SHOKO SCIENTIFIC Shodex CD-200 Manual de usuario

121056
READ
this operation manual
CAREFULLY
before using
Shodex® CD-200 Electric Conductivity Detector.
Electric Conductivity Detecto
r
Operation Manual
®CD-200

Copyright C 2011-2012, SHOKO SCIENTIFIC CO.,LTD. Printed in Japan.
All rights reserved. Contents of this publication may not be reproduced in any form or by
any means (including electronic storage and retrieval or translation into a foreign language)
without prior agreement and written consent from the copyright owner.

Notice
Users of Shodex®
CD-200
Electric Conductivity Detector (hereinafter Shodex
CD-200
) are requested to use it in strict accordance with the instructions in this manual.
Nothing contained in this manual shall be construed as giving any guarantee on
Shodex
CD-200
or granting or implying any license or immunity under any patent
or other rights.
Shodex
CD-200
must be used on the user's own responsibility and in strict
compliance with all applicable laws and regulations.
The information contained in this operation manual was obtained from sources
that we believe are reliable, but no warranty or representation is hereby given
as to its accuracy or completeness.
This operation manual does not expect Shodex
CD-200
to be used for clinical or
medical purposes. Shodex
CD-200
must not be used for such purposes.
The content of this operation manual is subject to change without notice.
Notice that the parts supply period for this product is 7 years from the end of
production. Please be aware that replacement parts may not be available beyond the
parts supply period.
Warranty
Except for a written warranty signed by its duly authorized representative and
specifically issued, SHOKO SCIENTIFIC CO.,LTD. makes no warranty,
express or implied,
written or oral, statutory or otherwise, as to the quality, performance,
workmanship,
fitness for a particular purpose, or merchantability of Shodex
CD-200
.

Precautions
Thank you for your continued patronage. Observe the following precautions
in order to make safe and stable use of the detector.
Precautions are divided into three groups in this operation manual,
depending on the degree of danger. The three groups are
: This sign is used where failure to follow strictly the instructions
and procedures indicated by this sign may cause injury or
damage to health.
: This sign is used where failure to follow strictly the instructions
and procedures indicated by this sign may cause this detector
and the peripheral apparatus to deteriorate or break down.
: This sign is used wherever information is given to ensure optimal
performance of the detector.
Precautions listed below are those of particular importance extracted
from
this operation manual :
Do not use the detector in places where combustible gas or any
source of fire or spark exists or might exist.
Prior to connection, make sure that the voltage of the power
socket into which the detector power cable is plugged is the
same as the power supply voltage indicated on the detector.
The power socket into which the detector power cable is
plugged should be of a 3P type with a grounding terminal.
Other types of power sockets should not be used.
When using organic solvent, wear safety goggles.
It is recommended that a sink or equipment for washing the eyes
be installed nearby in case the organic solvent in use comes into
contact with the eye(s) or skin.
When any abnormality, such as liquid leakage, is observed, immediately
turn off the power and unplug the detector from the main power
source.
!
Warning
Caution
Note
!
Warning
!
Warning
!
Warning
!
Warning
[ Important
! ]
!
Warning

Do not use the detector with the cover open; always unplug the
detector from the main power source before opening the cover.
If the detector is used in a manner not specified by this operation
manual, the protection provided by the detector may be impaired.
When using flammable organic solvent as mobile phase, be sure
to make necessary arrangement to prevent accidental ignition
(fire) by static electricity.
If the fuse blows after being replaced, power off the detector
immediately and disconnect the power cord. Please consult
our local representative in your area or SHOKO SCIENTIFIC.
As this detector is readily affected by the ambient temperature, use
it in places where there is little wind or change in the ambient
temperature. Do not use the detector near any source of vibration,
electrical noise, or in places where corrosive gas and a lot of dust
are present.
External input/output terminals are for contact closure.
Never apply voltage to them.
Do not connect any tube other than the provided "OUT tube"
to the eluent outlet joint of the detector. Put the exit end of
the "OUT tube" in the waste liquid bottle and do not apply back
pressure.
When connecting this detector to other detectors in series, put
it at the end.
Make sure that no back pressure is applied to the outlet of the
detector. Connect the tip of the Teflon tube on the cell outlet of the
detector to a receptacle for waste liquid.
If eluent freezes inside the detector, the flow line might break,
damaging the detector. If there is any chance of eluent freezing
during stoppage or storage of the detector, withdraw the eluent
completely from the flow line of the detector.
When using any eluent containing a salt at high concentration, make
sure that the flow line is washed with water thoroughly after use
.
Failure to do so may result in plugging of the line, causing the
detector to cease operation.
If the detector is not to be used for more than one week, prior
to storage, wash the flow line with pure water or acetone, and dry
the line by allowing nitrogen gas to flow through it.
Do not use any eluent which might corrode the material, such as
stainless steel, that it comes into contact with. Use of such eluent
might cause a base line drift and damage the detector.
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Caution
!
Warning
!
Warning
!
Warning
!
Warning
Caution

Use a thoroughly degassed eluent. It is recommended that a degassing
device be used to degas the eluent, as it is easy to use and permits
continuous degassing.
If an eluent is replaced with another eluent, replace all channels of the ion
chromatograph with new eluent. When eluent replacement is not sufficient,
it may become the cause of an unstabilized baseline.
When high voltage caused by static electricity is applied to the instrument,
some incorrect actions may be observed. Please take note of static
electricity.
Please do not set other equipment on this detector, in order to avoid
enlarging the baseline drift and losing control of the temperature.
Note
Note
Note
Note

Contents
1 . Accessories ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 1
2 . Features ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2
3 . Specifications ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3
4 . Measuring mechanism ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4
4-1 Detecting System ・・・・・・・・・・・・・・・・・・・・・・ 4
4-2 Electric System ・・・・・・・・・・・・・・・・・・・・・・ 5
5 . Nomenclature and Functions ・・・・・・・・・・・・・・・・・・・・・・ 6
5-1 Front Panel ・・・・・・・・・・・・・・・・・・・・・・・・・・ 6
5-2 Side Panel ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7
5-3 Back Panel ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 8
5-4 Display ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 10
6 . Installation and Connections ・・・・・・・・・・・・・・・・・・・・・・ 13
6-1 Power Connection and Grounding ・・・・・・・ 13
6-2 Signal cable Connections ・・・・・・・・・・・・・・・・ 14
6-3 Tube Connections ・・・・・・・・・・・・・・・・・・・・・・ 15
6-4 Operation Procedures ・・・・・・・・・・・・・・・・・・・・ 16
7 . Maintenance ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 18
7-1 Flow Line Cleaning ・・・・・・・・・・・・・・・・・・・・・・ 18
7-2 Replacing Fuse ・・・・・・・・・・・・・・・・・・・・・・・・ 18
7-3 Cleaning of detector
exterior ・・・・・・・・・・・・・ 18
8 . Troubleshooting ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 19
9 . WEEE Instructions ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 20

1. Standard accessories
When unpacking, please verify that all the following accessories are included:
【 Details of Accessories 】
No. Name Specification Quantity
1 Power cable
3-core cord with a grounding electrode (100V)
(1) For voltage of 100 or 120 VAC
Power supply cable set (125 V)
(2) For voltage of 220 or 240 VAC
Power supply cable set (250 V)
1*
2 Signal cable 2-core shield cable (approx. 2 m) 1
3 Fuse Time-lag type 500 mA (T500 mA L/250 V)
4 IN tube set
Teflon tube of 1.6 mm outer dia. x 0.25 mm inner dia.
x 1000 mm length
(PEEK setscrew with ferrule attached)
1
5 OUT tube set
Teflon tube of 2.5 mm outer dia. x 1.5 mm inner dia.
x 2000 mm length
(connecting tube, setscrew, and ferrule attached)
1
6 Discharge
tube set
Silicone rubber tube of 10 mm outer dia. x 7 mm
inner dia. x 2000 mm length
(connection joint attached)
1
7 Operation manual 1
*: Either 1-(1) or 1-(2) is attached according to the power supply voltage at the first destination.
When unpacking, please verify that all the following accessories are included:
- 1 -
1-(1) 1-(2)
4
3
5
2
6

2. Features
Shodex CD-200 is an Electric Conductivity Detector for HPLC and has the following features:
1) Stability
The double temperature control system provides for a very fast start up time
and excellent stable baseline performance.
2) Safety
Solvent leak sensor generates an output signal that stops pump flow in case of
an eluent leak within the detector.
3) Integration
Well-organized Output signal terminals and USB communication port guarantee
high-level integration and automation in conjunction with your HPLC system.
- 2 -

3. Specifications
1) Measuring method : Two-electrode system
2) Measuring limits : 0-600 mS/m
3) Measuring range : 0.0025-5.12/0.025-51.2/0.25-512 mS/m (12 steps)
4) Linearity range : 0-600 mS/m
5)
Noise
:
< 0.0005 mS/m (response; 1 sec) or less
(2.5 mM phthalic acid ; pH 4.0 ; 1.5 ml/min.)
6) Response : 0.1, 0.25, 0.5, 1.0, 1.5, 2, 3, 6 seconds
7) Zero adjustment : Automatic zeroing
8) Base line shift : Limit; 0-2 mS/m, Resolution; 0.01mS/m
9) Integrator output : 0-1 V (Sensitivity; 2/20/200 mV/mS/m)
10) Recorder output : 0-10 mV (for each range)
11) External contact input : ZERO IN/MARKER IN
12)
External contact output
:
・READY (Temperature control stabilized)/LEAK/
・ERROR (ROM/RAM/PARAMETER/LEAKAGE/ SENSOR
ERROR/OVER HEAT/ZERO OVER/ COMUNICATION)/
・MARKER OUT
(Contact capacity : 24 VDC, 0.1 A )
13)
Temperature control
:
OFF, 30-50℃ in increments of 1℃ (with a 77℃ thermal
fuse)
(Note) External cylinder is adjusted to a temperature
that is 5℃ lower than setting temperature.
14) External communication : USB
15) Cell volume : approx. 2.5 μl
16) Pressure rating : 1 MPa
17) Wetted part material SUS316, PTFE, PEEK
18)
Dead volume
:
Inlet → Cell ; approx. 75 μl
Cell → Outlet ; approx. 25 μl
Total dead volume ; approx. 103 μl
19) Power supply : 100-240 VAC ± 10% , 50/60 Hz
20) Demand : approx. 50 VA
21) Outside dimensions : approx. 260W x 400D x 150H mm (excluding rubber footing)
22) Weight : 8 kg
- 3 -