ideal-tek TEK-SCOPE Manual de usuario

www.ideal-tek.com
Switzerland
11
TEK-SCOPE User Manual
Ver. 1.0

www.ideal-tek.com
Switzerland
2
INDEX
System information 3
Assembly 3
Use with a monitor 4
- Remote control 4
- Remote control pairing 4
- TEK-SCOPE live streaming 5
- How to set your monitor 5
Use with a PC 6
- TEK-SCOPE software 6
- Toolbar 6
- System set up 7
- Measuring 7
- Calibration 7
- Favorite calibrations 8
- Save drawings 8
- Change colors 8
- Angle 8
- Diameter 8
- Lens 8
- Save images as sequential 9
- Save video installation encoder 10
Optional accessories 11
- TEK-TILT 11
- TEK-XY 11
- TEK-DBOX 11
- TEK-DBOX installation 12
Technical facts 13
Safety information 13
Warranty 13
Contact 13

www.ideal-tek.com
Switzerland
3
SYSTEM INFORMATION
TEK-SCOPE is a standalone high denition optical inspection system.
The camera system has the following functions:
• Auto Focus On/Off
• Digital Zoom
• LED lighting with adjustable brightness
• PC Connection Software via DVI / HDMI port
• Video / Image PC Storage Folder
• Measurement & drawing capabilities
TEK-SCOPE can be connected to any monitor with a DVI / HDMI port.
When used with a computer you can install the included TEK-SCOPE Software and make measurement,
drawing and much more.
ASSEMBLY
TEK-SCOPE package includes the following items:
• Inspection system unit
• Remote control
• Power Supply
• Monitor Cable (HDMI – Dvi) - Computer cable (USB – USB mini)
• TEK-SCOPE Software and manual on USB memory stick
1. Use with a monitor:
Place the camera unit on the desired stable location.
Position optional XY table or the tilt table if used.
Connect the monitor to the camera unit (HDMI/DVI) by the monitor cable.
Connect the power supply to a power outlet.
Connect the power supply to the camera unit with the DC cable.
You can now start the system by pressing the button on the top of the camera unit.
If needed, turn the LED lamp on/off by using the button on the back of the unit.
2. Use with a PC:
Connect your computer with the USB cable (USB mini for TEK-SCOPE, USB for computer).
Install the TEK-SCOPE PC software
You can now start the system by pressing the button on the top of the camera unit.
If needed, turn the LED lamp on/off by using the button on the back of the unit.

www.ideal-tek.com
Switzerland
4
1. USE WITH A MONITOR
Remote control
Use the remote control when the device is connected to the monitor.
Z+ Zoom in (+)
Z- Zoom out (-)
4Increase brightness
3Decrease brightness
Press MODE shortly and autofocus will go off / on. When autofocus is off, 3&4 will adjust focus manually.
OK changes magnication to wide angle and shows an articial laser cross. Press OK again to exit and go back
to previous magnication.
Press MODE for 5 seconds and a menu will appear with more choices (user or service). Press OK for more option
and settings:
• Language
• Camera Resolution
• Brightness / Contrast
• etc.
FUNC changes between color and greyscale. Press MODE again to exit.
Remote control pairing
Hold the remote control on the back of the TEK-SCOPE while the power is on, then press the OK button on the
remote control for 5 seconds. When a red light start ashing, the pairing procedure is complete.
Note: The remote control inside the package is paired by the Factory. Changing batteries or using remote with
more than one device may require additional pairing.

www.ideal-tek.com
Switzerland
5
TEK-SCOPE live streaming
Live streaming on TEK-SCOPE is only possible with a unique connection to a Monitor or to a PC.
The Device cannot work properly if connected to a PC and a Monitor at the same time.
How to set your monitor
1. To pair TEK-SCOPE with the screen you are using: press MODE button on the remote for a couple of seconds
to get into the menu.
Service – Camera – Camera resolution – Choose AUTO DETECT
2. To turn on the on-screen magnication display:
Press MODE button on the remote for a couple of seconds to get into the menu.
User – Zoom OSD Function – Choose ON
Your system is now paired with your screen and the magnication is shown on the screen.
If you are using a Tilt table or any higher object, you should replace the lens to a +4 lens.
You can do it using the removal tool. Clean the lens before using the system to avoid ngerprints etc. Notice that
you also need to change the lens settings in the TEK-SCOPE menu.
Press MODE button on the remote for a couple of seconds to get into the menu.
User – lens – Change to +4 or +3 by using the 34 buttons on the remote.

www.ideal-tek.com
Switzerland
6
2. USE WITH A PC
TEK-SCOPE software
Operating System: Windows 10.
After installing the Software, you can access TEK-SCOPE functions:
Toolbar
ZOOM + ,- : camera Zoom Wide ( Decrease ) or TELE ( Increase)
SAVE : left Click to save live picture, choose location folder and name.
Right click on this button will start and save the video. To start the video right click once;
to end the video right click once again. (Codec needs to be installed)
FREEZE : live picture will freeze, press the same button to go back to live again
BRIGHTNESS : increase or decrease brightness of the live image
AF ON/OFF : autofocus ON or OFF. When “off” it is possible to adjust manual focus with the34, this can
be useful if the object under the camera has different heights
CROSS : a cross devides the monitor in 4 parts
IR CUT : change to greyscale, it helps camera if used with IR light
STOP : turns off digital Zoom
DRAW/MEASURE :to activate text, drawing and measuring functions (buttons on the right of the screen)
TEXT : set some texts into the drawing, max 84 characters. Black text on white background
MARKTOOL : set spots on the drawing, for soldering marking, the size follows magnication
HORIZONTAL : horizontal measure, straight lines
VERTICAL : vertical measure, straight lines
FREE DRAWING : free drawing
CIRCLE : draw circle (2 options)
RECTANGLE : draw rectangle
ANGLED LINES : measure in non-horizontal or vertical mode. ( 2 options)
MOUSEWHEEL : up /down to handle Zoom in/out
WIN-CTRL-N : information about the software version
MOUSE LEFT CLICK : save images
Zoom in Freeze Brightness Focus Adjust IR Cut Draw/Measure
Zoom out Save live Picture Autofocus ON/OFF Cross Stop Digital
Mark tool Vertical Draw circle Angled lines
Text Horizontal Free Drawing Draw Rectangle

www.ideal-tek.com
Switzerland
7
System set up
Example of system setup when TEK-SCOPE Software is used:
DO NOT CONNECT THE DEVICE BOTH TO A MONITOR AND PC TOGETHER.
If the computer and the monitor are used together, like extended or dual, the algorithm for magnication/
measuring will be effected.
With the following screens sizes: 20”(1600x900); 22”(1920x1080); 22”(1680x1050); 24”(1920x1080);
24”(1920x1200) measuring will have a tolerance of 1-2% without calibration.
Measuring
Put your object under the camera.
Choose the magnication by the software buttons.
Notice: using the remote control, the Zoom factor will be not updated and measuring will be wrong.
DO NOT USE REMOT CONTROL WHEN CONNECTED TO THE PC BUT USE THE SOFTWARE TOOLBAR.
Press the draw button . Press LB (Left button) on your mouse, select the horizontal or vertical button
and click the starting point. Then go to the second measuring point and press LB on your mouse. The measure
value will appear on the screen.
Calibration
If you need a more accurate measuring, the calibration should be done.
Put a ruler horizontally on the subject you wish to measure. Verify that your magnication is the one you want,
select the drawing mode , right click on the mouse and select “calibration”. A message box will appear, telling
if you can use 1 mm or 10 mm calibration. The calibration setting (1mm or 10mm) depends on the magnication
you use. Press the LB on your mouse to select the length of 1mm or 10 mm. A yellow line will conrm the length
you selected. You can repeat it: just press the draw mode twice and start again.

www.ideal-tek.com
Switzerland
8
Favorite calibrations
If it is a repeating task, you can save your settings. In Draw mode , click RB (Right Button) and select “File”
Save setting, choose a le name and a folder. Your calibration and magnifying values will be saved and could be
used again. Press the OK button with mouse and don’t use the enter key.
Save drawing
Right click also has the functions to save the drawing. You can choose the name and the folder to save your
picture.
Change colors
If you wish to change the color of your drawing lines, right click and choose colors.
The active color will be indicated as (Active).
Angle
Option 1, shows length and angle vs the horizontal.
Option 2, right click and select angle vs line.
Diameter
Default:
Option 1: click the circle button and click two points of the round subject. A circle will show diameter and radius.
Option 2: right click, select “Diameter” and then “Diameter draw”, press left button and move the mouse.
Press 6to get an aiming start point, release 6, repeat if needed. Then press Left button move the mouse and
release.
Lens
First time the Software is started, it will have a default value for the lens that is set to have a +3 lens. If you use
the +3 lens, nothing need to be done.
If the +4 lens is used (with tilt table) do as follows: start the program, right click in the live image and choose lens,
change to lens +4. Done.
Next time the software will be started, it will remember this setting.
If lens it changed time to time, user needs to change the lens setting.

www.ideal-tek.com
Switzerland
9
Save images as sequential
By default images are saved as Index = year_Month_day_Hour_Minute_Second.
Example 170823_11.07.04
This can be changed by right click in the live image
Image Sequential = image name will be Noname_1 and added to the index of every saved image.
Example, NoName_1, Noname_2, Noname_3 ……
Set name for Image Sequential = choose the name of the image and the index will be added.
Example = MyBoard_1, MyBoard_2 ….
Reset index = the value after the name will be set to 1 again, Name will remain until user
changes it.

www.ideal-tek.com
Switzerland
10
Save video installation encoder
A 3part encoder needs to be installed onto the PC.
Available on the installation under folder \encoder. Run the setup .
Then go to Encode settings, ( program Xvid)
Verify that there is a folder named “\ Focus” under \\my pictures. If not, make a new folder and name it Focus.
Tabla de contenidos
Otros manuales de Instrumento analítico de ideal-tek
Manuales populares de Instrumento analítico de otras marcas

EXTOL PREMIUM
EXTOL PREMIUM 8820043 Instrucciones de instalación y funcionamiento

RIDGID
RIDGID SeeSnake Compact Manual de usuario

Teledyne
Teledyne 3000PA Manual de usuario

YDLIDAR
YDLIDAR TIA Manual de usuario

Olympus
Olympus IPLEX FX Series Manual de usuario

Bosch
Bosch GLM 50 C Professional Manual de usuario