After verifying that all of the necessary components were shipped and present, you can begin
the assembly of the system.
To assemble the 1812 Element into the housing
1. If you are using a female 1812 element, Insert the adaptors into the element (Figure 2,
#1,2)
2. Remove clamp (Figure 3, #1) and open up the Bottom End Cap (keep the Top End Cap
intact, Figure 4, #1).
3. Insert element into the housing in the correct orientation, i.e. inserted from the side of
the housing where the bottom end cap is, so that O-rings on the permeate line adaptor fits
snugly in the Top End Cap.
4. Push Bottom End Cap until element is fully inserted into the housing. This may take a bit
oorce.
Note: Applying Glycerin to the two O-rings installed on the permeate tube is
recommended to ensure that there will be no leaks.
5. Put high-pressure clamp back on the Bottom End Cap and tighten.
WARNING: Do not exceed 600 psi (41 bar) in the 1812 Housing.
To remove the 1812 Element from the housing:
1. Remove the clamp (Figure 4, #1) and open up Top End Cap
2. Remove the element by pulling the element through the Top End Cap (Figure 4, #1).
3. Separate the element from the Top End Cap by pulling it out.
4. If you are not using this element any more, Remove the adaptors and save them for next
element.
3. 1812 Housing Assembly