USER MANUAL
Centrifuge-Mixer CM-50M
4
General safety
•Use only as specied in the operating instruc-
tions provided.
•The unit should not be used if it has encoun-
tered a physical shock or has been dropped.
•The unit must be stored and transported in a
horizontal position (see package label).
•After transport or storage, allow the unit to dry
out (2-3 hrs) before connecting to mains power.
•It is necessary to observe the safety area of 300
mm around the centrifuge in accordance with
EN-61010-2-20. Persons and hazardous materi-
als must not be located in the safety area whilst
the centrifuge is in operation.
•Use only original accessories (rotors, adaptors,
etc.) provided by the manufacturer and ordered
specically for this model.
Electrical safety
•Connect to the mains only with a voltage cor-
responding to that on the serial number label.
•Ensure that the switch and plug are easily ac-
cessible during use.
•Do not plug the unit into the main outlet with-
out grounding, and do not use extension lead
without grounding.
•Before moving the unit, disconnect it from the
mains. To turn o the unit, disconnect the pow-
er plug from the mains outlet.
•It is the user’s responsibility to carry out appro-
priate decontamination if hazardous material is
spilt on or inside the equipment. If liquid is split
inside the unit, disconnect it from the mains
and have it checked by a competent person.
During operation
•Do not centrifuge ammable or chemically vig-
orously reactive materials. If such liquids are
spilled in the rotor or rotor chamber, the centri-
fuge must be cleaned with a moist cloth and a
mild soap solution.
•Do not use rotors with visible signs of corrosion,
wear or mechanical damage.
•Do not ll in the containers after they are insert-
ed in the rotor.
•Do not leave the operating unit unattended.
•Do not operate the unit in environments with
aggressive or explosive chemical mixtures.
•Do not operate the unit if it is faulty or has been
incorrectly installed.
•For indoor use only.
•Do not use outside laboratory rooms.
•Before using any cleaning or decontamination
method except those recommended by the
manufacturer, check with the manufacturer
that the proposed method will not damage the
equipment.
•Do not make modications to the unit.
Biological safety
•Without bioseal, the centrifuge is not a biosafe-
ty system in accordance with EN61010-2-20 and
cannot be used for centrifuging hazardous ma-
terials contaminated with toxic, radioactive or
pathogenic microorganisms.
•It is the user’s responsibility to carry out appro-
priate decontamination if hazardous material is
spilt on or inside the equipment.
Preparation for use
•Unpack the device.
•Inspect the appearance of the device, the adapt-
er and the power cord for damage.
•Open the lid of the CM-50M with the emergen-
cy lid opening lever and remove the packaging
material.( see emergency opening of the centri-
fuge lid p.8).
•In the cold season keep centrifuge at room tem-
perature for at least 2 hours.
•Assembly the rotor.
•Rotor must rotate freely without jamming.
•Centrifuge it self determines and congures to
the installed rotor.
•Connect the centrifuge to the adapter and the
adapter to the socket.
•Thus the light indicators will light up.
•In the absence of faults or damage the centri-
fuge can be considered ready to work.
•In the presence of faults and damage, do not
use the centrifuge without consulting a special-
ist.
ATTENTION! Automatic rotor detection system seamlessly detect a type of rotor installed and reprogram the pa-
rameters of the operation accordingly.
ATTENTION! Network socket must correspond to the adapter plug. In case of non compliance with this condition
the rules of electrical safety are violated.