Calibration procedure
CONDUCTIVITY (COND5 / PC5 / PC6)
1. Power ON the meter by pressing the key..
2. Rinse the probe with distilled water.
3. Immerse the probe in the calibration solution (84µS or 1413µS or 12.88mS), wait for stability.
4. Press CAL key, (blue backlight during calibration).
5. The meter will start calibration procedure and will recognise automatically the standard
used.
6. When calbration is stable press to conrm and complete the calibration.
7. The standard value will ash for 3 times and then the meter will go into the measurement
mode.
8. If the 2nd point calibration is required then rinse the electrode with distilled water and im-
merse in the 2nd Standard, wait for stability.
9. Repeat the points 4 to 7.
10. If the 3rd point calibration is required then rinse the electrode with distilled water and im-
merse in the 3rd Standard, wait for stability.
11. Repeat the points 4 to 7.
12. The calibration process is completed and the meter is ready to use.
1. Power ON the meter by pressing the key.
2. Rinse the electrode with distilled water.
3. Immerse the electrode in the 1st buer solution pH7.00 and wait for stability.
4. Press CAL key (blue backlight during calibration).
5. The meter will start calibration procedure and will recognise automatically the standard
used.
6. When calibration is stable, press to conrm and complete the calibration.
7. The standard value will ash for 3 times and then the meter will ask for next point for calibrat-
ion. If only 1-point calibration is required then press to nish and exist.
8. If the 2nd point calibration is required then rinse the electrode with distilled water and im-
merse in pH4.01 or pH10.01, wait for stability, otherwise press ESC to nish and exit.
9. Repeat the points 4 to 7.
10. If the 3rd point calibration is required then rinse the electrode with distilled water and im-
merse in the last buer, wait for stability, otherwise press ESC to nish and exit.
11. Repeat the points 4 to 7.
12. The calibration process is completed and the meter is ready to use.
Note: Anytime press ESC key to abort and exit from calibration procedure.
Note: Anytime press ESC key to abort and exit from calibration procedure.
Note 2: In case multipoint calibration is performed it is better to start from the lower value standard rst and
then go increasing.
PH (PX4 / PH5 / PC5 / PC6)
Calibration procedure