6 CRS AFS Purier I
Connecting the Manifold
NOTE: Do NOT remove the black plastic covers from the check
valves until manifold installation is complete!
1. The manifold block must be connected to the gas lines.
The compression ttings on the block t 1/8” stainless
steel tubing (202900) or 1/4” stainless steel tubing
(202902). If you need to connect the lter to lines of
another diameter, adapters will need to be installed rst.
2. Select the bracket, bench-mount (L-shaped) or wall-
mount (at), which you intend to use. Use the bracket
rst to mark the holes on the bench or wall where the
manifold will be mounted.
The bench-mount bracket can also be attached without
screws by using the supplied double sided tape (see
pictures on page 9).
3.Drill the holes to mount the bracket, either to t the
anchors provided or to t the screws without anchors,
depending on the requirements.
• In the case of the Bench Mount II conguration
(see pictures on page 8), the bracket needs to be
screwed into place rst, before mounting manifold to
the bracket.
4. If it is not already in position. Insert the long hex-head
screw into the back of the manifold. Fasten one of the
brackets to the manifold.
5. Fasten the bracket to the bench or wall in the correct
position.