Shimadzu LC-10/20 Pump 3 Installation Procedure
3 Installation Procedure
3.1 Shimadzu LC-10/20 Pump communication
The Shimadzu LC- 10/20 Pump is controlled by serial (RS232)
communication. The communication is provided via the Optical to USB
convertor connected to the computer via USB A-A cable (p/n SK10) and to
the pump via optical cable (p/n SK15). Both of these cables are part of the
DataApex Opt-USB converter set (p/n OPT01), but can also be ordered
separately.
Optical to USB converter
For communication an OPT-RS232 converter is needed. It can either be
the DataApex Opt-USB converter (p/n OPT01) or Shimadzu adaptor
(Shimadzu p/n 22-35480-91).
Pump Settings
Communication parameters should be set on the pump in the SERVICE
MODE (see pump documentation) as follows:
Baud rate 2400, parity N, bits 8, stop bit 1.
SYS (Set System Parameter) has to be set to 5for isocratic mode or to 6
for gradient mode from the pump keyboard.
Caution: These values are not documented in the pump manual.
The other Pump system parameters should be set to:
Local=0 (Remote)
ADRS= any value (not used)
CBM LINK=0 (not used)
Shimadzu LC-20 settings
Shimadzu LC-20 pumps have to be set to the LC-10 emulation mode in
order to be able to work in Clarity. This is done from the pump's keyboard
according to this procedure:
lPress the VP button on the pump's front keyboard until the CALIBRATION
inscription is shown on the pump's display.
lPress the func button to enter the pump's configuration. The password
has to be filled in.
Note: The default password is 00000.
lPress the func button repeatedly until the OP MODE inscription is shown
on the pump's display.
lEnter the desired number ( 1) to switch the pump to the SCL-10Avp
emulation mode and press the enter button.
- 3 -