9. Hose nipple The nozzle gives approximately 160 l/min constant flow at
7,5 bar (94 psi) pressure. Press in the Selector valve (12) to start the flow.
10. Not used.
11. Shutting-off screw with a outlet that gives a constant flow of 7 l/min at 2
bar (29 psi).
12. Selector valve press in to start the flow to the hose nipple (9) and
stopping the flow to the shutting-off screw (11).
13. Medium pressure connection to connect the breathing valve hose of the
regulator unit that is tested. Fitted with a safety valve that opens at 25 bar
(363 psi).
14. Needle valve
15. Secondary pressure gauge for reading the secondary pressure from the
tested pressure regulator.
16. Primary pressure gauge for reading the secondary pressure of the
pressure regulator (4).
17. Test adapter housing to connect the test adapter in.
18. Hose with cover for connection between the hose nipple and the test
adapter.
19. Adapter hose to connect between the 3-way adapter and the medium
pressure connection.
20. Test adapter Spiromatic to connect the breathing valve to the
measurement housing.
21. Adapter for connecting the breathing hose from the regulator unit to the
medium pressure connection.
22. Key for drag pointer to set the primary pressure gauge to zero.
23. 3-way adapter for testing breating valves.
The Spiromatic test is used to do performance tests of Interspiro breathing
apparatuses.
The Spiromatic test is mounted on a mounting plate in an aluminium transport
box. The mounting plate with the test equipment can be removed and secured
to the wall of the workspace if desired.
Remove the six screws to remove the plate.
Spiromatic test
7