This summary is intended only as a
quick reference and is not a substitute
for GlucoDr.S Blood Glucose Monitoring
System User Manual. Please read the
entire User Manual before you begin
the test.
TEST SUMMARY
Check the expiration date
printed on the strip vial
label.
Insert a test strip. The
meter will
automatically turn on.
The blood glucose
result will be displayed
on the meter in just
5 seconds.
Apply the blood drop to
the side edge of the
yellow window of the test
strip.
Wait for the blood drop to
completely ll the yellow
window of the test strip
and then the meter will
begin the countdown.
Remove the battery
cover and put the
battery with “+” sign
facing up.


Test Principle
The GlucoDr.S Blood Glucose Monitoring System is based on measurement of electric currents caused by
the reaction of glucose with the reagents on the gold electrode of the test strip. The blood sample is
drawn into the test strip's reaction chamber through capillary action. Glucose in the sample reacts with
glucose dehydrogenase and mediator on the test strip. This reaction creates electric currents. The
consequent electric currents are proportional to the glucose concentration in the blood and converted
to the equivalent glucose concentration by the algorithm programmed in the GlucoDr.S Meter.
Healthcare Professionals
Blood glucose determination with venous blood must be performed within 15 minutes of sample
collection to avoid error from glycolysis.
In order to avoid the eects of sample temperature, keep the venous blood sample to body temperature.
The blood should be collected in EDTA or heparin containing tubes.
Do not shake the tube vigorously.
For Self-testing: GlucoDr.S Blood Glucose Monitoring System is intended for self-testing outside the body
(in vitro diagnostic use) for quantitative measurement of glucose in fresh capillary whole blood drawn from
ngertips. The GlucoDr.S Blood Glucose Monitoring System is intended for use by diabetic lay users as an
aid to monitor their blood glucose level. The GlucoDr.S Blood Glucose Monitoring System should not be
used for the diagnosis of or screening of diabetes mellitus, or for neonatal use.
For professional use: GlucoDr.S Blood Glucose Monitoring System is intended for testing outside the
body (in vitro diagnostic use) for the quantitative measurement of glucose in venous whole blood and
fresh capillary whole blood drawn from ngertips. The GlucoDr.S Blood Glucose Monitoring System is
intended for use by healthcare professionals as an aid to monitor patients’blood glucose level.
The GlucoDr.S Blood Glucose Monitoring System should not be used for the diagnosis of or screening of
diabetes mellitus, or for neonatal use.
Intended Use
PRECAUTIONS

Precautions for Test Strip
- Store test strip vials in dry place with room temperature between 1~32℃(34~90℉).
- Avoid direct sunlight, heat and excessive humidity.
- Always close the vial cap immediately after removing a test strip from the vial. If the test strip vial is left
open for a long time, the test strips will become unusable.
Carefully read and follow the instructions in the User Manual and package inserts for the test strips and
control solution and practice the testing procedures before using the GlucoDr.S Blood Glucose
Monitoring System. It is very important to follow the instructions in order to avoid an incorrect test
result that leads to improper therapy.
Take note of information and cautions throughout this User Manual.
- The GlucoDr.S Meter is designed for use only with GlucoDr.S Test Strips.
(manufactured by All Medicus Co., Ltd.)
- The GlucoDr.S Blood Glucose Monitoring System is calibrated to produce values equivalent to the
results on plasma specimens on a laboratory analyzer (YSI 2300 STAT Plus).
- The GlucoDr.S Blood Glucose Monitoring System should not be used for the diagnosis of diabetes.
-Rapid change in temperature may cause inaccurate test results in reading. When taking the meter from
cold to warm area or from warm to cold area, let the meter sit for about 30 minutes to adjust to the
room temperature.
- The GlucoDr.S Blood Glucose Monitoring System is for in vitro diagnostic use.
-Do not drop the meter.
-Do not disassemble, repair or remodel without consultation. The sensitive parts could be damaged
and the warranty will then become invalid.
- A hematocrit that is either very high (above 65%) or very low (below 20%) can cause false test results.
- Handle all the reagents with care knowing that they are capable of transmitting infection.
- Used test strips, lancets and meter may be considered bio-hazardous waste in your area. Be sure to
follow your local regulation for proper disposal of used strips, lancets and meter.
Precautions for Use

Safekeeping
- Store the GlucoDr.S Meter in a room temperature between 0~40℃(32~104℉).
- Store the GlucoDr.S Meter where relative humidity is between 15~85%.
- Keep the GlucoDr.S Meter in sanitary environment.
- Keep the meter away from direct sunlight for a long time.
- Keep the test port away from dirt, blood or water.
- Do not store your meter and test strips in the car, the bathroom, or the refrigerator.
(sensitive to temperature and humidity)
- Keep the meter, test strip vials and lancing materials away from children.
- Clean the outside of the meter using a moist (not wet) cloth or tissue with isopropyl alcohol or mild
detergent with water. Do not immerse the meter in water or other liquid.
- Use the test strip immediately after removing it from the vial.
- Use all the test strips within 4 months after the rst opening.
- Store test strips only in their original vial and do not transfer them to a new bottle or any other container.
- Do not touch the test strip with wet hands.
- Do not use the test strip beyond the expiration date printed on the vial.
- Do not bend, cut, or damage the test strips.
- Do not insert the same test strip into the test port multiple times. It may cause an improper operation.
Please read the literature packaged with test strip and meter to nd the
product specication and limitations.
If the meter does not work properly and you have followed the storage and
handling instructions in this User Manual, contact your local distributor.

Display Message Guide l 34
Troubleshooting Guide l 40
Specications l 41
Performance Characteristics l 42
Explanation of Symbols l 45

GlucoDr.STM Meter
Battery
Carrying case
User Manual
GlucoDr.S Blood Glucose Monitoring
System includes the following items:
Upon request of distributor, GlucoDr.S
Blood Glucose Monitoring System may
include the following items:
GlucoDr.STM Test Strips
Lancing Device
Lancets
All the including items are listed on the main box.
Check the items after opening GlucoDr.S Blood Glucose Monitoring System package.
If any of the items are missing from your kit, Contact your local distributor.
7
GlucoDr.STM Blood Glucose Monitoring System
1

- Test results, icons, symbols and simple messages appear here.
LCD Display
Power Button
- Turn the meter on/o.
- Enter into SET MODE while pressing the power button for more
than 3 seconds.
- Delete all the stored test results in the AVERAGE MODE.
- Delete each test results in the MEMORY MODE.
Test Port
- Insert the GlucoDr.S Test Strip for testing. Then the meter turns
on and BLOOD DROP ICON will blink.
</> Button
- Recall the stored test result in the MEMORY MODE.
- Select an event in the EVENT RECORD MODE.
- Change the date, time and options in the SET MODE.
GlucoDr.STM Meter
2

Battery Cover
Micro USB Port
- Transfer your test results from the
meter to a computer.
Test Strip Eject Button
GlucoDr.STM Meter
3
Este manual sirve para los siguientes modelos
1
Tabla de contenidos
Otros manuales de Medidor de glucosa en sangre de Medicus
Manuales populares de Medidor de glucosa en sangre de otras marcas

ReliOn
ReliOn Prime Manual del operador

AgaMatrix
AgaMatrix Presto Manual de usuario

Polymer Technology Systems
Polymer Technology Systems A1C Now Self Check Manual de usuario

Medisana
Medisana GlucoDock Manual de usuario

Arkray
Arkray ASSURE PLATINUM Manual de usuario

Dexcom
Dexcom G4 Professional Manual de usuario

LabStyle
LabStyle Dario Manual de usuario

ACONLABS
ACONLABS ON CALL EXPRESS Manual de usuario

Higi Technology
Higi Technology iGlucose Manual de usuario

Bayer HealthCare
Bayer HealthCare CONTOUR NEXT Manual de usuario

Accu-Chek
Accu-Chek ADVANTAGE Manual de usuario

Sanofi-aventis
Sanofi-aventis BG STAR Especificaciones técnicas