Bioscreen C Pro Manual de usuario

Bioscreen C° ProTM
Operating Manual
Cat. No. 220021
Version 1.0.18
Oy Growth Curves Ab Ltd
2020

2
BioscreenC°ProTM
Bioscreen C° Pro is a photometric microplate
absorbance reader. Experiment protocols for the
Bioscreen C° Pro are created with BioScreener
PRO Software running on a computer connected
to the instrument. Experiment protocol controls
all the functions of the Bioscreen C° Pro
instrument. For a complete description of the
modes of operation, how to choose instrument
settings, etc. refer to the BioScreener PRO
Software User Guide.
Bioscreen C° Pro is used to measure absorbance
from sample liquids in special Honeycomb 2
Microplates. Incubation of the samples can be
carried out in a controlled temperature. The
instrument has three different shaking modes:
linear, orbital and double-orbital shaking.
This manual is designed to guide Bioscreen C° Pro
users through the basic hardware features of the
instrument. Although these instructions were
carefully written and checked, we cannot accept
responsibility for problems encountered when
using this manual. Oy Growth Curves Ab Ltd
reserves the right to make technical
improvements to this equipment and
documentation without prior notice as part of a
continuous program of product development.
This manual supersedes all previous editions.
Product manufactured by:
Oy Growth Curves Ab Ltd
Telekatu 12
FI-20360 Turku
Finland
www.bioscreen.fi
Oy Growth Curves Ab Ltd shall not be liable for
any damages whatsoever arising out of the use or
inability to use this product.
Copyright © Oy Growth Curves Ab Ltd. All rights
reserved. Oy Growth Curves Ab Ltd brand and
Bioscreen C° Pro and BioScreener PRO product
names are trademarks of Oy Growth Curves Ab
Ltd. Other brand and product names are
trademarks or registered trademarks of their
respective holders.

3
Table of Contents
1 Technical Specifications 4
2 Safety Information 5
2.1 Safety symbols and markings used on the
Bioscreen C° Pro instrument 5
2.2 General Information 5
2.3 Environmental Safety Standards 5
2.4 Electrical Safety 5
2.5 Photo biological Safety 5
2.6 Chemical and Biological Safety 6
2.7 Disposal of Packing Material and the
Instrument 6
2.7.1 Disposal of Transport Packing 6
2.7.2 WEEE Compliance 6
2.8 Cleaning and Decontamination of the
Instrument 6
2.8.1 Disinfection Steps 7
3 Intended Use 7
4 Instrument Overview 7
5 Installation 8
5.1 Instrument Installation 8
5.2 Software Installation 8
5.2.1 BioScreener PRO Software 8
5.2.2 Software Installation 9
5.3 Power and Communication Connections 9
5.3.1 Power Connection 9
5.3.2 Ethernet Communication Connection 10
5.3.3 Manual Setup for Direct Connection between the
Bioscreen C° Pro and the Computer 10
5.3.4 Connecting the Bioscreen C° Pro to a Local
Network with DHCP Server 12
5.3.5 Connecting the Bioscreen C° Pro to a Local
Network Manually 13
6 Device User Interface 14
6.1 Changing the Settings of the Instrument 14
7 Running an Experiment 17
7.1 Switching On 17
7.2 Installing the Thermo Cassette 18
7.3 Loading a Microplate 18
7.4 Opening the Cover during the
Experiment 20
7.5 Shutting Down the Instrument 20
8 Maintenance 21
8.1 Preventive and Regular Maintenance 21
8.2 How to Clean the Thermo Cassette 22
9 Service Protocols 23
9.1 Service Contracts 23
9.2 Service Request Protocol 23
9.3 How to Prepare for Service 23
Appendix A - Certificate of Decontamination 25
Appendix B - Guide to manual pipetting 26

4
1TechnicalSpecifications
Bioscreen C° Pro
Vertical filter based photometry
Light Source
Xenon flash lamp
Detector
Photodiode detector
Detection mode
Absorbance
Filters
Filter wheel with up to 8 filters
3 filters included as standard:
405 nm, 600 nm and Wide Band
filter (400-600 nm)
5 empty slots for extra filters purchased
(350, 420, 450, 492, 540, 580, 750 nm)
Optional filters on special request
Thermo Cassette
1-2 pcs simultaneous 100 well
Honeycomb 2 Microplates
Shaking
Linear, orbital and double-orbital shaking
patterns
User-selectable time, speed and
amplitude for each pattern
Incubation
Incubation range (in steps of 0.1°C):
Minimum: 9°C below ambient but
not under +15°C
Maximum: +65°C
Temperature accuracy +/- 0.1°C
Filter based Absorbance
Wavelength range: 350 - 800 nm
OD range: 0 to 4 OD (400 - 800 nm)
OD range: 0 to 3 OD (350 - 400 nm)
Computer Interface
Ethernet
Recommended PC Requirements
Operating System: Windows 10
Processor: Intel Core i5 or equivalent
Memory: 8GB DDR4
Display: 1920 x 1080
Storage: 500GB HDD
Power Requirements
External: 100-240 VAC, 50/60 Hz, max
input current 1.85A/115VAC, 1A/230VAC
Dimensions
Height: 44 cm (17.3 inches)
Width: 30 cm (11.8 inches)
Depth: 59,5 cm (23.4 inches)
Weight
26,6 kg (57.3 pounds)
Ambient conditions
Operating temperature: 5°C to 40°C
Transportation and storage temperature:
-25°C to 50°C
Humidity of atmosphere: 20% to 80%
Non-condensing
Instrument Conformity
Over voltage category II
Specifications are subject to change without
notice.

5
2SafetyInformation
2.1Safetysymbolsandmarkings
usedontheBioscreenC°Pro
instrument
Warning. Hot surface, risk of burns.
Warning. Optical radiation, risk of
injure to eyes or skin.
Manufacturer.
Date of manufacture.
Consult Instructions for use.
Do not treat this equipment as
unsorted municipal waste. Collect
waste electrical and electronic
equipment separately.
2.2GeneralInformation
This instrument is designed to provide full user
protection. When correctly installed, operated
and maintained it will present no hazard to the
user. Read this user manual entirely prior to
operating the instrument. Observe all Warning,
Caution and Note statements as well as safety
symbols and markings on the instrument and in
the documentation.
Warning:
This instrument must be installed and used as
outlined in this Operating Manual. Installation,
service and any operation which requires opening
of the instrument must be performed only by
trained and certified personnel from Oy Growth
Curves Ab Ltd or its representatives. Failure to
comply with these instructions will invalidate the
warranty and can lead to unsafe operation of this
equipment.
Warning:
The area designated for the instrument should be
free of dust, liquids and acidic vapor. The surface
of the table should be flat and even. Avoid areas
subject to vibrations and direct sunlight.
Warning:
Microplates operated at temperatures higher
than 55°C must cool down below 55°C before
being handled and removed from the instrument.
2.3EnvironmentalSafetyStandards
The environmental safety standards for operation
under norm IEC/EN 61010-1 and IEC/EN 61010-2-
010 are met under the following conditions:
Indoor use
Altitude: up to 2000 m
Temperature: +15°C to +35°C
Relative Humidity: maximum 80% at 31°C
non condensing then decreasing linearly
to 50% at 40°C
Mains supply voltage fluctuation: +/- 10%
from nominal
Overvoltage category: II (acc. to IEC
60364-4-443)
Pollution degree: 2 (acc. to IEC 61010-1)
2.4ElectricalSafety
Connect the unit only to a grounded
supply socket.
Connect the unit only to a power supply
with a designated voltage rating
corresponding to the label on the back of
the instrument.
2.5PhotobiologicalSafety
Xenon high power flash lamp light source is
incorporated and encapsulated in the Bioscreen
C° Pro instrument. The intended and routine use
of this instrument and its high power xenon flash
lamp does not pose a photo biological hazard,
given that the user strictly follows these
precautions:

6
Do not operate the instrument unless the
instrument shell is mounted and all
screws are in place and tightened.
Warning:
Xenon high power flash lamp emits ultraviolet
light during operation at levels that could injure
the eyes or skin. Do not look directly at the light
and avoid exposing any skin to the light. During
normal operation the user will not be exposed to
any light from the xenon flash lamp.
2.6ChemicalandBiologicalSafety
Ensure that the bottom of each
microplate is dry. Fluid on the bottom of a
microplate may present a contamination
hazard.
Wash your hands thoroughly after
handling test fluids.
Observe normal laboratory procedures
for handling potentially dangerous
samples.
Use proper protective clothing. Use
disposable gloves.
Ensure that the working area is well
ventilated.
Never spill fluids in or on the equipment.
Do not spill any alkaline onto any
instrument surfaces to avoid damage of
the instrument.
Do not autoclave any part of this
instrument.
Decontaminate the instrument
thoroughly before it is removed from the
laboratory or any servicing is performed
on it. (Follow decontamination
instructions in Chapter 2.8.)
Warning:
Use only mild detergent or 70% ethanol for
cleaning the instrument.
2.7DisposalofPackingMaterialand
theInstrument
The Bioscreen C° Pro is packed in a specially
designed carton. The carton is designed to assure
safe transport and to minimize transit damage.
Please save the original carton, the packing
materials and the pallet for future transportation.
If the unit should need to be returned for repair,
you must use the original packing materials and
carton for shipping.
2.7.1DisposalofTransportPacking
Straps to fix the packing on pallet, packaging
foam and plastic pallet must be disposed of and
recycled according to your local policy of plastic
recycling, if they can’t be saved and stored.
Transport cardboard box must be disposed of and
recycled according to your local policy of paper
recycling, if it can’t be saved and stored.
2.7.2WEEECompliance
The European Union Waste Electrical and
Electronic Equipment Directive (WEEE) is a
Producer Responsibility Directive aimed at
reducing the waste from electrical equipment,
increasing recovery and recycling rates of WEEE
and improving environmental performance of all
operators involved in the life cycle of electrical
and electronic equipment.
A product with WEEE symbol must be collected
separately according to the regulatory guidelines
of specific area or country.
2.8CleaningandDecontaminationof
theInstrument
If there is a risk of contamination with
biohazardous material, the procedure
recommended below or some other
corresponding decontamination procedure must
be performed. It is strongly recommended to
perform the complete decontamination
procedure before relocating the instrument from
one laboratory to another or shipping it to
service.
Warning:
The decontamination procedure should be
performed by authorized trained personnel in a
well-ventilated room wearing disposable gloves,
protective glasses and clothing.

7
Warning:
Working with decontamination agents may be
hazardous. Always observe the safety measures
given by the manufacturer of decontamination
agents. Use only agents specified in this manual.
Warning:
In case there is any doubt about the compatibility
of decontamination agents with parts of the
Bioscreen C° Pro or with materials being in
contact with the agent consult your local
Bioscreen C° Pro representative.
Use suitable disinfectants, e.g. Ethanol (70%).
2.8.1DisinfectionSteps
Basic procedure:
1. Prepare the decontaminant: ethanol 70%
or another agent recommended by your
safety officer.
2. Eject the plate carrier.
3. Take out the thermo cassette.
4. Disinfect the plate carrier and the door
mechanism below it.
5. Disinfect the thermo cassette and its
cover on both inside and outside using a
cloth dampened with 70% ethanol.
6. If you are shipping the device for service,
leave the thermos cassette outside the
device for packaging. Otherwise install
the thermo cassette back in the plate
carrier.
7. Close the door.
8. Switch OFF the power and disconnect the
power cable.
9. Disinfect the outside of the instrument
using a cloth dampened with 70%
ethanol.
10. After performing this decontamination
procedure, enclose a signed and dated
Certificate of Decontamination (Appendix
A, page 25) both inside the possible
transport package and attached to the
outside of the package.
3IntendedUse
The Bioscreen C° Pro photometric microplate
absorbance reader is intended for research or
routine-test laboratories to be used by
professional personnel, who understand the
nature of photometry. The Bioscreen C° Pro is a
microplate photometer for measuring
absorbance from samples in Honeycomb 2
Microplates in 100-well (10x10) plate format. It
also has incubation and shaking capabilities.
Note:
The instrument may only be used for research
and development or for other non-clinical
purposes.
If the assay performance is essential for the
analysis, the test result has to be ensured with
internal quality controls or with an alternative
test. For validation of the entire system, it is
recommended that Good Laboratory Practices
(GLP) are followed to guarantee reliable analyses.
4InstrumentOverview
Figure 4-1: Bioscreen C° Pro front view.

8
Figure 4-2: Bioscreen C° Pro back view.
5Installation
5.1InstrumentInstallation
Note:
When moving the Bioscreen C° Pro, it is
recommended that two people lift the instrument
together to avoid personal injury. Be sure to lift
the instrument from the metal frame at the
bottom to avoid injury and/or damage to the
instrument. Always lift from the metal frame only.
Lifting from the white plastic covers is not
allowed as the plastic parts will break.
Note:
Unpack the instrument according to printed
Bioscreen C° Pro Quick Guide which you can find
inside the shipping carton.
When unpacking the instrument, please check to
ensure that all of the following parts are included:
1. Bioscreen C° Pro Instrument
2. Bioscreen Memory Stick
Bioscreen C° Pro Operating Manual
BioScreener PRO SW User Guide
Bioscreen C° Pro Quick Guide
BioScreener PRO Software
3. Thermo Cassette
4. Power Adapter
5. Power Cable
6. Ethernet Cable
7. Honeycomb Plates (4 pcs)
8. Bioscreen C° Pro Quick Guide
Contact your Oy Growth Curves Ab Ltd
representative immediately if any of these items
are missing.
The area designated for the instrument should be
free of dust, liquids and acidic vapors. The table's
surface should be flat and even. Avoid areas
subject to vibrations and direct sunlight.
A minimum of 10 centimeters of open space is
required on both sides of the instrument. Space
in front of the instrument is required to access
the measurement chamber door.
After unpacking and positioning the instrument,
please do the following steps in the given order:
1. Install the BioScreener PRO software.
2. Plug in power and Ethernet cable.
3. Check settings for Bioscreen C° Pro with
the device user interface (refer to
Chapter 5.3.5).
Follow the instructions in the next chapters.
5.2SoftwareInstallation
5.2.1BioScreenerPROSoftware
BioScreener PRO Software that is installed on a
local computer is needed to access Bioscreen C°
Pro instrument. Software is used to create
experiment protocols and upload them to the
Bioscreen C° Pro. After the experiment run is
completed software is used to download the
results from the Bioscreen C° Pro to the computer
for further processing.
Note:
BioScreener PRO Software is compatible with
Windows 10 operating system.

9
5.2.2SoftwareInstallation
1. Remove the Bioscreen USB memory stick
from the shipping box:
2. Rotate the tail of the USB stick to release
the connector:
3. Insert the USB stick to a USB port of the
computer you want to use to run the
Bioscreen C° Pro instrument.
4. Double-click the software installer (the
file that has .msi file type) in the window
that opens:
Note:
Numbers in parentheses indicates the version
number of the software.
5. Click “Yes” to the window that possibly
opens:
6. The software is installed automatically
and a BioScreener PRO icon appears into
a desktop of your computer:
7. Start the software by double-clicking the
icon.
5.3PowerandCommunication
Connections
5.3.1PowerConnection
Inspect the voltage information on the power
adapter to ensure that it corresponds to the local
main power specifications. Also make sure the
power cable is grounded. Hereafter, the power
cable can be connected to the instrument.
Note:
To prevent power spikes, always connect the
power cable to the Bioscreen C° Pro first and only
after that connect the power adapter to main
power supply.
Warning:
Bioscreen C° Pro instrument must always be used
with the original power adapter only.

10
Figure 5-1: How to connect the power cable.
Figure 5-2: How to connect the power adapter.
Note:
Whenever disconnecting the power cable, please
note that the connector is equipped with a cable
strain relief, and must only be detached by
releasing the lock, sliding the connector mantle
backwards.
5.3.2EthernetCommunicationConnection
Connect the RJ45 Ethernet cable between the
Bioscreen C° Pro and local Ethernet network
(directly to computer or via network router based
on local policies).
Figure 5-3: How to connect the Ethernet cable.
Network settings for the Bioscreen C° Pro are
either entered manually via device user interface
(default) or collected automatically from a DHCP
server.
Note:
If you are unsure which to use, contact your local
IT support for help.
Communication between the Bioscreen C° Pro
and the software is done with normal TCP/IP
protocol. Default connection method is a direct
point-to-point connection between the Bioscreen
C° Pro instrument and the computer. This works
well in environments where the computer does
not need access to other local network resources.
It is also possible to connect the Bioscreen C° Pro
to a local network. In this case the IP settings
need to be set to comply with local network
policies. Consult your local IT specialist for
network details prior connecting the Bioscreen C°
Pro to local network.
5.3.3ManualSetupforDirectConnection
betweentheBioscreenC°Proandthe
Computer
By default, Bioscreen C° Pro is set for direct point-
to-point connection with your computer. If you
are using this connection method, you will only
need to configure your computer’s network
settings to comply with Bioscreen C° Pro’s.
Default IP address for Bioscreen C° Pro is
192.168.1.10
Tabla de contenidos
Manuales populares de Equipo de laboratorio de otras marcas

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manual de usuario

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manual de usuario

NI
NI PXI-5422 Manual

Collomix
Collomix Aqix Manual de usuario

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manual de usuario

Ocean Insight
Ocean Insight FLAME-NIR+ Manual de usuario