6. THE PARAMETER
6.1 The beat number (Frequency number) Manual
and Automatic selection
The automatic select beat number by the instrument as below:
12000, 14400, 18000, 19800, 21600, 25200, 28800, 36000, 43200
The instrument automatic display the closet beat number if the watch beat
number is not in the automatic selection range, but the test result is incorrect,
so need to select the beat number manually. The manual selection can select
the special beat number.
The manual select beat number as below:
3600, 6000, 7200, 7380, 7440, 7800, 9000, 9100, 10800, 11880, 12000,
12342, 12480, 12600, 13320, 13440, 13500, 14000, 14040, 14160, 14200,
14280, 14400, 14520, 14580, 14760, 14850, 15000, 15360, 15600, 16200,
16320, 16800, 17196, 17258, 17280, 17786, 17897, 18000, 18049, 18514,
19332, 19440, 19800, 20160, 20222, 20944, 21000, 21031, 21306, 21600,
25200, 28800, 32400, 36000, 43200
6.2 The lift angle
Most of the movement lift angle is 52°, so the instrument setting lift angle is
52°.
The lift angle degrees can set between 30° and 70° when calculating the
amplitude.
6.3 The test period
The test period is: 2S, 4S, 8S, 12S, 20S, 30S, 60S; the base time for calculate
the average test result.