3. CarefullymeshthesplinesontheanglewiththesplinesoftheMotor-to-AngleContra-Angle
Attachment
(seeillustrationbelow).
4. Tightenknurledringontothreadsuntilsnug.
5. Installburaccordingtotheangle’sinstructionmanual.
7LWDQ (
USP/XEH)UHH
0DGHLQWKH86$
-;;;;;
6WDU'HQWDO
Splines
Angle
Motor-to-Angle
Contra-AngleAttachment
Titan-EMotor
TransmissionGear
CAUTIONWhenangledoesnotseatonattachment,gearsarenotmeshedcorrectly.Remove
angle;spinangletransmissiongearandrepeatStep3.
Titan E
®
StarDental
®
Internal
Surfaces
SwivelBearing/
SpringAssembly
Clean
Cleaning & Sterilization
CAUTION
INSTRUCTIONS
Maintenance &Lubrication: TheTitan®-EMotorisLubeFreeand,therefore,requiresnolubrication.Motorshouldberoutinely
maintained,checkedforcorrosion,wornparts,etc.andshouldbereplacedonanasneededbasis.
WARNING
Donotexceed135°C
Donotuse:
Limits on Reprocessing
Thehandpiecemotormustbecleanedandsterilizedbeforerstuseandbetweenpatientuse.
Cleaning-Manual: TheTitan®-EMotorisLubeFreeand,therefore,requiresnolubrication.
Attimesdebrisfromthedeliverysystem’sdriveairtubingcollectsinthemotor.Thisbuild-up
maycausesluggishmotoroperation.Toease,place3dropsofDentaLube II,StarDental®Part
#262539intothedriveairtubeofmotor,andalternatelyoperatethemotorinforwardand
reverse.
Cleaning the Swivel, the Mating Motor Part
Cleanallinternalsurfacesatrearofhandpiecemotorwithacottonswabsoakedin
isopropylalcoholuntilclean.
Cleanallexternalsurfacesusingasmalltoothbrushwithisopropylalcohol.
Tooptimizeoperationoftheswivelconnector,cleanallinternalsurfacesofthemotorwith
acottonswabsoakedinisopropylalcoholuntilclean.Cleanareaaroundbearing/spring
assemblyatrearofswivelusingisopropylalcoholandatoothbrush.Drybothusingsyringeair.
Cleaning-Automated: Washer-disinfectorsarenotrecommended.
Wrapped Instrument
Sterilization:
GravityTypeAutoclave:Pre-VacuumAutoclave:
•Minimumof15minutesat132°C•Minimumof4minutesat132°C
•Minimumof30minutesat121°C
Donotexceed135°C
•dryheatsterilizer
•immersiontechniques
•coldsterilizers •ultrasoniccleaners
•sterilizerbagscontainingdisinfectants •handsoaps
•detergents
•chemicaldisinfectors
•surfacesprays •surgicalmilk
•surfacewipes
Additional Information: Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only paper or
paper/plasticbags.
Storage: Storeinbagaftersterilizationuntilreadytouse.
Repeatedprocessinghasminimaleffectontheseattachments.Endoflifeisnormally
determinedbywearanddamageduetouse.
Drythemotorfor30minutesminimumusingthesterilizationdrycycleandallowthemotorto
cooltoroomtemperatureinbagpriortouse.Ifmoistureispresent,increaseyoursterilizerdry
cycleappropriately.
Drying:
Containment andtransportation: Itisrecommendedthathandpiecesarereprocessedassoonasisreasonablypracticalfollowinguse.
Preparation for cleaning: Wearpuncture-resistantglovesandprotectiveeyewear.Aftereachpatient,ushairbyrunning
handpiecemotorfor20seconds.Removemotorfromswivelifapplicableortubing.
Usesteamsterilizationexclusivelyforcompletesterilityasoutlinedbelow.
Disinfection:
SeeInstallation,OperatingRecommendations,and/orTroubleshootingsectionsofthismanual.
Inspection & function testing:
Single:Astandardsterilizationpackagingmaterialmustbeused.Ensurethatthepackislarge
enoughtocontaintheitemwithoutstressingtheseals.
Packaging:
Manufacturer Contact: Seerstorlastpanelsofthismanual.
StarDental®
Instrument Solutions
StarDental®
Instrument Solutions
7LWDQ (
6WDU'HQWDO
7LWDQ (
6WDU'HQWDO
Drive
Air
RedTip
Applicator
ItistheresponsibilityoftheusertouseonlyFDAcompliantsterilizingequipmentandaccessories
thatarecapableofmeetingthemanufacturersrecommendationsbelow.
CAUTION
Troubleshooting
Problem: LACK OF POWER
Cause: Airpressure
Correction: Verifyproperhandpieceairpressure.RefertoInstallation.
Cause: Crimpedtubing
Correction: Straightentubinginsideandoutsidedeliveryhead.
Cause: Swivelnotthreadedfullyontotubing
Correction: RefertoInstallation.
Cause: Brittlemotororswivelsiliconerubberseal
Correction: Replacesiliconerubbersealfor2/3-lineswivelwithStarDentalPart#203103
and4-linewithStarDentalPart#202897.
Cause: SwivelO-ringsworn
Correction: ReplaceswivelO-rings,StarDentalPart#259884.
Cause: Debrisbuild-upinhandpiece
Correction: Cleanhandpiece.RefertoCleaning.
Problem: SWIVEL WILL NOT DISENGAGE FROM HANDPIECE
Cause: Build-upofdebris
Correction: Place3dropsofDentaLube II,StarDentalPart#262539,intotheswivel/tubing
connector.Rotateswiveltocirculatelube.Disengagehandpiece.Contactdealer
ifunabletodisengageswivel.Topreventreoccurrence,refertoCleaning.
Note: Ifaproblempersists,contactdealerforfurtherassistanceorcontactStarDental technical
serviceat1-866-DTE-INFO.
AlwaysinsistonStarDentalpartsforbestperformance.
Replacement Part Numbers
Titan 3 Titan T
Description Item Number Item Number
AirPressureGaugefor4/6-Line.....................................................262591............................................................262591
SiliconeRubberSeal,4-Line(Pkg.of3).......................................202897...........................................................202897
DentaLube II............................................................................................262539...........................................................262539
ACCESSORIES
Swivel,Motor2/3-Line.......................................................................260152............................................................260152
Swivel,Motor4-Line............................................................................260144............................................................260144
LubeFreeBallBearingAutoLatchAngle......................................264146........................................................... 263950
LubeFreeBallBearingManualLatchAngle................................262738........................................................... 263948
LubeFreeBallBearingAutochuckFrictionGripAngle..................264151........................................................... 263954
LubeFreeBallBearingManualChuckFrictionGripAngle...........262740
BallBearingAutoLatchAngle..........................................................264182
BallBearingManualLatch.................................................................259005...........................................................263947
BallBearingAutochuckFrictionGripAngle...............................264149
TitanEndodonticAngle,AutoLatch..............................................262570............................................................ 262570
TitanEndodonticAngle,ManualLatch........................................262571............................................................ 262571
For new products, manuals and technical information, call
1-866-DTE-INFO or visit our web site at www.dentalez.com.
Important Notice About
Returning Product To The DentalEZ Inc.
Anyone wishing to return any DentalEZ Inc. product to a DentalEZ facility MUST obtain the proper
RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must
appear clearly marked on both the outside and inside of the returned product carton. When calling for
a return authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation
with this procedure will help us expedite processing of any returns.
• Product Name, Model Number or Part Number
• Serial Number of Product
• Dealer Return Purchase Order Number
• Reason For Return
• Copy of Invoice Showing Purchase
Limited Warranty
O-RingsandSealsarespecicallyexcludedfromthiswarranty.
DentalEZ®warrantstheTitan
®-EMotorfor1yearfromthedateofpurchase*.
Pleasenotethefollowingadditionaltermsofourwarrantypolicy:
• Warrantiescovermanufacturingdefectsonlyanddonotcoverdefectsresultingfrom
abuse,improperhandling,cleaning,careormaintenance,normalwearandtearor
non-observanceofoperating,maintenanceorinstallationinstructions.Failure to use
authorized parts or an authorized repair facility voids this warranty.
• Liability is limited to repair or replacement of the defective product at our sole
discretion. All other liabilities, in particular liability for damages, including,
without limitation, consequential or incidental damages are excluded.
• THISWARRANTYISINLIEUOFALLOTHERWARRANTIES,EXPRESSEDORIMPLIED,
INCLUDINGANYIMPLIEDWARRANTIESOFMERCHANTABILITYORFITNESSFORA
PARTICULARPURPOSE.NOEMPLOYEE,REPRESENTATIVEORDEALERISAUTHORIZED
TOCHANGETHISWARRANTYINANYWAYORTOGRANTANYOTHERWARRANTY.
*Providedconditionsinwarrantyaremet.
WARRANTY REGISTRATION
PLEASEVISITTHEWEBSITEBELOWTOREGISTERYOURSTARDENTALPRODUCT.THISWILL
ENSUREPROPERREGISTRATIONANDPROMPTANDACCURATESERVICESHOULDANYOFYOUR
PURCHASESNEEDEVALUATIONORREPAIRDURINGTHEWARRANTYPERIOD.
StarDental®Products:www.dentalez.com/warranty