6. Test Area: The test area is a general term used to describe a grouping of components in or near the
particle removal area. It includes: magnet assembly, delivery arm assembly, slide lifter, and vacuum
drain (complete descriptions of each of these items are contained in Section 8, Run FM Procedures).
7. Waste Bottle: The waste bottle acts as a reservoir for sample fluid passing over the glass substrate. It
is installed on the left side of the FM-6 unit.
8. Fixer Bottle: The fixer bottle is located on the right side of the instrument and provides solvent to
both slide making stations. The Fixer Heptane (Heptane can be purchased from VWR at
https://us.vwr.com or call 1-800-932-5000) is filtered to a specified micron level to insure additional
particulate is not present on the sample Ferrogram. It is strongly suggested to use the fixer supplied by
Trico when running the FM-6 unit.
NOTE: Methanol is NOT compatible with the pumps in this unit.
9. Leveling Feet: Two adjustable leveling feet are located on the bottom front of the case and are used
to level the magnet assembly.
10. Bubble Levels: There are two bubble level indicators located between station one and station two
magnet assemblies. The horizontal level indicates adjustments made in the leveling feet along the X
axis while the vertical level indicates adjustments made by the leveling feet in the Z axis.
Adjustments should be made to move the center of the bubbles as closes as possible to the black
center mark line on the level bubble.
11. Dispensing Arm: The arm raises and lowers the dispensing assembly onto the glass substrate in
order to dispense fluid. When in operation, the arm is lowered and the sample tube is inserted into the
tube holder at the tip.
12. Model and Serial Number Label: When calling Trico Corporation for repair work on the FM-6, this
information will be needed.
13. Power Input Jack: The AC line socket accepts the 3-conductor IEC type AC line cord provided with
the FM-6 instrument.
14. Ethernet Network Jack: Ethernet cable can be used to utilize internet/intranet Windows 7 functions.
15. USB Ports: External peripheral devices such as the provided mouse, a keyboard or monitor may be
utilized through the USB connections provided as long as they are Windows 7 compatible.
16. On/Off Power Switch: The on/off power switch manually controls the 115/220 VAC input power to
the FM-6 instrument. Red indicates ON position, Black indicates OFF position.
17. Back Panel: Taking the back panel off allows the operator to access the fixer pumps and peristaltic
tubing. Check with Trico before removing this component to avoid voiding the warranty.