Aesculap Suhl GmbH
Add-On Set for Hybrid Shearing Machines
4
2.3 Operating principle of charger
GT203/Li-ion battery GT201
The charger GT203 is designed for a mains voltage/
frequency range of 100 V to 240 V and 50 Hz to 60 Hz.
To make the charger GT203 ready for operation, con-
nect it to mains power by plugging in the mains plug.
The charger GT203 is equipped with to charging bays.
The first charging bay is for charging the battery-pow-
ered shearing machine FAV5 CL, while in the second
charging bay the Li-ion battery is charged.
Each charging bay has an indicator field on the charger
front panel assigned to it.
Charging starts automatically, as soon as a Li-ion bat-
tery is inserted in the charging station.
The charging time in the charging bays depends on the
charge state and capacity of the Li-ion batteries.
2.3.1 Charging principle
The Li-ion batteries are charged by constant current
pulses.
The charge state of the Li-ion battery is monitored
during charging. Monitoring of the charging curve
ensures that the battery is fully charged without over-
charging.
Additionally, the battery temperature and charging
time are monitored as well.
The charger monitors the battery status and indicates
the charge state via two LED indicators.
LED indicators:
• green = Battery fully charged
• red = Battery completely discharged
2.3.2 Charging time
Charging is stopped as soon as the maximum charging
time is reached.
The charging time is approx. 50 min.
3. Preparation and setup
Non-compliance with these rules will result in com-
plete exclusion of liability on the part of Aesculap.
¾Always observe the following guidelines when set-
ting up and operating the product:
– national regulations for installation and opera-
tion,
– national regulations on fire and explosion protec-
tion.
– operating advisories according to IEC-/VDE regu-
lations.
DANGER
Fire and explosion hazard!
¾Do not operate the product in
explosion-hazard areas.
CAUTION
Malfunction of, or damage to the
product caused by improper setup!
¾Do not expose the product to
direct sunlight or moisture.
¾Do not cover the ventilation
slots of the product.
¾Connect the product to power
mains with a voltage range of
100 V to 240 V and a frequency
range of 50 Hz to 60 Hz.
¾Ensure that the mains power
socket used for supplying the
product is freely accessible.