9
Caution! For model PST-60HL: Load only pairs of microplates for best
fixation.
Correct loading
Wrong loading
Caution! Do not fill in the microplates after they have been inserted in
the unit.
Caution! For model PST-60HL-4: The microplate fixation screws must
always be tightened to avoid damage. Tighten the fixation
screws completely when microplates are removed from or
placed on the platform. Do not close the lid if the microplate
fixation screws are not tightened to avoid damage.
4.4.2. Press the RPM-RUN/STOP key (Fig. 1/4). The platform will start rotating and
the timer indicator will start counting up the time interval (with 1 min precision).
4.4.3. After finishing the program, the platform motion will stop and the timer will
show the flashing reading STOP accompanied by the repetitive sound signal
until the RPM-RUN/STOP key is pressed.
4.5. If the working time is not set (or is reset) and the timer indicator in the upper line
shows 00:00, pressing the RPM-RUN/STOP key will start continuous operation
of the Thermo-shaker (timer indicator will start counting up the time interval in
the lower line (Actual)) until the RPM-RUN/STOP key is pressed again.
Please note! The platform temperature will be constantly maintained in
accordance with the set temperature. This allows using the
device again without pre-heating.
4.6. The timer can be reset during operation if required. Press the TIME-
RUN/STOP key once (Fig. 1/5) to stop the timer. Press the TIME-RUN/STOP
key again to restart the timer.
4.7. The platform motion can be stopped at any time by pressing the RPM-RUN
/STOP key. In this case the program realisation will stop and the timer will
switch into the STOP mode saving previously set time. Press the RPM-
RUN/STOP key to repeat the operation with the same time and speed.
Caution! At the end of the set time period the platform movement is
stopped automatically, but the heating can be stopped only by
, lower
button) till the OFF sign appears in the upper part of the display.
4.8. After finishing the operation set the power switch, located on the rear panel of
the unit, in position O (Off) and disconnect the external power supply from
electric circuit.