAUDT SW GII 6406 Series Guía rápida

Spektrum
Spektrum Air Module System Manual de usuario

BOC Edwards
BOC Edwards Q Series Manual de usuario