SAS-3 Operator Manual
6
■Always check the parameters set in the SAS-3 program before starting.
TheparametersmaybedierentthanthosedescribedintheInstructions
For Use. The parameters can be accessed via the edit function.
■Do not block air vents and intakes.
■The instrument should not be moved after installation. Before moving,
the instrument’s shipping bracket, used to lock the applicator sled
in place, must be installed. Improper handling of the applicator sled
could result in the instrument requiring re-alignment, which can only be
completed by a representative of Helena Biosciences Europe.
■The SAS-3 must not be used in areas of high electromagnetic
phenomenon as temporary loss of function may occur. Contact Helena
Biosciences Europe for advice.
1.4 Safety Considerations
The SAS-3 is fully overload protected. However, the following guidelines
should be observed to ensure safe operation of the unit.
■The SAS-3 must only be operated when connected to an earthed supply
of the correct voltage and frequency as indicated on the serial number
label.
■For emergency shutdown, use either the ON/OFF (switch on the right
sideofthedevice)orswitchoatthemainssupply.
■Wait at least 5 seconds after stopping a run before handling output leads
or connected apparatus.
■Always make sure your work area is clean and dry before operating the
SAS-3.
■Iftheequipmentisusedinamannernotspeciedbythemanufacturer
the protection provided by the equipment may be impaired.
■To protect the operator against unintentional contact with possible high
voltage between the electrodes a safety precaution has been introduced.
The high voltage will only be applied to the electrodes when the door is
closed. By opening the door the high voltage will end and a warning
signal will be displayed. Act according to the instructions to continue the
process.
To avoid damage to the instrument:
■For proper ventilation, leave at least 10 cm of space behind the
instrument and on either side.
■Do not operate the SAS-3 in high humidity environments (>90%), or
where condensation may occur.
■To avoid condensation, the unit should be left on between use during
operation in a cold room or in high humidity environments.
■Do not lubricate any parts of the instrument.
■Never place objects on top of the instrument.
■Keepammableliquidsandammablevaporsawayfromtheinstrument
at all times.
HL-2-1514P 2017-01(4) [SAS-3 Operators Manual] EN.indd 6 11/01/2017 10:23