We thank you very much for your buying AUTOMATIC ELIMINATION PROBE MODEL 01-00013B FOR
ESS-Series. This unit is connected to ESS so as to automatically discharge EUT (Equipment under test)
that is charged by applying static electricity. It is recommended that this manual be thoroughly
understood and used as ready reference for operation.
For the safety precautions, warranty terms and so on, refer to the relevant section of the manual of
your simulator.
1. CONTENTS IN THE PACKAGE
Automatic elimination probe··················· 1 pc.
Instruction manual (this manual)············· 1 pc.
2. NAME AND FUNCTION OF EACH PART
①
①①
①
UX connector: Connects to AUX CONNECTOR on the rear panel of ESS.
②
②②
②
AUX connecting cable
③
③③
③
Elimination switch/lamp: The pilot lamp is on during elimination. When this switch is pressed to
set to no-elimination position (Lamp is off), manual elimination can be performed.
④
④④
④
HOT side clip (Red): Connect the cabinet/earth terminal of EUT.
⑤
⑤⑤
⑤
HOT side elimination resistance: Incorporates resistance of 470kΩ.
⑥
⑥⑥
⑥
HOT side cable (White): 30kV high-voltage cable is used.
⑦
⑦⑦
⑦
GND side clip (Black): Connects to the ground of electrostatic simulator system
⑧
⑧⑧
⑧
GND side elimination resistance: Incorporates resistance of 470kΩ.
⑨
⑨⑨
⑨
GND side cable (Black)
⑤
④
⑨
③
①
②