Avoiding Injury
Review safety instructions before installing, using, or maintaining the equipment.
♦ Before moving unit, remove contents from the chamber.
♦ Before moving unit, ensure door is closed and latched, and casters are unlocked and free of debris.
♦ Before moving unit, disconnect the AC power cord and secure the cord.
♦ When moving unit, use assistance from a second person.
♦Never physically restrict any moving component.
♦Avoid removing electrical service panels and access panels unless so instructed.
♦ Use appropriate gloves when handling cold internal components and stored inventory.
♦ Keep hands away from pinch points when closing the door.
♦ Avoid sharp edges when working inside the electrical compartment and refrigeration compartment.
♦ Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good
laboratory practices.
♦ Total freezer weight (including contents) is not to exceed 1400 lbs (635 kg).
♦ Use manufacturer supplied power cord only.
♦ Using the equipment in a manner not specied by Helmer may impair the protection provided by the equipment.
♦ Ensure biological materials are stored safely, in accordance with all applicable organizational, regulatory, and
legal requirements.
♦ The freezer is not considered to be a storage cabinet for ammable or hazardous materials.
CAUTION
Decontaminate parts prior to sending for service or repair. Contact Helmer or your distributor for decontamination instructions
and a Return Authorization Number.
1.2 General Recommendations
Intended Use
Helmer ultra-low freezers are intended to provide a controlled temperature environment at ultra-low temperatures required for the
storage of biological materials, pharmaceuticals, and reagents used in a research or clinical laboratory.
The devices referenced in this manual are intended to be operated by personnel who have procedures in place for meeting FDA,
AABB, or any other applicable regulations for the processing and storage of biological materials, pharmaceuticals, and reagents.
General Use
Allow freezer to come to room temperature before switching power on.
During initial startup, the high temperature alarm may sound while the freezer reaches operating temperature.
NOTE
This unit is not a “rapid-freezing” device. Freezing large quantities of liquid, or high-water content items, will temporarily
increase the chamber temperature and will cause the compressors to operate for a prolonged period of time.
Initial Loading
Allow chamber temperature to stabilize at the setpoint before storing pre-frozen product.
Product Loading Guidelines
When loading your freezer, take care to observe the following guidelines:
♦ Never load freezers beyond capacity.
♦ Always store items within shelves.