00020EN Rev.09 – 12/2012 4
2INSTALLATION
2.A INSTRUCTIONS AND CHECKS ON DELIVERY
Considering the critical nature of the use of the FLOWFAS H cabinet and the need to keep it in optimum condition,
installation is very important.
FLOWFAS H cabinets are positioned on a pallet, wrapped in an extensible film and contained in a package of
multi-layer strapped cardboard.
After placing the cabinet in its site of use, opening the package and removing the extensible film, check that the
equipment has not suffered any dents or scratches due to transport or improper handling of the package.
In case of any further transport, packing and storage by the user after the initial period of use (e.g.: change of
laboratory or factory), contact the technical assistance service or the distributor for more accurate and precise
instructions or for assistance by specialized technicians.
2.B INSTALLATION REQUIREMENTS
Install the cabinet away from drafts and heat sources (radiators, ventilators/convectors), to ensure proper
functioning.
Install the cabinet in a well-ventilated room with a low degree of dust.
he distance between the cabinet and the room ceiling should be at least 10 cm.
Place the cabinet away from doors and windows, which may cause malfunctions
Place the cabinet in places where there is little human traffic.
he door of the room should be in such a position relative to the cabinet that drafts are prevented.
Use the cabinet whenever possible, together with the external exhaust system package, which should always be
sheltered from rain.
he temperature should never fall below 0°C to prevent the humidity present in the filtering screen from freezing,
which might damage the filters' meshes:
-
max. temperature: 40 °C
-
max. humidity: 80 % at 30°C, linear drop in relative humidity down to 50% relative humidity at 40°C.
Before connect ng the cab net to the ma ns power supply, check the necessary voltage and power
nd cated on the plate near the power cable.
The room must be equ pped w th an earth connect on and connect ons to the gas and/or vacuum
networks, and an a r exhaust duct must be nstalled outs de the bu ld ng.
For the connect on to the gas and/or vacuum networks, read carefully chapter 2C.
The nstallat on s made by techn c ans author zed by Faster s.r.l.