-7-
END TRIGGER MODE
Selecting this will return you to the SETUP menu.
EVENT RECALL
To recall an event and display the results onthe LCD, choose EVENT RECALL from the
SETUP menu. Pressing theêkey enables you to scroll from the latest event occurrence,
through to the oldest event in memory. This will enable the on site user to physically record
all data if necessary. When you have completed viewing all records, press the ENTER
button to exitback to the main menu. Below is a listing and explanation of information
displayed on the LCD of the VMS-200E.
Line 1:
Line 2:
Line 3:
Line 4:
Line 1: In the example above VER indicates the vertical axis recorded the highest
peak value and it's value was 166.7 millig's. The peak value for H1 12 millig’s.
Line 2: displays the peak value for the MIC and H2 channels.
Line 3: (2) Indicates that this event is the second stored in memory. To view the next
event occurrence, press theêkey.
Line 4: indicates the event occurred on December 19, 1996 and the time of the
event was on the 13th hour, 38th minute, 15th second of that day. (1:38:15 PM)
If the trigger level was never exceeded, and the VMS-200E was exited from MONITOR
mode, a message such as the following will be stored in memory:
Line 2 shows (3) indicating this occurrence (or lack thereof) is third in memory and the
VMS-200E was removed from monitor mode on November 11, 1996 at 10:45:01 am. If
you miss the desired selection, continue pressing theêkey and the selection will "roll
VER 166.7mg H1 12mg
MIC 64dBA H2 16mg
(2)
12-19-96 13:38:15
OPERATOR EXITED
FROM MONITORING (3)
WITHOUT A TRIGGER AT
11-11-96 10:45:01