6 microAeth®Model AE51 Operating Manual microAeth®Model AE51 Operating Manual 7
AethLabsAethLabs
1.4 Technical
Specications
Measurement Principle
Real-time analysis by measuring the rate of change
in absorption of transmitted light due to continuous
collection of aerosol deposit on lter. Measurement
at 880 nm interpreted as concentration of Black
Carbon (‘BC’).
Measurement Range
0-1 mg BC/m3, lter life time dependent on
concentration and ow rate setting:
avg. 5 µg BC/m3 for 24 hours @ 100 ml/min
avg. 100 µg BC/m3 for 3 hours @ 50 ml/min
avg. 1 mg BC/m3 for 15 min @ 50 ml/min
Measurement Resolution
0.001 µg BC/m3
Measurement Precision
±0.1 µg BC/m3, 1 min avg., 150 ml/min ow rate
Measurement Timebase (User setting)
1, 10, 30, 60, or 300 seconds
Flow Rate (User setting)
Internal pump provides 50, 100, 150, 200 ml/min,
monitored by mass ow meter and stabilized by
closed-loop control.
Sampling
3 mm spot created on lter strip containing insert of
T60 Teon-coated borosilicate glass ber lter material.
PM2.5 size selective inlet available.
Consumables
Filter strip: 1 lter strip per sampling event, typically
one per day.
Data Storage
4 MB internal ash memory, providing up to 1 month
data storage when operating on a 300 second
timebase, and 1 week when operating on a 60
second timebase.
Communications
USB connectivity to Windows®-based PC with
microAethCOM, or data stream via digital output
through mini USB port (command protocols can be
supplied).
Data Output
Internal data les are uploaded to microAethCOM
PC software and stored on local disk.
PC Software
microAethCOM software is included. Provides visual
interface including real-time BC mass concentration
values. Facilitates settings conguration, calibration
routines, downloading data, and uploading new
instrument rmware.
Dimensions
4.6 in (117 mm) L x 2.6 in (66 mm) W x 1.5 in
(38 mm) D
Weight
Approximately 0.62 lbs (280 g).
Power
Internal rechargeable lithium-ion battery.
Power Supply Adapter
Input: 100~240 VAC 50/60 Hz 0.2 A
Output: 5VDC / 0.5A
Charging Time
4 hours to full charge (using AC adapter, instrument
turned off).
Total Run Time (single battery charge)
Minimum 24 hours @ 300 second time base at 100
ml/min ow rate.
Operation Environment
0 ~ 40 °C operating, non-condensing.
Please note: Specications are subject to change without
notice.
1.5 Symbols and Cautions
1.5.1 Explanation Operation Symbols
Operation indicator
Charging indicator
Aerosol inlet
Aerosol outlet
System alert indicator
Filter strip orientation arrow (point
indicates orientation of upstream
face of lter strip)
On/Off
USB port
1.5.2 Important Safeguards
Please read these safety instructions completely
before operating the instrument, and keep this
manual for future reference. Carefully observe all
warnings, precautions and instructions on the
instrument, or as described in the operating manual
and product literature.
Do not expose the microAeth or its batteries to
sources of excessive heat such as sunshine or re.
1.5.2.1 Power Source
The microAeth should be operated only from the
type of power source indicated in the instrument
specications. If you are not sure of the type of
electrical power supplied to your home, consult your
dealer or local power company. For those devices
designed to operate from battery power, or other
sources, refer to the operating instructions. Also,
the connections on both ends of the USB interface
cable are designed to be inserted into the AC
power/charger adapter or the microAeth only one
way. These are safety features. If you are unable to
insert the AC plug fully into the outlet, try reversing
the plug. If the plug should still fail to t, contact
AethLabs.
1.5.2.2 Object and Liquid Entry
Never push objects of any kind into the AC power/
charger adapter or into the microAeth (except for
the lter ticket) through openings as they may touch
dangerous voltage points or short out parts that
could result in a re or electric shock. Never spill
liquid of any kind on the microAeth or its electrical
accessories. This instrument should not be exposed
to rain or moisture, and objects lled with liquids,
such as vases, should not be placed on this
instrument.
1.5.2.3 Accessories
Do not use accessories not recommended by the
manufacturer, as they may cause hazards.
1.5.2.4 Servicing
Use extra care when servicing the instrument
yourself as opening or removing covers exposes
sensitive internal hardware to potential damage.
Refer to all service documentation and trained,
authorized service personnel for assistance.
1.5.2.5 Replacement Parts
Only genuine AethLabs parts should be used in
the microAeth. Only trained, authorized service
personnel should make repairs or install replacement
parts.
Lithium-Ion batteries are recyclable and should
be disposed of properly. Caution: Do not handle
damaged or leaking Lithium-Ion batteries.