821S OPERATING MANUAL
Page 7 of 13
S\masters\manuals\821S
THE GAS DIVIDER MUST NOT BE USED WITHOUT A SAFETY
EARTH CONNECTION
If hazardous gases are being used, ensure that the OUTPUT port is ALWAYS connected to a
safe vent, either via the equipment being calibrated, or directly from the divider.
3 INSTALLATION
3.1 Introduction
3.1.1 Installation requires the use of a tool set compatible with electrical and pneumatic skills.
A suitable set of tools for a minimum installation consists of an electrician’s flat bladed
screwdriver for the mains connections, a sharp knife for cutting PTFE tubing, and a 9/16”
(14.3mm) A/F spanner. Full installation of remote control will require the use of a
soldering iron plus solder, wire cutters, wire strippers, small pliers, and a working
knowledge of the equipment to be connected. Plumbing in stainless steel will require the
use of pipe cutters and benders. We, or our local agent, can offer an installation service
if you do not have the necessary skills.
3.1.2 The pneumatic and electrical connectors are found on the rear panel.
3.2 Location
3.2.1 The gas divider should be placed in a dry and sheltered location out of direct sunlight,
avoiding drafts, and protected from rain. All mains power connections must be protected
from water ingress.
3.3 Mains Power Connection
3.3.1 The gas divider is factory set for a particular mains voltage. Check that the voltage
quoted on the rear panel is compatible with your local supply. The divider is not
sensitive to mains frequency variations and may be used over the full 50 Hz to 60 Hz
range.
3.3.2 The gas divider is supplied with a 2m long mains lead with an IEC320 socket at one end
for connection to the analyser. The other end should be connected to a mains plug to
suit the local supply.
3.3.3 Standard Wiring
•Connect the BROWN wire to the LIVE (L) pin.
•Connect the BLUE wire to the NEUTRAL (N) pin.
•Connect the GREEN & YELLOW wire the EARTH (E) pin.
•If the plug has provision for a fuse, one of 2 or 3 Amp rating should be used.
3.4 Gas Connections
3.4.1 Before the gases are connected to the 821, switch the mains supply on and check that
the pressure difference meter reads zero (centre zero). If necessary use a screwdriver
to adjust the potentiometer mounted to the right of the meter to obtain a zero indication.