Main Frame
Returns automatically to this
display 30 sec after last key
operation
Actual Set 180°C
Continuous Mode
180 °C
Fan Speed
08
Air Valve
080%
Min Aver Max
179 180 181
Logg °C
Model TS9053
Rel: 001 003 008
180 °C
Real Time
Day Hrs:Mins
Mo 20:15
Status Messages:
*TimeLeft 1h 25min
*Cycle Complete
*Heating Up
*Cooling Down
*Sleep Mode Alarm Messages:
*High Temp Alarm
*Low Temp Alarm
*High Temp Safety
*Sensor Failure
*Adjust Real Time!
*Data Error
Temperature Target Actual Temperature
Actual Fan Speed
Not allways equal to Set Value
May have been automatically
adjusted for optimal control
Display Only Frames,overview
Selected with Up/Down Keys when no light at the Set Key
Actual Air Valve Opening
Not allways equal to Set Value
Closed during Heating Up and
when Cyce Completed
Logged Temperature
Start logging when Set
Temperature is reached (30
minutes delayed)
Cleared at temperature change
and when the timer is turned on
Show Minimum, Average and
Maximum temperature since
last temperature change or
timer start
Zeros only if logging not yet
started
Actual Day and Time
This frame comes up only when
the Real Time Program is chosen
in the System Setup
Model and Firmware
Show actual Chamber Model
and Firmware Release
numbers for the different
microcontrollers in the system
Temperature Logging
•Starts when the temperature has been reached, about 30 minutes delayed
•Used to evaluate the temperature control
•Updates every minute
•Door openings and Air Valve adjustment will affect the Max and Min values
•All zeros means that the logging has not yet started
•Cleared when Power Up, wakeup from Standby, entering new Temperature Set and
when turning on the Timer