Sterling Scale DW800 Manual de usuario

DW800/820 Operation Manual
STERLING SCALE COMPANY
DW800 HIGH ACCURACY WEIGHT INDICATOR
AND
820 HIGH ACCURACY PIECE COUNTER
USERS MANUAL
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]

DW800/820 Operation Manual
Table of Contents
Pages
INSTALLATION-------------------------------------------------------------------------------------------------------1 - 4
Unpacking/Inspecting 1
Load Cell Wiring 2
Universal Bracket 3
880/820 Assembl Drawing 3
Installing the Platter Assembl 2
Connecting External Power Suppl 2
Jumper Settings 4
OPERATION-----------------------------------------------------------------------------------------------------------5 - 6
Weighing Operation 5
Weigh Ke s 5
Counting Operations 6
Count Ke s 6
PARAMETER SETUP AND CALIBRATION------------------------------------------------------------------7 - 10
Parameter Setup 7
Parameter List with Parameter Explanation 7 - 9
Calibration Setup 9-10
PRINTERS--------------------------------------------------------------------------------------------------------------11
Ticket and Tape Printer Driver 11
Model 795 and 895 Printer Driver 11
Model LD-2 Printer Driver 11
COMMUNICATION------------------------------------------------------------------------------------------------- 12 - 16
RS-232 Serial Interface 12
RS-232 Connector 12
Output Protocol 12
Computer Output Format 143
Scoreboard Output Format 14
Remote Control Codes 14
External Zero Switch 14
SPECIFICATIONS--------------------------------------------------------------------------------------------------15 - 16
WARRANY INFORMATION--------------------------------------------------------------------------------------17
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]

DW800/820 Operation Manual
PLEASE READ BEFORE ASSEMBLY INSTALLATION
Unpacking and Inspecting
Unpacking and Inspecting
1. Upon receipt of the shipment, check for an shipping damage. If there is damage then
report it to our shipping carrier immediatel .
2. Carefull unpack all contents from the shipment.
3. Save all shipping material in case contents require transporting at a later date.
4. The content of the shipment should include the following:
Qty: Item:
1 Instrument with options installed.
1 External power suppl if desk mount or NEMA 1 enclosure is provided.
1 Instruction sheet.
1 Hardware kit.
If an of the above items are missing, contact our dealer immediatel .
Assembly
The instrumentation is shipped full assembled. If the unit shipped to ou is a desk mount
indicator, it contains a built-in scale and connecting the load cell is not necessar . If ou
received a wall mount, a stainless steel or a standard enclosure the load cell ma need to be
connected.
Connecting Load Cell
1. Disconnect power to the indicator.
2. Set the unit on a flat surface, open unit and locate “P4” on the lower part of the board.
3. Feed the load cell cable through the strain relief on the left.
4. Connect the load cell cable to the load cell interface connector, “P4” on the
circuit board.
5. Verif that jumpers “J2 and ”J3” are installed on the circuit board.
INSTALLATION
Page 1
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]

DW800/820 Operation Manual
Pin Name Description
1 +EX + Excitation voltage (+10 VDC)
2 +SN(+EX2) + Sense line
3 +SG + Signal line
4 -SG - Signal line
5 -SN(-EX2) - Sense line
6 -EX - Excitation voltage (GND)
7 SHD Shield
8 +SG2 Do not connect
9 -SG2 Do not connect
The load cell interface connector is “P4” on the circuit board,
located in the lower left hand corner.
Connecting External Power Supply
NEMA 1 enclosures are supplied with a transformer power suppl , which plugs into the wall.
Power Required – 117 VAC +/- 10 VAC, 60 Hz
NEMA 12 and NEMA 4 enclosures are supplied with a 6’ power cord with a standard 3-prong
connector.
Power Required – 117 VAC +/- 10 VAC, 60 Hz.
Desk enclosures are supplied with a transformer power suppl , which plugs into the back of
the instrument and the transformer plugs into the wall.
Power Required – 117 VAC +/- 10 VAC, 60 Hz
Mounting brackets are supplied with a NEMA 1 standard and NEMA 4 stainless steel
enclosures.
(See diagram on the next page for proper positioning)
INSTALLATION
Page 2
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]

DW800/820 Operation Manual
Desktop Enclosure Assembly Drawing
INSTALLATION
PAGE
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]

DW800/820 Operation Manual
Jumper Settings
In order for the 800/820 to operate as desired hardware, Jumpers must be set properl .
Most of the jumper setting’s are set correctl at the factor and never need to be altered, but
occasionall there is a need to alter a jumper setting, therefore the are shown below
Jumper List
Number Name Setting Description
J1 Back light On Back light on
Off Back light off
J2* Load Cell Sense (+) On Not using load cell sense wires (+)
Off Using load cell sense wires (+)
J3* Load Cell Sense (-) On Not using load cell sense wires (-)
Off Using load cell sense wires (-)
J4* EPROM 256 Use 27C256 EPROM
512 Use 27C512 EPROM
J5 Ground On Chassis ground tied to digital ground
Off Chassis ground not tied to digital ground
J6 RAM 2K Use 2K RAM
8K Use 8K RAM
J7 Not available
J8 NTEP/CTS NTEP Enable NTEP operation
CTS Enable CTS flow control
J9 RAM 8K Use 8K RAM
32K Use 32K RAM
*NOTE: Jumpers J2 and is J8 must both be either connected or disconnected to avoid
improper weighing and/or possible damage to the indicator.
INSTALLATION
Page 4
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]

DW800/820 Operation Manual
Weighing Operation
Weigh Keys: During weighing operations the ke s are used as follows:
OPERATION
PAGE 5
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]
The ZERO ke is used to return the scale to the center of zero.
The TARE ke is used to tare the weight on the scale. Place the
empt container on the scale and then press the TARE ke . The
displa will indicate the new weight.
The PRINT ke is used to print a weigh ticket.
The GROSS/NET ke is used to change the displa ed weight from
Gross Wt. (tare + net = gross) to the Net Wt.
(gross – tare = net).
The UNITS ke is used to switch between weights, units, lbs., kg.,
grams.

DW800/820 Operation Manual
Counting Operation
The Counting Operation cannot be accessed, if the NTEP parameter is set to True
(see parameter setup) or the Units of Weight parameter is set to LB/KG, LB/G, KG/G, or ALL.
The Count Operation is started b pressing and holding the UNIT ke . The indicator will toggle
thru all allowable sample sizes, starting with the current sample size. When the desired
sample size is displa ed release the UNIT ke . The indicator will sample the pieces and
displa the count.
Count Key Operation: During counting operations the ke are used as follows:
OPERATION
PAGE 6
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]
The ZERO ke is used to zero the scale.
The TARE ke is used to tare the weight on the scale. Place the
empt container on the scale and then press the TARE ke . The
displa will indicate the new weight.
The PRINT ke is used to print a weight ticket.
The GROSS/NET ke is used to change the displa ed weight from
Gross Wt. (tare + net = gross) to the Net Wt. (gross – tare = net).
The UNITS ke is used to switch count and weigh operation. If held
down the count displa will show the average piece weight.

DW800/820 Operation Manual
Parameter Setup Operation
Parameter Setup ma be accessed using the following method:
1. From the Weigh Mode press and hold the GROSS/NET ke . The DW800/820 displa s
“GRSNET” to signif that the GROSS/NET ke has been pressed. After approximatel
3 seconds the indicator displa s “PARAMS” at which time the ke ma be released.
NOTE: Jumper J8 must be removed from the NTEP position, if NTEP is being utilized, before
access to Parameter Setup is allowed.
2. Once access is allowed to Parameter Setup, the NTEP F parameter
is displa ed.
NOTE: NTEP value should alwa s be “F”, unless ou are using the NTEP standards.
3. Press the PRINT ke to select the desired parameter.
4. Press the TARE ke to select the value for the currentl selected
parameter.
5. Press the PRINT ke to save the currentl selected parameter.
6. Repeat steps 3 thru 5 until parameters are set as desired.
7. Press the GROSS/NET ke or RESET ke to exit Parameter Setup.
Parameter List
Name Values Description
NTEP F If NTEP operation not required
T If NTEP operation required
FCAL F If not calibrating and/or configuring indicator
(Calibration) T if calibrating and/or configuring indicator
NOTE: If True then the DW800/820 will start the calibration routine
PCT 0 Minimum load % disabled
(% of loading) 0.2 If minimum is .2% of full scale capacit
0.5 If minimum is .5% of full scale capacit
1 If minimum is 1% of full scale capacit
NOTE: The value is the minimum weight allowed for parts counting
MSQ 1 Allows an number of samples
(Minimum 2 2 samples required
sample 5 5 samples required
quantit ) 10 10 samples required
20 20 samples required
NOTE: The value is the minimum number of samples allowed for an entered sample
size.
_____________________________________________________________________________________________________________
Revision 4. Copyright ©2004 Sterling Scale Co., Inc. May 2004
Web Site: www.sterlingscale.com E-mail: [email protected]

DW800/820 Operation Manual
PARAMETER SETUP AND CALIBRATION
PAGE 7
Name Values Description
ZT 0 Zero tracking disabled
(Zero Tracking) 0.2 Zero tracking at ± 0.2 graduations
0.6 Zero tracking at ± 0.6 graduations
1 Zero tracking at ± 1 graduations
3 Zero tracking at ± 3 graduations
NOTE: This scale will be re-zeroed within the chosen graduation
DA OFF Digital averaging disabled (Fast displa update)
(Digital Average LOW Digital averaging at low (Normal setting for
counting/weighing)
MED Digital averaging at medium Counting/weighing moderate
vibrations)
HI Digital averaging at high (Counting and/weighing
heav vibrations)
ZR 1.5 Zero range at 1.5 percent of full scale
(Zero Range) 100 Zero range at 100 percent of full scale
U LB. Pounds onl
(Units) KG Kilograms onl
G Grams onl
LB/KG Pounds to Kilogram conversion, counting operation is
disabled
LB/G Pounds to Grams conversion, counting operation is
disabled
KG/G Kilogram to Gram conversion, counting operation is
disabled
ALL Pounds, Kilogram, Grams conversion counting operation
is disabled
MB 0 Motion band disabled (Not recommended for normal
operations)
(Motion Band) 0.1 Motion band at 0.1 graduations (Most sensitive setting
counting)
0.2 Motion band at 0.2 graduations (Counting/weighing with
printer
0.3 Motion band at 0.3 graduations
(Counting/weighing with
printer)
0.4 Motion band at 0.4 graduations (Standard setting)
0.5 Motion band at 0.5 graduations
0.6 Motion band at 0.6 graduations
0.7 Motion band at 0.7 graduations
0.8 Motion band at 0.8 graduations
0.9 Motion band at 0.9 graduations (Least sensitive)
1.0 Motion band at 1.0 graduations (Least sensitive)
OF CMP Computer output format (See Data Interface Section)
(Output Format) SCB Scoreboard output format (See Data Interface Section)
290 Output format for 290 and 695 printers (See Printer Section)
795 Output format for 795 printers * (See Printer Section)