ATORN DFT Series Manual de usuario

Operators manual
DFT
Gages for coating thickness measurements

ATORN DFT FE and ATORN DFT FE+NF
Gages for coating thickness measurements
© 2019 by ATORN Service, Ludwigsburg, Germany
This document is protected by copyright. All rights reserved. This document
may not be reproduced by any means (print, photocopy, microfilm or any oth-
er method) in full or in part, or processed, multiplied or distributed to third par-
ties by electronic means without the written consent of ATORN Service.
Subject to correction and technical changes.
Note: Designations not marked with ® or ™ may also be protected by law.
Document no. 1002229
Issue date 2019-11-22 (for software version 1.0.2)
Distribution Further sources of supply
ATORN Service www.atorn.de
Schliefenstraße 40
71636 Ludwigsburg
Germany

ATORN DFT i
1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Environmental conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Safety of the electrical equipment . . . . . . . . . . . . . . . . . . . . 2
2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1 Measurement reading view (example) . . . . . . . . . . . . . . . . . 3
2.2 Gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4 Menus - function overview . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.5 Gage concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.6 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1 Installing batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2 Switching on the gage . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3 Switching off the gage . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4 Getting started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5 Settings for measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1 Creating a new batch . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.2 Opening a batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.3 Assigning a different calibration to the opened batch . . . . . 22
6 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.1 Notes on coating thickness measurement . . . . . . . . . . . . . . 23
6.2 Before you start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.3 Measuring - Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.1 Calibration - When necessary?. . . . . . . . . . . . . . . . . . . . . 26
7.2 Selecting the appropriate calibration method . . . . . . . . . . . 26
7.3 Performing a calibration . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.4 Calibration- Assigning/changing names . . . . . . . . . . . . . . 35
7.5 Calibration - Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8 Data transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.1 Transfer batch files in the PHASCOPE PAINT app. . . . . . . . 36
8.2 Transfer single readings online to an Excel file via PC-Datex 38

ii ATORN DFT
9 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
9.1 Glossary - Display symbols . . . . . . . . . . . . . . . . . . . . . . . . 43
9.2 Glossary - Display texts. . . . . . . . . . . . . . . . . . . . . . . . . . . 44
10 About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

ATORN DFT 1
Safety information
1 Safety information
If you use the instrument as intended and observe the safety information, the
instrument poses no danger.
Please read and follow this Operator's Manual and observe the safety infor-
mation. Also observe generally applicable safety and accident prevention reg-
ulations.
1.1 Intended use
The gage is intended solely for measurement of coating thicknesses.
Only accessories approved or recommended by ATORN Service may be con-
nected to the gage.
Any use beyond this is not the intended use. The risk of damage ensuing there-
from is borne solely by the user.
1.2 Environmental conditions
Storage and transport temperature: 0 … +50 °C
Temperature
When exposed to sunshine, the areas behind glass windows (e.g. in an
automobile) can easily reach temperatures in excess of +50 °C. This can
cause damage to the instrument.
►Do not keep or store the instrument and accessories behind glass
windows, or near to sources of heat such as radiators etc.!
Acid
The gage and accessories are not acid-proof.
►Do not place the instrument or accessories in contact with acids or liquids
which contain acid!

2 ATORN DFT
Safety information
Potentially explosive environment
The gage and accessories are not suitable for use in potentially posted
environments.
►Operate the gage and accessories only outside of potentially explosive
areas!
1.3 Safety of the electrical equipment
Only accessories approved or recommended by ATORN Service may be con-
nected to the gage!
USB cable
Damaged USB cable
Kinking or pinching the USB cable can result in a broken wire. Data trans-
mission is then no longer possible.
►Connect only an undamaged USB cable with a max. length of 3 m to the
gage.
►Always coil up the USB cable for storage.
Batteries/rechargeable batteries
•Use the following alkaline or lithium battery type: Mignon, 1.5 V,
LR6 - AA
or
•Use the following NiMH rechargeable battery type: Mignon, 1.2 V,
HR6 - AA
Servicing and repairs
Modifications, repairs as well as maintenance and service work on the instru-
ment and accessories may be carried out only by service personnel authorized
by ATORN Service.
Exception: Changing the batteries/rechargeable batteries.

ATORN DFT 3
Description
2 Description
The gages in the ATORN DFT series
measure coating thicknesses easily,
quickly and in a nondestructive man-
ner. Gage construction with an inte-
grated measuring probe allows sin-
gle-hand measurements, preferably
on flat and cylindrical samples.
This operator's manual describes the
following gage types in the ATORN
DFT series:
•ATORN DFT FE
Gage type for measurements on fer-
rous base materials (Fe) only
•ATORN DFT FE+NF
Gage type for measurements on
both as measurements on ferrous
(FE) as well as on non-ferrous metal
base materials (NF)
Additional information can be found
in the data sheet, page 9
2.1 Measurement reading view
(example)
1 Measuring application
NF/FE: Nonferrous coating material on
ferrous base material
NC/NF: Electrically nonconducting and
nonferrous coating material on electri-
cally conducting nonferrous metals
2 Batch name (batch type abbrev.)
3 Tolerance limits
4 Measurement reading
5 Statistics values
6 Key assignment line (example:
delete symbol, open block statis-
tics, symbol for opening calibra-
tion)
7 Battery indicator
8 Unit of measurement
A list of all display symbols and texts
can be found starting on page 43.
1
2
3
4
3
5
6
NF/FE
Batch 01 (IDV)
No 1 n 10
X 46.8 s 1.81
Min 44.9 Max 48.5
BLOCK
47.3 µm
44.5
49.0
Example of the measurement
reading view showing statis-
tics with set tolerance limits
7
8

4 ATORN DFT
Description
2.2 Gage
1 Eyes for a carrying strap
2 Signal lamp to indicate measurement acquisition and limit violations
3 Display
4 Keys, On/Off key, for description see page 5
5 Positioning support for reliable placement of the gage on the surface
6 Probe
7 USB port
8 Battery compartment cover
Gage dimensions can be found in the data sheet, page 14
Front of gage
1
2
3
4
5
6
7
8
Rear side of gage

ATORN DFT 5
Description
2.3 Keys
There are 4 keys for operating the gage.
The bottom line of the display always shows the functions of the 4 keys (see
illustration below). The assignment depends on the opened menu page.
The function shown on the display is assigned to the key directly underneath (example).
The key has two functions:
•It switches the gage on and off when the key is held for at least 1 s
•Function shown in the bottom line of the display
The key has two functions:
•It moves the cursor/mark down when is shown in the bottom line of
the display
•It opens the main menu when is not shown in the bottom line of the
display
A description of the remaining key symbols can be found starting at page 43
SEL

6 ATORN DFT
Description
2.4 Menus - function overview
Main menu for the "Industrial" batch template package (Standard)
Statistics > Display of the statistics for the opened batch
Batch
Modify >
Settings and en-
tries for the
opened batch
MEASURE > Switch to the measurement view
Tolerance Limits > Activation of limit monitoring and
setting of the limits
Batch Info > Input of information
View Settings > Measurement view
Batch - statistics display
Block - statistics display
Units > Set the unit of measurement
Block Size > Activation of automatic block for-
mation and setting of the block
size
Offset > Set the offset value
More Settings > MEASURE
Counted Average
Measurement Mode
(Single Reading, Free Running)
Measurement acquisition
Value Resolution
Save Measurements
Air Reference
Calibration assign
Batches > New >
List of saved batches
(After a batch has been selected, it can be opened, copied
or deleted.)
Calibrations > New >
List of saved calibrations
Este manual sirve para los siguientes modelos
2
Tabla de contenidos
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario