medBike
This manual covers operation procedures for the following products:
950-192 medBike
950-193 medBike Calf Supports
Contact Information
Biodex Medical Systems, Inc.
20 Ramsey Road, Shirley, New York, 11967-4704
Tel: 800-224-6339 (Int’l 631-924-9000)
Fax: 631-924-8355
email: supportservices@biodex.com
www.biodex.com

3 Biodex Medical Systems, Inc. © 2017
Table of Contents
Definition of Symbols ................................................................................................................4
Before Proceeding .....................................................................................................................5
Important Safety Information.....................................................................................................6
Biodex Warranty ........................................................................................................................8
1.Introduction ................................................................................................. 10
Intended Use................................................................................................ 10
Indications For Use ....................................................................................... 10
General Cleaning and Maintenance .................................................................. 10
2.Set-up and Operation .................................................................................... 12
Unpackaging and Assembling the medBike ....................................................... 12
Adjusting Pedals on the MedBike ..................................................................... 16
Positioning or Relocating the medBike.............................................................. 17
Display and Handlebars Height Adjustment....................................................... 17
On / Off....................................................................................................... 17
Changing Pedaling Directions While Exercising .................................................. 19
Spasm Control.............................................................................................. 21
Emergency Stop ........................................................................................... 21
3.Active Modes................................................................................................ 23
Active Manual Mode ...................................................................................... 23
Active Resistance Mode (Program A)................................................................ 25
Continuous Watt Mode (Program B)................................................................. 28
4.Passive Modes .............................................................................................. 31
Passive Manual Mode..................................................................................... 31
Passive User Profile Mode............................................................................... 34
Passive Pedal Assisted Mode........................................................................... 38
5.Specifications ............................................................................................... 42
6.Assembly Drawings and Schematics ................................................................ 43
Calf Support Assembly................................................................................... 46
Calf Support Assembly................................................................................... 47

4 Biodex Medical Systems, Inc. © 2017
Definition of Symbols
The following symbols and their associated definitions are used and implied throughout this
manual.
S
S
y
y
m
m
b
b
o
o
l
l
D
D
e
e
f
f
i
i
n
n
i
i
t
t
i
i
o
o
n
n
Carefully read these instructions prior to use
Caution
General Warning
General Mandatory Action
Dangerous Voltage
“On” Power
“Off” Power
Earth (ground)
Alternating Current
Fuse
USB Connector/Cable
Waste in Electrical Equipment
Date of Manufacture
Type B Applied Part
CE Mark
CE Mark for products with EC Certificate
Certified for Safety by ETL Intertek

5 Biodex Medical Systems, Inc. © 2017
Before Proceeding
NOTE: The warnings, cautions and instructions provided in this manual must be
read, followed and kept available for consultation at all times. Observing the
information, instructions and procedures presented throughout this manual is
essential for using this product both properly and safely.
SPECIFIC CAUTIONS
●Allow only qualified, trained personnel to operate or service this product.
●If the equipment is used in a manner other than specified in this operation
manual, the protection provided by the equipment may be impaired and
results could be compromised.
●
Never leave user unattended on device.
EN GARDE SPÉCIFIQUES
●Permettez au personnel seulement autorisé, entraîné de faire marcher ou
assurer l'entretien de ce produit.
●Si l'équipement est utilisé dans une manière autre qu'indiqué dans ce
manuel d'opération, la protection fournie par l'équipement peut être
diminuée et les résultats pourraient être compromis.
●Ne quittent Jamais le patient sans surveillance.
CAUTION: Unauthorized modifications to this product are not permitted and will
void the manufacturer’s warranty. Unauthorized modification of the product may
result in a hazard to the user and/or patient. Do not modify this equipment
without authorization from the manufacturer.
ATTENTION: Les modifications faites sans autorisation à ce produit ne sont pas
permises et va faire le vide la garantie du fabricant. La modification faite sans
autorisation du produit peut s'ensuivre dans un hasard à l'utilisateur et-ou le
patient. Ne modifiez pas cet équipement sans autorisation du fabricant.
Training
This operation manual includes assembly and operating instructions. Operating/assembly
questions can be directed to the service department during business hours.

6 Biodex Medical Systems, Inc. © 2017
Important Safety Information
CAUTION: Federal Law restricts this device to sale by or on the order of a
physician, sonographer or other licensed professional.
ATTENTION: La Loi Fédérale restreint cet artifice à la vente par ou sur l'ordre
d'un docteur, sonographer ou d'autre professionnel agréé.
Follow the unpacking and assembly instructions document.
Before using this equipment, read the entire operation manual carefully.
Failure to read the manual may result in user error or injury. Be sure to save all
provided documents for future reference.
Make certain to understand all warning and caution labels as explained in the
Before Proceeding section of this manual.
This product should be used only as specified in the operation manual.
CAUTION: Biodex devices are designed for use in a client environment.
ATTENTION: Artifices Biodex sont conçus pour une utilisation dans un
environnement du client.
For product specifications, refer to the Table of Contents.
This medical electrical equipment requires special precautions regarding EMC
and must be assembled and placed into service according to EMC information
provided in this manual.
CAUTION: Operation for 950-192: 115 VAC.
ATTENTION: Opération pour 950-192: 115 VAC.
WARNING: Only use approved power supplies.
AVERTISSEMENT: N'utiliser que les alimentations homologuées
CAUTION: To avoid risk of electric shock, this equipment must only be
connected to supply mains with protective earth.
ATTENTION: Pour éviter le risque de choc électrique, cet équipement doit
uniquement être connecté à un approvisionnement conduites avec la terre
protectrice.

7 Biodex Medical Systems, Inc. © 2017
CAUTION: The plug is considered the method of disconnecting the product
from main power. Do not place the product in a position where the plug is not
easily accessible.
ATTENTION: Le bouchon est considérée comme la méthode de déconnexion du
produit d'alimentation. Ne placez pas le produit dans une position où le
bouchon n'est pas facilement accessible.
CAUTION: This product is intended to remain in one location during operation.
It is provided with wheels for relocation which should be used when moving.
ATTENTION: Ce produit est destiné à rester à un endroit pendant le
fonctionnement. Il est muni de roues pour la réinstallation qui devraient être
utilisés lors du déplacement.

8 Biodex Medical Systems, Inc. © 2017
Biodex Warranty
1. Product Warranty
A. This equipment and its accessories are warranted by BIODEX MEDICAL SYSTEMS, INC.
against defects in materials and workmanship for a period of two years from the date of
shipment from BIODEX MEDICAL SYSTEMS, INC. During the warranty period, BIODEX
MEDICAL SYSTEMS, INC. will in its sole discretion, repair, send replacement parts or
replace the equipment found to have such defects, at no charge to the customer.
EXCEPT AS STATED ABOVE, THERE ARE NO WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING WITHOUT LIMITATION WARRANTIES OR MERCHANTABILITY OR FITNESS FOR
USE. BIODEX DOES NOT ASSUME LIABILITY FOR INCIDENTAL, CONSEQUENTIAL OR
INDIRECT DAMAGES INCLUDING LOSS OF USE, SALES, PROFITS OR BUSINESS
INTERRUPTION.
B. This warranty does not apply if the product, as determined by BIODEX MEDICAL
SYSTEMS, INC., is defective due to abuse, misuse, modification or service performed by
other than a BIODEX MEDICAL SYSTEMS, INC. authorized repair representative. Misuse
and abuse include, but are not limited to, subjecting limits and allowing the equipment
to become contaminated by fluid materials.
C. In order to obtain warranty repair service and to expedite repair process, please contact
BIODEX MEDICAL SYSTEMS, INC. Support Services Dept. at 800-224-6339, and select
product support as prompted.
2. Warranty is non-transferable
3. Non-Warranty Service
A. Repairs and/or replacements not covered by this warranty may be performed by
BIODEX MEDICAL SYSTEMS, INC. authorized service representatives.
B. The cost of transportation to and from the service location will be the responsibility of
the customer.

9 Biodex Medical Systems, Inc. © 2017
Service Procedure
If you think you have a service problem, take the following action:
1. Check to see that the problem occurs more than once.
2. Refer to the instruction manual and operations procedure.
If you still think you have a service problem, call BIODEX MEDICAL SYSTEMS, INC.,
Service Department at (800) 224-6339 and select product service as prompted.
1. Keep yourself and the phone next to the equipment.
2. Service will ask you for a brief description of the problem. We will ask specific questions
about the malfunction that occurred. This diagnostic process may take a few minutes,
so call us when you can set aside an uninterrupted block of time.
3. After taking the information, we will advise on the action we will take.
4. Sometimes service personnel must consult with engineering and it may take time to get
back to you. Be sure to let the service representative know your schedule so that we can
call at a convenient time.
5. The return call may be from a person other than whom you first reported the problem
to.
6. After analyzing the problem, we will decide if the unit can be repaired on site, or
replacement parts will be sent.
7. If the unit must be returned, Biodex will provide a return materials authorization
number (R.M.A. #.) Pack the table in the carton that it was originally shipped in. It is the
customer's responsibility for any damage that occurs during shipping.
8. Non-warranty/non-service contract charges for repair are as follows:
a. Materials
+
b. Time
+
c. Travel Zone
Contact information
Biodex Medical Systems, Inc.
20 Ramsey Road, Shirley, New York, 11967-4704
Tel: 800-224-6339 (Int’l 631-924-9000)
Fax: 631-924-8355
email: supportservices@biodex.com
www.biodex.com

10 Biodex Medical Systems, Inc. © 2017
1. Introduction
Intended Use
The medBike is designed to provide a safe ergonomic environment for the clinician and
customer. The design features are intended to relieve musculoskeletal stress while providing
the utmost comfort during exercises.
Indications For Use
The medBike is typically used in cardiovascular facilities, skilled nursing facilities, elder care
fitness centers, and private practice clinics to achieve quality exercise sessions for a variety of
health wellness applications.
General Cleaning and Maintenance
1. As required, cleanse all exterior surfaces and seat pad with a mild detergent solution such
as Parker Laboratories Protex Disinfectant or any one-step disinfectant that does not contain
bleach.
2. Keep pedal/wheel assemblies free of foreign materials and dirt accumulation.
3. Periodically inspect all welds.
4. Periodically check bolts, tighten if necessary.
5. Periodically inspect pedal strap holders.