Kistler 5495C Series Manual de usuario

Instruction
Manual
Sensor Tester
Type 5495C...
5495C_002-886e-02.20


Foreword
5495C_002-886e-02.20 Page 1
Foreword
Thank you for choosing a Kistler quality product charac-
terized by technical innovation, precision and long life.
Information in this document is subject to change without
notice. Kistler reserves the right to change or improve its
products and make changes in the content without
obligation to notify any person or organization of such
changes or improvements.
© 2020 Kistler Group. Products of the Kistler Group are
protected by various intellectual property rights. For more
information, visit: www.kistler.com. The Kistler Group
includes Kistler Holding AG and all its subsidiaries in
Europe, Asia, the Americas and Australia.
Kistler Group
Eulachstraße 22
8408 Winterthur
Schweiz
Tel. +41 52 224 11 11
www.kistler.com

Sensor Tester Type 5495C…
Page 2 5495C_002-886e-02.20
Content
1.Introduction....................................................................................................................................... 3
1.1Tips for using this instruction manual.................................................................................. 4
1.2Warning symbols and their significance.............................................................................. 4
1.3Basic information................................................................................................................. 5
1.4On transport, storage and handling .................................................................................... 5
2.Important information ....................................................................................................................... 6
2.1Disposal instructions for electrical and electronic equipment ............................................. 6
2.2Software upgrades and updates ......................................................................................... 6
2.3Charging rechargeable batteries......................................................................................... 6
3.Sensor Tester Type 5495C… .......................................................................................................... 7
3.1Accessories......................................................................................................................... 7
3.2Application of the adapter ................................................................................................... 8
3.3Description of hardware/connections.................................................................................. 9
4.General operating button and symbols .......................................................................................... 11
4.1Buttons .............................................................................................................................. 11
4.2Colors................................................................................................................................ 12
4.3Numeric input .................................................................................................................... 13
4.4Alphanumeric input ........................................................................................................... 13
4.5Selection of a value from a list .......................................................................................... 14
4.6Power on/off and sleep mode ........................................................................................... 14
4.7Rechargeable battery status display................................................................................. 15
5.Start screen .................................................................................................................................... 16
6.Analysis of cavity pressure sensor................................................................................................. 17
6.1Sensitivity measurement ................................................................................................... 17
6.2Isolation Test..................................................................................................................... 22
7.Analysis of temperature sensor...................................................................................................... 25
7.1Direct connection of sensor tester to temperature sensor ................................................ 25
7.2Indirect connection of sensor tester to external amplifier ................................................. 28
8.Analyze amplifier ............................................................................................................................ 30
8.1Charge output ................................................................................................................... 30
9.Create/delete/change new customer-specific sensor .................................................................... 33
10.Service............................................................................................................................................ 34
11.Attachment: Example Test Protocol............................................................................................... 35
Total pages 35

Introduction
5495C_002-886e-02.20 Page 3
1. Introduction
Please take the time to thoroughly read this instruction
manual. It will help you with the installation, maintenance,
and use of this product.
To the extent permitted by law Kistler does not accept
any liability if this instruction manual is not followed or
products other than those listed under Accessories are
used.
Kistler offers a wide range of products for use in
measuring technology:
Piezoelectric sensors for measuring force, torque,
strain, pressure, acceleration, shock, vibration and
acoustic emission
Strain gage sensor systems for measuring force and
torque
Piezoresistive pressure sensors and transmitters
Signal conditioners, indicators and calibrators
Electronic control and monitoring systems as well as
software for specific measurement applications
Data transmission modules (telemetry)
Kistler also develops and produces measuring solutions
for the engines, vehicles, manufacturing, plastics and
biomechanics sectors.
Our product and application brochures will provide you
with an overview of our product range. Detailed data
sheets are available for almost all products.
If you need additional help beyond what can be found
either on-line or in this manual, please contact Kistler's
extensive support organization.

Sensor Tester Type 5495C…
Seite 4 5495C_002-886e-02.20
1.1 Tips for using this instruction manual
These instructions describe the handling of the sensor
tester Type 5495C... .
If these instructions become lost, please contact your
Kistler Customer Service Office and request a
replacement.
All of the information and directives contained in these
instructions may be changed in the interests of technical
progress at any time and without advance notification.
Modifications (conversions, retrofittings, etc.) will as a
rule also result in changes to the Instruction Manual.
Inquire at your Customer Service office in such cases to
inform yourself of possible updates in your
documentation.
1.2 Warning symbols and their significance
Noncompliance with the precautionary measures des-
cribed here can lead to malfunctions or to damage to
system components (see table below).
Symbol Noncompliance can result in
Mechanical damage or destruction with possible
accident consequences.
Damage to modules, possibly resulting in system
failure.
Worsening of signal quality, falsification of measure-
ment data or diminishment of efficiency.

Introduction
5495C_002-886e-02.20 Page 5
1.3 Basic information
Check all of the packaging of the sensor tester for
transport damage. Report any damage to the
transport company and the responsible Kistler
representative
Work with the sensor tester should be carried out only
under the specified operating conditions; it should be
protected against the effects of excessive mechanical
strain (shock, vibration)
Protect the inputs and outputs of the sensor tester
against impurities. Attach the covers included in the
delivery if a connection is not occupied and if such
covering is provided
1.4 On transport, storage and handling
Check all of the packaging of the sensor tester for
transport damage. Report any damage to the
transport company and the responsible Kistler
representative without delay when the merchandise
is received
Protect the input and outputs of the sensor tester
against impurities. Attach the covers included in the
delivery to connections and plugs
Use the sensor tester only under the specified
operating conditions. It is to be protected against
excessive mechanical strain (shock, vibration) and
contamination

Sensor Tester Type 5495C…
Seite 6 5495C_002-886e-02.20
2. Important information
2.1 Disposal instructions for electrical and electronic equipment
Do not discard old electronic instruments in municipal
trash. For disposal at end of life, please return this
product to an authorized local electronic waste
disposal service or contact the nearest Kistler
Instrument sales office for return instructions.
2.2 Software upgrades and updates
Kistler may from time to time supply upgrades or updates
for embedded software. Such upgrades or updates must
always be installed.
Kistler declines any liability whatsoever for any direct or
consequential damage caused by products running on
embedded software which has not been upgraded or
updated with the latest software supplied
2.3 Charging rechargeable batteries
The device must be completely charged before
starting operating for the first time.
The rechargeable batteries may only be charged with
the charging device provided.

Sensor Tester Type 5495C…
5495C_002-886e-02.20 Page 7
3. Sensor Tester Type 5495C…
The sensor tester Type 5495C… can measure
the sensitivity of cavity pressure sensors
designed to measure directly or indirectly. The
operation of direct and indirect connected
temperature sensors can also be analyzed. It
can also analyze operation and determine the
quality of the insulation. The measuring chain
can be fully tested, by generating a defined
charge.
3.1 Accessories
1. Case with sheath
2. Sensor Tester (Handheld)
3. Pressure test pin
4. Bluetooth antennas (2 pcs.)
5. Power supply
6. Adapter for pressure test pin
7. Connecting cable for 4-channel connector
8. Connecting cable for 8-channel connector
9. Extension cable for Fischer-BNC
10. Kistler USB Memory Stick 8 GB
11. Cable for charge generation

Sensor Tester Type 5495C…
Seite 8 5495C_002-886e-02.20
All parts in the equipment list can be ordered
individually.
3.2 Application of the adapter
For some sensors, a special adapter must be attached to
the pressure test pin for checking in order not to damage
the sensor.
Using no adapter or using the wrong adapter could
cause the sensor to be damaged or destroyed..
The corresponding adapter is attached to the front part
of the pressure test pin for this purpose.
Tabla de contenidos
Manuales populares de Equipo de prueba de otras marcas

SMART
SMART KANAAD SBT XTREME 3G Series Manual de usuario

Agilent Technologies
Agilent Technologies BERT Serial Manual de usuario

Agilent Technologies
Agilent Technologies N3280A Manual de usuario

Vernier
Vernier Go Direct Voltage Manual de usuario

Lifeloc
Lifeloc R.A.D.A.R. Manual de usuario

Fluke
Fluke T5-600 Instrucciones de funcionamiento e instalación