Larson Davis CAL200 Manual de usuario

CAL200
Precision Acoustic Calibrator
Manual


I200.1 Rev H
APPENDIX
Larson Davis
CAL200
Manual

Copyright
Copyright 2006-2015 by PCB Piezotronics, Inc. This manual is copyrighted, with all rights reserved.
The manual may not be copied in whole or in part for any use without prior written consent of PCB
Piezotronics, Inc.
Trademarks
PCB®is a registered trademark of PCB, Inc.
Disclaimer
The following paragraph does not apply in any state or country where such statements are not
agreeable with local law:
Although PCB Piezotronics, Inc. has reviewed this documentation, PCB Piezotronics Inc. makes no
warranty or representation, either expressed or implied, with respect to this instrument and
documentation, its quality, performance, merchantability, or fitness for a particular purpose. This
documentation is subject to change without notice, and should not be construed as a commitment or
representation by PCB Piezotronics, Inc.
This publication may contain inaccuracies or typographical errors. PCB Piezotronics, Inc. will
periodically update the material for inclusion in new editions. Changes and improvements to the
information described in this manual may be made at any time.
Recycling
PCB Piezotronics, Inc. is an environmentally friendly organization and encourages our customers to
be environmentally conscious. When this product reaches its end of life, please recycle the product
through a local recycling center or return the product to:
PCB Piezotronics, Inc.
Attn: Recycling Coordinator
1681 West 820 North
Provo, Utah, USA 84601-1341
where it will be accepted for disposal
Product Warranty
For warranty information, refer to our Terms and Conditions of Sale on our website at
www.larsondavis.com/TermsConditions.aspx.

CAL200 Manual Introduction 1-1
CHAPTER
1Introduction
This chapter describes the features of the Larson Davis
CAL200 Sound Level Calibrator.
Features
The Larson Davis CAL200 sound level calibrator is a
battery operated, precision instrument used for calibrating
sound level meters and other sound measurement
equipment. It provides an output level of either 94.0 or 114.0
dB (switch-selectable) at a frequency of 1 kHz. It has been
designed for both field and laboratory use and the accuracy
has been calibrated to a reference traceable to the National
Institute of Standards and Technology.
The Larson Davis CAL200 Sound Level Calibrator has the
following features:
• Precision sound level calibrator
• Dual output levels: 94.0 dB and 114.0 dB (user
selectable)
• Output frequency 1 kHz
• Internal batteries for portable operation
• Opening for use with 1/2” microphones
• Adaptor ADP031 available for use with 3/8”
microphones
• Adaptor ADP024 available for use with 1/4”
microphones
• Adaptor ADP075 available for use with 1/8”
microphones

1-2 Features CAL200 Technical Reference Manual

CAL200 Manual Using the CAL200 Calibrator 2-1
CHAPTER
2Using the CAL200 Calibrator
This chapter describes how to operate the CAL200
calibrator.
Installing the Battery
The CAL200 uses a nine-volt battery and it is recommended
that you use an alkaline battery to extend the running time of
your calibrator. The CAL200 will run for approximately 100
hours on one nine-volt battery when using the 114 dB output
level. This will give you nearly 4200 calibrations.
Follow the steps below to install the battery:
Step 1 Slide the battery door on the lower section of the
back panel up to remove it.
Step 2 Gently pull the battery connector clip sufficiently
outside the interior to permit it to be snapped onto the
battery terminals.
Step 3 Tuck the battery, with the battery connector clip
attached, into the battery compartment
Step 4 Slide the battery door back into place, pressing
until the tab snaps the door securely shut.
Microphone Calibration
The CAL200 provides a nominal tone of 1 kHz, which
requires no weighting corrections when used with sound
level meters utilizing A, B, C or Z (Flat) frequency
weighting.

2-2 Microphone Calibration CAL200 Manual
Select the Output Level
The output level is user-selected to be 94.0 or 114.0 dB
using the slide selector on the side of the CAL200. Slide the
switch to lower position for 114.0 dB or to the upper
position for 94.0 dB, as indicated by the label on the lower
left of the front panel.
CAL200 Used to Calibrate an Instrument with ½" Free-Field Microphone
The CAL200 provides a nominal pressure level of 94 dB or
114 dB. The exact levels are printed on the Larson Davis
calibration sheet that came with the calibrator. When using a
free-field microphone, the pressure level at the microphone
diaphragm will be slightly different. Thus, a free field
correction of -0.12 dB should be applied to either of these
levels. Pressure and random incidence microphone do not
require a correction of this type. If the calibrator and
instrument are near room temperature (23° C) and near sea
level (101.3 kPa) then no other corrections need to be made.
If the calibration sheet for the CAL200 indicates 113.98 dB
for its level when set to 114 dB then set the Cal Level in the
instrument to 113.86 dB and 1 kHz.
When the microphone and instrument are at a temperature
other than near room temperature or static pressures not near
sea level, then corrections will need to be added for the
ambient temperature and the prevailing static pressure.
Check the calibration data shipped from Larson Davis with
the CAL200 to get these corrections. The corrections can be
added to the level obtained in the previous paragraph to get
the actual level of the CAL200. The microphone's sensitivity
varies with static pressure. If the instrument is calibrated in
one environment and moved to another, then the sensitivity
will change (after stabilization) depending on the change of
temperature and pressure. The coefficient of static pressure
is typically -0.013 dB/kPa for PCB®½” free-field
microphones. If the system is calibrated at 85 kPa, for
instance, then it will be 0.21dB less sensitive at sea level.
The sensitivity of the microphone varies slightly with
temperature also. The coefficient of temperature is typically
-0.009 dB/°C for PCB ½” free-field microphones. If the
system is calibrated at 18 °C then it will be 0.05 dB less
sensitive at 23 °C.

CAL200 Technical Reference Manual MicrophoneCalibration 2-3
Perform the Calibration
To calibrate the microphone, follow the steps below:
WARNING! Do not remove the
microphone gridcap while
calibrating as this may lead to
microphone damage.
Step 1 Place the instrument so that you can perform the
calibration vertically, as shown in Figure 2-1.
Step 2 Insert the proper microphone adapter fully into the
calibrator. Make sure it fits snugly.
Step 3 Insert the microphone fully into the adapter. Make
sure it fits snugly.
The preferred method for calibrating
with the CAL200 is to mount the
preamplifier in a fixed, vertical
position.
FIGURE 2-1 Vertical Positioning
Step 4 With the microphone connected to the instrument
being calibrated, press the CAL200’s ON button. With a
fresh battery, the calibrator will issue a tone for more than 60
seconds before automatically shutting off (see note below).
Top: calibrator stands
vertically above mounted
preamplifier.
Bottom: mounted
preamplifier,
fixed vertically

2-4 Microphone Calibration CAL200 Manual
Step 5 Make a reading. If the reading is not within
tolerance, refer to the instrument’s manual for instructions
on how to adjust the instrument.
Step 6 Check the ambient noise level to insure that the
calibration was not influenced by noise from external
sources. The CAL200 needs to be on the microphone but
with the CAL200 signal off. External noise levels lower than
89 dB SPL for the 114 dB level calibration, and 69 dB SPL
for the 94 dB calibration level will influence the calibration
by less than 0.015 dB.
Each time the ON button is pressed,
the CAL200 calibrator will run
approximately 60 seconds on a fresh
battery. In order for the calibrator to
turn off, do not press the ON button
again. Wait until the calibrator times
out. It is not an ON/OFF button.
As the battery becomes weaker, the calibration tone will not
deteriorate, but the operating time will decrease until the
time is too short to accomplish an adequate calibration.
When making a sequence of measurements, a calibration
check and an adjustment (if necessary) of the instrument
should be made at the beginning. At the end of the
measurement sequence, the calibration should be checked
again. The inaccuracy of the measurements will be at least
as large as the difference between the level measured for the
initial calibration (or calibration check) and the level
measured for the final calibration check.
Calibration History
Larson Davis strongly recommends that a history of each
calibration adjustment be kept for each piece of equipment.
Normally, most modern equipment requires little or no
adjustment once the initial calibration is performed.
Systematic drifts are possible, and these should be recorded
for corrective action.
Most Larson Davis sound level meters keep a history of each
calibration change that can be printed before an overall reset.
Please refer to the individual instrument manuals for details.
Environmental Precautions
While the CAL200 will perform normally under a wide
variety of gradually changing environmental conditions,
Tabla de contenidos
Otros manuales de Equipo de prueba de Larson Davis
Manuales populares de Equipo de prueba de otras marcas

SMART
SMART KANAAD SBT XTREME 3G Series Manual de usuario

Agilent Technologies
Agilent Technologies BERT Serial Manual de usuario

Agilent Technologies
Agilent Technologies N3280A Manual de usuario

Vernier
Vernier Go Direct Voltage Manual de usuario

Lifeloc
Lifeloc R.A.D.A.R. Manual de usuario

Fluke
Fluke T5-600 Instrucciones de funcionamiento e instalación