ADWA AD330 Manual de usuario

USER MANUAL
AD 330
EC/TDS/Temperature
Portable Meter
www.adwainstruments.com

2
Dear Customer,
Thank you for choosing an Adwa product.
Please read carefully this manual before starting operations.
This instrument is in compliance with the EMC directive 2004/
108/EC and its standards, and Low Voltage Directive 2006/95/
EC and its standards for electrical equipments.
For additional technical information, please e-mail us at
WARRANTY
Adwa warrants this product to be free of defects in material
and workmanship as stated in the operating manual. If repair
or adjustment is necessary and has not been the result of
abuse, misuse or improper handling within the warranty period,
please contact your dealer or nearest Adwa Office for the RGA
(Return Goods Authorization) number to put on the outside of
your package. Warranted service will be made without charge.
The meter is warranted for a period of 2 years, while probes
are warranted for 6 months. The warranty period commences
from the original date of sale. Warranty is only valid when the
product is used under normal conditions and in accordance
with the instruction manual. The warranty is void if the instrument
is repaired or serviced by unauthorized personnel, not used in
accordance to the instructions, or if non-Adwa accessories
such as buffer solutions, probes, etc. are used in conjunction
with the meter. Adwa will not be held responsible for any accident
whether directly or indirectly, caused by the use of this instrument.

3
Introduction ................................................................ 4
Technical Data .......................................................... 6
Front and Rear Panels .............................................. 8
Operational Guide .................................................... 10
Autoranging ............................................................. 13
Temperature Compensation .................................... 14
EC Calibration ......................................................... 16
Temperature Adjustment ........................................ 21
Setup ........................................................................ 22
Good Laboratory Practice ....................................... 25
Hold ......................................................................... 28
Battery Replacement .............................................. 29
Probe Maintenance ................................................. 30
Probes and Solutions ............................................... 31
TABLE OF CONTENTS

4
AD330 is a heavy-duty portable microprocessor-based
instrument for measuring conductivity, TDS and
temperature.
The autoranging feature of the EC and TDS readings
automatically sets the instrument to the scale with the
highest resolution.
Measurements are compensated for temperature effect
automatically (ATC) or manually (MTC).
The temperature compensation feature can also be
disabled to measure the actual conductivity.
The temperature coefficient is user selectable.
The instrument also features a measurement stability
indicator, GLP capability, and a user selectable ID
code to uniquely identify the instrument.
INTRODUCTION

5
This model is supplied complete with:
•AD76309 conductivity probe with built-in temperature
sensor and 1 m cable
• Calibration solution (20 ml sachet each):
• 1413 μS/cm EC standard solution
• 12,88 μS/cm EC standard solution
•1.5 V AA alkaline batteries (4 pcs)
•User manual

6
TECHNICAL DATA
Range 0.00 to 19.99 μS/cm; 0.00 to 9.99 ppm
20.0 to 199.9 μS/cm; 10.0 to 99.9 ppm
200 to 1999 μS/cm; 100 to 999 ppm
2.00 to 19.99 mS/cm; 1.00 to 9.99 ppt
20.0 to 199.9 mS/cm; 10.0 to 99.9 ppt
-9.9 to 120.0 °C
Resolution 0.01, 0.1, 1 μS/cm; ppm
0.01, 0.1 mS/cm; ppt
0.1 °C
Accuracy ±1% f.s. (EC/TDS)
(@25 °C/77 °F) ±0.5 °C
EC Calibration Offset at 0.00 μS/cm;
Slope at 1 point with 6 memorized values
(84.0, 1413 μS/cm; 5.00, 12.88, 80.0, 111.8 mS/cm)
or with custom value
Temperature Automatic or manual,
Compensation -9.9 to 120 °C
Temperature User selectable
Coefficient from 0.00 to 10.00%/°C
TDS Factor User selectable from 0.40 to 1.00
(default value: 0.50)

7
Reference Temperature
Selectable from 15.0 to 30.0 °C
Probe (included) AD76309 EC probe
with built-in temperature sensor
Battery Type & Life
4x1.5 V AA (Alkaline batteries)
approx. 200 hours of continuous use
Auto-Off User selectable
5 minutes or disabled
Environment 0 to 50 °C;
RH max 95% non-condensing
Dimensions 188 x 96 x 70 mm
Weight 460 g

8
FRONT AND TOP PANELS

9
1. Probe holder
2. Connector for EC probe
3. ON/OFF key, to turn the instrument ON and OFF
4. HOLD/GLP key, to freeze the first stable reading
on the LCD and to display Good Laboratory Practice
information
5. CAL/SETUP key, to enter/exit calibration and setup
modes
6. CFM key, to confirm values
7. SHIFT key, to activate the key alternate function
Press and hold first the SHIFT key and then the
second desired key
8. Down arrow key, to manually decrease the value of
temperature or other parameters
9. Up arrow key, to manually increase the value of
temperature or other parameters
10. RANGE/“lock” key, to select measurement unit
or switch the focused data, and to freeze current
range on the LCD
11. Secondary display
12. Primary display

10
• The meter is supplied complete with four 1.5 V AA
alkaline batteries. Remove the cover, unwrap the
batteries and install them while paying attention to their
polarity (also see “ Battery Replacement” section).
• To prepare the instrument for use, connect the EC probe
to the DIN connector. Tighten the threaded ring and
make sure the probe sleeve is properly inserted.
INSTRUMENT START-UP
• Turn the instrument on by pressing the
ON/OFF button.
• At start-up the display will show all used segments for
a few seconds (or while the button is held), followed
by the reference temperature value with “rEF”
indication, then enters the measurement mode.
OPERATIONAL GUIDE
Otros manuales para AD330
1
Tabla de contenidos