-12-
-03-
Make sure thetest pin isconnected with thedevice terminal
safely
Make sure theinstrument is turnedoff after use.
Take out the battery, if theinstrument is expectednot to be
used for along time infuture.
Do not putthe instrument inthe harsh environmentof
sunlight, extreme temperatureand humidity.
Use wet clothof water orneutral cleanser toclean the unit,
do not useany abrasive orsolvent.
When the instrumentis wet,make surestore it afterit is
dried.
Note
Do not usethe instrument ifthere is anyabnormal run ofthe
instrument such asbroken housing andtest pin, orexposed
metal parts.
Do not shiftthe test rangeduring operating afterthe test pin
connected.
Do not disassemblethis instrument bythe user,please
contact our servicedepartment in caseof need.
Do not openthe battery dooror replace thebattery when the
instrument is wetand slippery,it must bedried with dry
cloth.
Turn offthe instrument incase of openingthe battery door
or replacing thebattery.
Warning
Do not operateunder environment offlammable gas, the
unit may generatespark during operationwhich may lead
to blast.
Do not connectthe test pinif the operatorhas a wethand.
Do not applyany power supplythat is abovethe allowed
limit or testrange of theinstrument.
Do not openthe battery doorduring operation.
Dangerous
Warning
Earthed
Electrode
Under Test
Rx re
Where earth for mains supply is used
Supply Transformer
Be sure to
check earthed
side of a
commercial
power supply.
Secondary
Side
Primary
Side
Red
Green
E
CP
In this method,the terminal Pand C hasbeen short connected.
This simple methodcan also beadopted to measurevoltage
with the greenwire connected withV connector.
Use volt meterto measure voltageof the commonterminal
to the earth,this instrument cannotbe used tomeasurethat,
it isvery dangerous,because uncertain stateof theterminal
and connecting wire,the voltage maynot be displayed,
which leads toelectric shock ordamage of theunit. This
unit cannot beused to measurethe common powervoltage,
this unit isnot designed forthis operation, whenusing quick
test head tomeasure(measure earth resistanceonly), the
termial Pand C willbe short connectedwhich will lowerthe
input resistance ofthe product.
4. In operation,press PRESS TO TEST key and turn it
clockwise to LOCK position can perform measure
without manual operation.
5. Aftermeasurement, turn theswitch to OFFif the unitis
expected not tobe used fora while later.
AR4105A