1. Caution before use
This product is used for jetting powder. It may clog by moist air or by water. Do not use a delivery system in which the air pressure
may increase rapidly when it clogged.
How to check the air pressure of the delivery system.
•Run a high speed hand piece; listen to the sound it generates. If the sound stays consistent then there should be no variations
in the air pressure.
•Check the regulator in the junction box and check for proper settings. Generally your air setting will be set to approximately
80 PSI. There are several variations so the setting does not have to be exact.
2. Operation
(1) Connecting your air prophy to your HP tubing.
4-Hole, or 2-Hole type Powder Case
Connect the powder case directly to the corresponding handpiece hose.
(2) Mounting nozzle to Powder Case
Slide back the handpiece release ring on the powder case joint and
hold. Insert the nozzle into the powder case joint, and release the ring.
(3) Filling Cleaning Powder
Unscrew the powder case cap and remove. Fill the powder chamber to
small groove on the chamber. Recap the powder case cap securely.
•Do not use left-over powder or powder that has been sitting open in the
package. Moisture can solidify
(4) Jetting Powder
Aim the nozzle at the cuspidor, and jet for 3 seconds to verify that the
powder jets uniformly. And use on the patient. Use with the nozzle 5-10 mm
•Do not jet the powder directly to the oral soft tissues, gum or
sublingual.
•Always use the high volume evacuator to prevent the patient from
swallowing the powder. Use intermittently and have the patient wash
his or her mouth as often as needed.
3. Care after Use
Open
1. Unscrew the powder case cap, and completely empty the powder case.
2. Slide back the nozzle release ring, and remove the nozzle from the powder
case.
3. Wipe saliva and debris off the nozzle with alcohol-immersed cloth.
4. Remove the powder case from the coupling, or from the nozzle.
5. Remove the 0-ring from the powder case cap, plastic powder case cover,
and thin gasket. Clean the powder case cover, 0-ring, thin gasket and the
serrated powder case cover ring.
6. Clean the powder case passages with the supplied wire tool. Clean the
powder case, its body and nozzle. If the powder passages are clogged
badly, you may place the nozzle in boiling water for 10 min to loosen the
obstructions.
7. Dry completely so that your unit is ready for its next use.