6
Intended Use
The Airo 24 by Mangar Health is a low pressure
air compressor, which has been designed to
inate the Mangar ‘Camel’ and ‘ELK’ pneumac
liing equipment through a 4-Way pneumac
hand control. The 4-Way hand control is
supplied with the Camel and ELK.
Applied Parts (Accessories) for the Airo 24:
Mangar ELK (Emergency liing cushion)
Mangar Camel (Emergency liing cushion)
Contents
Intended Use Pg 6
Warnings Pg 6
Baery Status Indicator Pg 7
Baery Charging Pg 8
Mains Power Adaptor (oponal) Pg 9
Operaon Pg 9
Cleaning & Storage Pg 10
Maintenance Pg 10
Trouble shoong guide Pg 11
Specicaons Pg 11
Symbols used on Airo 24 and Charger Pg 12
Symbols used on packaging Pg 12
Warranty Pg 13
Declaraon of Conformity Pg 51
WARNINGS
•Be sure that you have read and understood
these instructions and those provided with any
other Mangar Health product, before
operating the equipment.
•To avoid the risk of electric shock, NEVER
recharge the battery in the bathroom,
outdoors or in a damp environment.
•Always disconnect from the battery charger
before cleaning.
•Use only the supplied battery charger.
•Do not operate the Airflo 24 compressor
in the presence of flammable anaesthetic
mixtures with air, oxygen or nitrous oxide, as
these mixtures can create a combustible
environment leading to the potential for fire or
even an explosion in the presence of a spark.
•Ensure that when in use, the Airflo 24 and its
connected hoses are laid out in such a way as
to avoid any risk of tripping or strangulation.
•Ensure that children and pets do not interfere
with the safe operation of the Airflo 24 and
inflatable product.
•Keep the Airflo 24 and inflatable product away
from sources of heat.
•Disconnect the Airflo 24 compressor from the
inflatable product before transporting so as to
avoid accidental inflation.
• The Airflo 24 should only be used for its stated
purpose and should not be modified in any way.
•The optional mains power adaptor is specific
to the Airflo 24 and is the only approved unit
to be used for this purpose.
•Do not operate the compressor outside of the
specified duty cycle of 20% ‘Operating Time’
and 80% ‘Non-Operating Time’. Maximum
continuous operating time = 5 minutes.
•To calibrate the battery and battery indicator, it
is essential before the first use to charge for 12
hours and then disconnect from the charger
and leave for 90 minutes to allow the battery
management software to correctly calibrate.
Failure to do this could lead to incorrect
battery charging and battery ladder indication.
This product and associated packaging does not
contain latex.
The ELK requires assistance to help li a paent
from the oor to an elevated sing posion.
Once in a sing posion, the paent can be
transferred to a chair or helped to a standing
posion.
The Camel can be operated with or without
assistance and when inated, lis the paent
into a raised seated posion ready to stand.
No special skills or training are required to
operate the Airo 24 and its accessories.
Further guidance on operaon can be found in
the individual accessory User Instrucons.
The Airo compressor and applied parts are
intended for connual re-use as required and
no special acons or operaons are needed to
achieve this.
English
Airo 24