8
CHARGE THE 1.2V RECHARGEABLE BATTERY WITH THE
LINE-CHARGING UNIT
As soon as the symbol <BAT> is flashing the rechargeable 1.2V Mignon
battery must be charged.
Important: Before inserting the charging connector into the interface
plug of EASY-CHECK FE-S the device must be switched off.
The charger must not be connected when the 1.5V battery is inside
the device. The battery may run out and destroy the device !
With the device switched off the charging connector is inserted into the
interface plug of EASY-CHECK FE-S. The device switches on by itself
and a bar diagram with the battery symbol is displayed. With the charger
connected you can go on taking measurements.
After approx. 20 sec the state of charge is displayed again provided no
measurement is taken. When the charger is connected the automatic
switch off is deactivated.
The rechargeable battery needs approx. 8 hours to be charged com-
pletely. A constant bar diagram is flashing together with the battery sym-
bol and the charger can be removed.
GENERAL REMARKS
Thickness of the base material: at least 300 µm
Automatic switch-off
The device switches off automatically one minute after the last measu-
rement. The instrument can also be switched off with the red key.
IMPORTANT
The probe should not be drawn across the testing surface but reset at
different spots, i.e. after each measurement hold the instrument in the air
for about 1 sec. In doing so the stored calibration is automatically che-
cked and corrected if necessary.
Make sure that the probe surface and the base plate are kept clean
and polished at all times.