TK84 Temporary Transducer
Interface
OPERATION
LL
WARNING
Contact with rotating machinery causes
severe injury. Be careful; keep safe
distance.
POWERING TRANSDUCER INTERFACES
The TK84 Temporary Transducer Interface powers transducers such as
Proximitors
or
Bently Nevada Corporation accelerometer interface units.
The output power
is
-24 Vdc at 12 mA maximum. This
is
adequate to
power one Proximitor
or
accelerometer interface unit. Figure 2 shows the
necessary connections. Test leads and alligator clips are supplied with the
package for making these connections.
OPTICAL KEYPHASOR TRANSDUCER
Refer to Figure 3 and the steps below to install the mounting kit.
1.
Clean shaft surface.
2.
Place reflective tape on shaft. A one-inch square of reflective tape
is
adequate for most shafts.
3.
Install optical Keyphasor sensor in pickup holder. Secure with lock
nuts.
4.
Install pickup holder, with optical Keyphasor sensor attached, in the
end of the gooseneck. Tighten pickup holder locking knob.
5.
Connect gooseneck to magnetic base using the threaded adapter. (An
alternate mounting method
is
with the vise grip pliers.
Do
not use the
threaded adapter when connecting gooseneck to vise grip pliers.)
6.
Mount the magnetic base (or vise grip pliers) on the machine.
7.
Loosen lever and adjust gooseneck so pickup
is
about 3 inches from
shaft.
Do
not
bend
gooseneck more than
90
degrees.
8. Tighten lever to secure gooseneck.
4