Press and release to increase the control set-point (while in normal control
mode) or to change lower display program parameter (while in User Menu or
Factory Mode). Press and hold to accelerate increment speed.
Press and release to decrease the control set-point (while in normal control
mode) or to change lower display program parameter (while in User Menu or
Factory Mode). Press and hold to accelerate decrement speed.
Pressing
key while in
normal control
mode
Press and hold for at least 2 seconds and release (while in normal control
mode) to access operator level parameters. Press and release to cycle
through all user parameters. Press and hold for 2 seconds and release to
silence audible alarm under normal power or on battery power. While unit is in
an alarm condition, the external alarm relay contacts will remain energized
until the alarm condition no longer exists. Press and hold to display chamber
temperature while the controller is on battery power. Chamber temperature will
be displayed until key is released. Alarm contact will remain energized while
operating on battery power.
Press both keys
simultaneously
Displays current AC power (i.e. 110VAC) as long as keys are pressed. If
power is 110VAC or 220VAC, unit will display 110. Mode is only active during
normal control mode (when top display = PV, Lower display = SP). N/A on
battery power.
Press both keys
simultaneously
Energize audible alarm and alarm relay output as long as keys are pressed.
Mode is only active during normal control mode (when top display = PV,
Lower display = SP). N/A on battery power.