Lambda 365 6-Position Peltier Controlled Cell Changer,
Controller Installation Instructions
This instruction sheet describes the installation of this accessory which is used with the Lambda 365
Spectrophotometer.
NOTE:
Read these instructions before you install this accessory.
Contacting PerkinElmer
Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part
numbers.
See our website:
http://perkinelmer.com
PerkinElmer's catalog service offers a full selection of high-quality supplies.
To place an order for supplies and many replacement parts, request a free catalog, or ask for
information:
If you are located within the U.S., call toll free 1-800-762-4000, 8 a.m. to 8 p.m. EST. Your order will be
shipped promptly, usually within 24 hours.
If you are located outside of the U.S., call your local PerkinElmer sales or service office.
Features
•Full software control
•Liquid cooling system
•N2gas purging available
Figure 1 6-Position Peltier Controlled Cell Changer, Controller (with heated
reference) [P/N: N4101029] or Peltier Temp Unit (Multi) w/o Reference
Heating [P/N: N4101008]

09931286E
2
Dimensions and Specifications
Peltier Controller
Physical Characteristic Specification
Power 100-240 VAC, 50/60 Hz, 300W, Free Voltage
Temperature Range -5 to 100°C (Maximum Internal Temperature)
Maximum Ambient Operating Temperature 40°C
Dimensions 310 (W) x 410 (D) x 275 (H) mm
(12.2 (W) x 16.1 (D) x 10.8 (H) in)
Weight 20.0 kg (44 lb)
Coolant Volume 2 L
N2Gas Available 100 psi (6.9 bar)
Liquid Cooling System
Magnetic Stirrer Control Available
Emergency Power Off Switch Available
6-Position Peltier Controlled Cell Changer (with and without heated reference)
Physical Characteristic Specification
The number of Sample Cells 6
The number of Reference Cell 1 (with or without heated reference)
Dimensions 135 (W) x 384 (D) x 143 (H) mm
(5.3 (W) x 15.1 (D) x 5.6 (H) in)
Weight With heated reference: 3.4 Kg (7.5 lb)
Without heated reference: 3.1 Kg (6.8 lb)
Temperature Accuracy ±0.15°C
Temperature Precision ±0.1°C
Temperature Stability ±0.1°C
Ramping Time (from 0 to 100°C) 5 minutes

09931286E
3
Safety Warnings
WARNING
When this label is attached to an instrument it means refer to the
manual.
Lorsque cette étiquette est attachée à un instrument, il est nécessaire
de voir le manuel.
WARNING
There is risk of receiving a fatal electric shock if the fuses are replaced
with the power cord connected.
Il y a un risque d’électrocution si les fusibles sont remplacés tandis que
le cordon d'alimentation est encore branché.
See the following figure for the location of warning labels on the back of the instrument:
Warning
FUSE
Avertissement
FUSE
Warnin
g
To prevent electrical shock,
do not open cover.
Refer all servicing to qualified
personnel.
Avertissement
Afin d’éviter tout risque
d’électrocution ne pas enlever ce
couvercle.
Faire appel au service Après-Vente.
Warnin
g
For continued protection
against electrical shock,
replace with same type and
ratings of fuse.
Avertissement
Pour assurer la protection
contre le risque de choc
électrique, remplacez le fusible
seulement avec le même type
et la même valeur nominale.

09931286E
4
Configuration of Peltier Temp. Control Unit (Multi)
6-Position Peltier Controlled Cell Changer (with heated reference or without heated reference)
Peltier Controller
Power Cord, 3ea
Interface Cable (RS-232 and USB to RS-232), 1each
Spare Fuse (AC 250V T5AL), 2ea
Temperature Probe, 1ea
Magnetic Stirrer, 6ea
Coolant Hose, 1ea
Coolant, 2ea
Waste Basket for Coolant, 1ea
Waste Hose for Coolant, 1ea
Phillips round head screw with washer, 2ea
Description
6-Position Peltier Controlled Cell Changer (with heated reference or without
heated reference)
Figure 2-1 Lambda 365 6-Position Peltier Controlled Cell Changer (with heated
reference)

09931286E
5
Figure 2-2 Lambda 365 6-Position Peltier Controlled Cell Changer (without heated
reference)
a. Interface connector between 6-Position Peltier Controlled Cell Changer and Peltier Controller
b. Coolant Inlet/Outlet Quick Coupler
c. N2Gas fitting
d. Interface slot for Temperature Probe
e. Cell Lifting knob for sample cells
f. Cell Lifting knob for reference cell
g. Phillips round headscrew with washer
Used to fix a cell holder, a base plate or a front plate for Lambda 365.
Spare screws (2ea) are enveloped with the accessory.
Figure 3 Phillips round headscrew with washer (M4 *12L)
h. Temperature Probe
Figure 4 Temperature probe

09931286E
6
Peltier Controller
Front View of Peltier Controller
Figure 5 Front view
a. Manual touch pad with Display: Temperature Control Unit (Heating/Cooling Block).
b. LED Indicators: Display the status of the operation of Coolant Circulation, Fan, Pump, etc.
(If there is any problem in the components, Red LED will flash with beep alarm.).
c. Manual ON/OFF button: As the Manual ON/OFF button is ON, the temperature can be
controlled manually by a) Manual touch pad with Display.
NOTE:
Manual operation is used only for the manufacturer so it should be turned off during the measurement.
d. N
2
gas ON/OFF button: It is used for using N
2
gas to remove the frost from the heating
and cooling.
e. Stirrer ON/OFF button: It is used for controlling of the stirring.
f. Stirrer Speed Controlling Knob: Control the stirring speed (rpm).
g. Emergency switch: Stop all operations in the emergency situation.
NOTE:
During the operation, Emergency switch should be turned off.
h. Coolant Inlet

09931286E
7
Rear View of Peltier Controller
Figure 6 Rear view
a. RS-232 port
b. AC Power: Main Power ON/OFF
c. Fuse: AC socket + Fuse Holder
d. Interface: Interface cable for connecting with 6-Position Peltier Controlled Cell Changer
e. Air Vent Manual button: It is used for removing the air in the coolant hose.
f. Buzzer: It makes an alarm sound. If there is any problem before the measurement, Buzzer
beeps 6 times at every 10 sec. If the malfunction occurs during the operation, Buzzer beeps
twice at every 10 sec.
g. Quick Coupler of Coolant lnlet/Outlet
h. Quick Coupler of Coolant Drain
i. N2Gas Inlet/Outlet ports
NOTE:
N
2
gas is not always required [Tube for N
2
gas is not supplied with the accessory].
j. Flow Gauge: Indicator of the flow rate of N2gas.

09931286E
8
Installation
1. Prepare the Lambda 365 and Peltier Controller in a location that is compatible with the required
environmental conditions for the operation.
2. Open the lid of the coolant inlet on top of the Peltier Controller.
Figure 7 Lid of coolant inlet
3. Fill up the liquid coolant.
CAUTION
ATTENTION
Liquid coolant should be used as refrigerant. Do not use any water (tab water,
DI, etc.). If water is used as refrigerant, the Peltier Controller’s components might
be corroded and its performance may be deteriorated.
Use any kind of liquid coolant of normal grade that is available. (e.g. liquid coolant
including Distilled Water 70 wt%, Propylene Glycol 27 wt%, Additives 3 wt%).
Le liquide de refroidissement doit être utilisé comme réfrigérant. N'utilisez jamais
d'eau (eau du robinet, eau déminéralisée, etc.) L'eau risque en effet de corroder
les composants du contrôleur Peltier et de réduire les performances de ce
dernier.
Utilisez un liquide de refroidissement du commerce de qualité normale (par
exemple, un liquide de refroidissement à base d'eau distillée à 70 % M/V, de
propylène glycol à 27 % M/V et d'additifs à 3 % M/V).
NOTE:
Keep the amount of liquid coolant to the level where the indicator is located between the ‘L’(low) and
‘H’(high) mark in the scale on the left side of the Peltier Controller. When the coolant is running short for
the operation, LED of Low Water Level is changed red with alarm sounding.
4. Connect the power cord and communication cable of the Lambda 365. DO NOT turn on the power
of the instrument!
CAUTION
ATTENTION
Make sure the instrument is turned off while installing this accessory.
Assurez-vous que l’instrument est éteint lors de l’installation de cet accessoire.

09931286E
9
5. Remove the two Phillips round head screws with washer (M4*12L) to disassemble the existing cell
holder and base plate.
Figure 8 Location of Phillips round head screws with washer
6. Pull out the cell holder and base plate by hand.
Figure 9 Pulling out the cell holder and base plate
7. Insert the 6-Position Peltier Controlled Cell Changer in the sample compartment.
Figure 10 Inserting the 6-Position Peltier Controlled Cell Changer

09931286E
10
8. Gently press the 6-Position Peltier Controlled Cell Changer to connect the communication port (male)
under the bottom of the 6-Position Peltier Controlled Cell Changer to the port (female) in the
sample compartment.
Figure 11. Connecting the communication ports
9. Fix and tighten the 6-Position Peltier Controlled Cell Changer with the Phillips round head screws.
or
Figure 12 Location of screws (with heated reference or without heated reference)