idi 1000F Series Manual del producto

Installation and Maintenance Guide
Information for Installation and Users
IDI 1000F Series
Monitor Suspension System
Model IDI 1000F-1H
Adapter to IDI-03 Rails

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 2
The text of this manual was originally written, approved and published by the manufacturer in English.
The information in this manual is subject to change without notice.
CONTENTS
SECTION TITLE
A INTRODUCTION, PRODUCT DATA, AND OPERATION
B SYMBOLS
C GENERAL SAFETY
D SAFETY HAZARDS
E INSTALLATION
F CLEANING
G MAINTENANCE
H CUSTOMER SUPPORT
I DISPOSAL GUIDELINES
J APPROVALS

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 3
A. INTRODUCTION
The IDI 1000F-1H Flat Panel Suspension System transports monitors used in medical imaging
procedures. The system features a counterbalanced arm with vertical travel and internal cable
routing. The mobile package includes clip-mounted ceiling rails.
PRODUCT DATA
Package Includes:
Mobile gantry and counterbalanced swing arm with 36" sweep radius, 20"
vertical travel and +180o rotation
Anodized aluminum ceiling rails with Easymount clip system
Flat panel array with +180o rotation
Complete installation package; meter box mount, all mounting hardware and
cable management components; brackets, covers and “flexhaust”hose
OPERATION
The monitor is moved along the overhead rails, positioned vertically, and
rotated manually using the handle.

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 4
B. SYMBOLS
Attention, consult accompanying documents.
Failure to follow these instructions can cause accidents
resulting in serious injury to patient, user, and damage to
equipment.
Warning! Information or instructions shown near this symbol must be
adhered to in order to prevent a potentially hazardous situation which if not
avoided, could result in death, personal injury or damage to the equipment.
Caution, risk of electrical shock
Protective Ground
This is the common tie point between the AC monitor
cord grounds, frame ground, and service (main) ground.
Alternating Current
Weight Limit
Recycle
Some of the materials can be recycled rather than discarded.

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 5
C. GENERAL SAFETY
Attention, consult accompanying documents.
Failure to follow these instructions can cause accidents
resulting in serious injury to patient, user, and damage to
equipment.
Only qualified persons may install, operate or maintain this equipment.
Installation of this unit must adhere to applicable codes and authorities having any jurisdiction
over this installation. The customer's architect or engineer is responsible for assuring that the
structural support plans comply with all applicable codes and regulations.
The unit should be used only in rooms that comply with state, federal and local recommendations
concerning electrical safety when used in medical installations. All electrical connections shall be done
by licensed/approved electrician per national electric codes.
CAUTION: •Always switch off mains before performing
nonelectrical tests or maintenance.
•Many positions exist where equipment or
patient collision may occur. Care must be
exercised during equipment positioning to
avoid a patient or equipment collision.
Changes and additions to the equipment may be performed only by an authorized representative. These
changes must conform to regulations and accepted standards of good practice. To prevent defeat of the
built-in safety mechanisms, changes must be submitted in writing to the manufacturer for review.
Use only parts specified by Image Diagnostics, Inc. when repairing or servicing this equipment.
This equipment is intended to only be used with Image Diagnostics, Inc. rails.

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 6
D. SAFETY HAZARDS
Installers and Operators using this equipment should understand the safety issues and operating
instructions provided.
Comments and questions regarding safety should be addressed to:
Customer Support
Image Diagnostics, Inc.
310 AUTHORITY DRIVE
FITCHBURG, MA 01420
USA
Or call IDI at (978)829-0009 or send fax to (978)829-0027
Or call Toll Free at (877)304-5434
SAFETY HAZARDS ALERTS
Alert Circumstances for use
DANGER Indicates an imminentlyhazardous situation which,
if not avoided, will result in death or serious injury.
WARNING Indicates a potentially hazardous situation which,
if not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation which,
if not avoided, may result in minor or moderate injury
or equipment damage.
E. INSTALLATION
UNPACK
No special handling is required for unpacking this equipment at the site.
Conventional shipping materials are used.
Recycle or disposal of shipping material per local regulations

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 7
INSTALL CEILING RAILS
1. Refer to the Adapter Plan drawing IDI #D240-224 to determine the rail-to-grid attachment
points.
2. Attach the pair of ceiling rails to the room mounting grid with Rail Mounting Kits. (Refer to the
Longitudinal Ceiling Rail Assembly IDI #D240-514 and Rail Mounting Kit diagram shown
below for information on orientation of rails and mounting kits.) Hand tighten hardware.
3. Position the ceiling rails so that the ends line up with each other and the rails are parallel, then
tighten all mounting hardware.

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 8
INSTALL CEILING RAIL END STOPS
1. The two adjustable end stops provided are to be positioned at diagonal ends of the rails. Load
one Rail End Stop into a rail with the Rubber Bumper facing inward as shown below. Load
the other Rail End Stop into the opposite side of the other rail also with the Rubber Bumper
facing inward. Do not tighten socket set screws on Rail End Stop until Transverse Bridge
Assembly is mounted.
2. Install all four Rail End Caps onto the ends of the Ceiling Rails. Use thread locker Loctite
#242 on all threaded hardware used for the Rail End Caps.
WARNING! Failure to install both pairs of end caps
could result in serious injury to patient,
user, and damage to equipment.

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 9
BUILD TRANSVERSE BRIDGE
1. Remove the two
Ceiling Rail Guide
Wheel Units from the
Rail Adapter Carriage
that has them as shown
to the right.
2. Mount the two Guide Wheel Units into the Ceiling Rail
by tilting them until the horizontal guide wheel can seat
into the bottom cavity of the Ceiling Rail.
3. Remove the two Ceiling Rail Wheel Units from the
other Rail Adapter Carriage as shown to the right. Tilt
those two Wheel Units until they mount into the other
Ceiling Rail.
4. Bolt the two Ceiling Rail Guide Wheel Units to their
respective Adapter Carriage and bolt the two Ceiling
Rail Wheel Units to their respective Adapter Carriage with socket head cap screws and lock
washers. Use thread locker Loctite #242 and tighten hardware to 170 inch-lbs.

Installation and Maintenance Guide for IDI 1000F Series Suspension System Model
IDI 1000F-1H Adapter to IDI-03 Rails
D000-1098 REV C 06-SEP-2018 Image Diagnostics, Inc. 10
INSTALL TRANSVERSE RAILS
1. Raise up one 6ft. 8in. long IDI Rail perpendicular to the Adapter Carriages and hold in place
approximately located in the diagram shown below. Place two pairs of Rail Clips over the
rail flanges under the four-hole mounting patterns in the Adapter Carriages. Insert painted
bolts through Rail Clips and Adapter Carriages and install flat washer, lock washer and
nylon imbedded nuts from the opposite side. Hand tighten bolts. Mount the second IDI Rail
in the same manner.
2. Place the slots of the two Triangulation Bars over the Painted Bolt Studs on the Adapter
Carriages and under the washers and nuts as shown above. Hand tighten the rail mounting
hardware.
3. Slide the IDI Rails back and forth to center them under the Adapter Carriages and to align both
ends. Make adjustments to the Triangulation Bars and Rail Clips until the rails are spaced 17
1/8" apart at both ends as shown below. Tighten hardware.